Detailed Specificaitions optical spectrum analyzer.docx
DOCX document 15 KB Posted
- Attached to
- Optical Spectrum Analyzer Federal contract opportunity
- Solicitation number
- NB686100-24-02254
About this file
This document contains the detailed technical specifications for an Optical Spectrum Analyzer required by the National Institute of Standards and Technology (NIST) under Solicitation Number NB686100-24-02254. The key requirements include:
The system must cover a wavelength range of 350 nm - 1100 nm, with a wavelength resolution better than 30 pm. It must accept multimode optical fibers up to 800 um core size, and have remote control capabilities via RS232 and GPIB ports. Required performance specifications include wavelength accuracy better than ±0.3 nm, measurement noise floor of -75 dBm or better, dynamic range of 60 dB or better, and the ability to handle up to 100 mW input power. Delivery is required within 8 weeks after receipt of order, with a 2-week acceptance period. The system must be delivered to NIST in Boulder, CO.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| NB686100-24-02254 LPTA RFQ - .docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Optical Spectrum Analyzer Detailed specifications
1. The system must cover the wavelength range of 350 nm – 1100 nm.
2. The system must have demonstrated wavelength resolution of better than 30 pm across the operating wavelength range.
3. The system must have demonstrated wavelength resolution of better than 30 pm across the operating wavelength range.
4. The system must be able to accept multimode optical fibers with the fiber core sizes as large as 800 um at the input port.
5. The system must be equipped with RS232 and GPIB ports for remote control of the equipment.
6. We require wavelength accuracy of better than +/- 0.3 nm at the operating wavelength range of 700 nm - 1000 nm.
7. The system must have at least one USB port to be able to accept external USB drives for data storage.
8. The system must have a built-in calibration light source to be used for frequent recalibration of the system.
9. The system must have a measurement noise floor (measurement sensitivity) equal to or better than -75 dBm in the wavelength range of 700 nm to 1000 nm.
10. The system must achieve a measurement dynamic range of equal to or better than 60 dB.
11. The system must be capable of wavelength calibration with an external reference source.
12. The system must be capable of handling upto 100 mW (20 dBm) input power at the wavelength range of 700 nm – 1000 nm.
13. Delivery of 8 weeks after receipt of order.
14. Acceptance is 2 weeks after receipt of order.
15. Deliver to: NIST, 325 Broadway BLVD, Boulder, CO 80305