Nuclear Magnetic Resonance (NMR) Spectrometer Specifications 2-6-18
The contractor shall provide a 300 MHz NMR spectrometer system that is fully computer controlled and has solution-state characterization capability. The spectrometer shall have z-gradient ability and sufficient amplifiers for 1H decoupling for solution-state NMR. The contractor shall provide for delivery and installation of the spectrometer, as well as disassembly, removal, and disposal of the current 300 MHz Bruker Avance NMR system with magnet. Complete specifications are as follows:
Solution-state There shall be two direct detect broad-band probes. One must be broad band down to N14 (X(F)/H with fluorine on BB coil and the other bbo probe (X/H(F) must have access down to at least N15 with 19F on the outer coil. Both must be able to lock on and characterize deuterium. Their operable temperature range must be -100 degrees Celsius to +130 degrees Celsius. They must have z-gradient capability. They must be able to conduct 2 dimensional experiments including inadequate (13C-13C), COSY(1H,1H), HMQC(X, H), and HMQC(X, H).
Autoshim, autolock capabilities for solution-state 1H 2H gradient shimming for solution-state Real time data acquistion Ethernet TCP/IP interface, host to console Automated deuterium gradient shimming for solution state Software and Hardware with ability to run direct and indirect detect experiments including but not limited to: HMQC, HMBC, HSQC, COSY, TOCSY, INADEQUATE, NOESY, DEPT, HETCOR, and direct observation of 1H, and 13C, 15N, 19F, 29Si, 11B, and 31P with 1H decoupling for solution-state Software and Hardware with ability to run direct detect experiments including but not limited to: Direct observation of 13C, 1H, 15N, 29Si, 31P, 27Al, 19F and include 1H (and 19F) decoupling of broadband nuclei.
Automated system and probe calibration capabilities Complete operator’s manual and documentation Samples for calibration and systems tests, sample depth gauge Sample room temperature spinner and high temperature spinner Software for automation of lock, shims, phase, gain, sample insert/spin/eject, acquisiton Acquisition computer with hardware/software installed Installation and delivery of entire system including freight costs Ability to readily insert processed data into Word files
Broadband RF System, 300 MHz Channel 1: 1H and 19F and Channel 2: 15N to 31P capabilities minimum Both channels at least 0.1 Hz frequency resolution
Narrow Bore Magnet 300 MHz Superconducting magnet Flexible Helium transfer line and magnetic field hazard signs Liquid helium hold time of 80 or more days Pulsed field gradient Helium probe and level meter Minimum14 Channel Shims
Digital Variable Temperature Control Unit Accurate to +/- 0.2ºC Accessories to allow for a temperature range of –100ºC to 130ºC in designated probes
Solutions Probes Direct detect PFG (z) Broadband probe, two-channel, X=19F to N14 (X(F)/H coil, Variable Temperature (-100 to +110 deg C). Deuterium locking ability.
Direct detect PFG (z) Broadband probe, two-channel (X/H(F) X= 31P to at least N15 with 19F/1H on the outer coil, Variable temperature (-100 to +110 deg C). Deuterium locking ability.
Installation The spectrometer console must be compatible with a 208V / 30 amp / single phase power source. During installation, the contractor shall supply the power to bring the superconducting magnet up to field, without the use of electrical power from our facility.
Workstation 1, complete with the following:
Acquisition computer Acquisition computer software license Flat panel display Ethernet interface Ethernet cable kit Workstation license Software and hardware installed and a CD read-write for data archiving Laserjet printer/plotter
Workstation 2, complete with the following:
Computer, processing Software license Flat panel display Ethernet interface Ethernet cable kit Workstation license Software and hardware installed and a CD read-write for data archiving Laserjet printer/plotter
Training New user training at vendor’s facility (3 day minimum) On-site training included within a day of installation (1 day minimum)
Warranty The contractor shall provide for a standard commercial warranty.