Minimum_Requirements_OSA.docx

DOCX document 18 KB Posted

Attached to
Optical Spectrum Analyzer
Solicitation number
N66604-18-Q-2000
Issued by

About this file

Optical Spectrum Analyzer minimum requirements.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Features / Minimum Requirements PR 1300707473 Optical Spectrum Analyzer 4/3/18 Optical Spectrum Analyzer

Function
Salient Features / Min Requirements
Wavelength range
1100 – 1700 nm
Wavelength Linearity
+/- 0.02 nm
Wavelength Accuracy
+/- 0.02 nm from 1520-1620nm
Dynamic Range
>55 dB at +/- 0.5 nm from peak
Min Resolution
0.05 nm
Level Sensitivity
-90 dB (1520-1620 nm)
Level Linearity
+/- 0.05 dB
Level Accuracy
+/- 0.4 dB
Level Flatness
+/- 0.2 dB
Max Input power
At least 10 dBm
Optical input
SMF
Connectors
FC
Data Logging
Internal storage & external data transfer
Condition
fully tested and assembled
Calibration
Certificate of calibration required
Warranty
1 year factory warranty (parts and labor)

File details come from the government source that posted it.