Salient Features / Minimum Requirements PR 1300707473 Optical Spectrum Analyzer 4/3/18 Optical Spectrum Analyzer
| Function |
| Salient Features / Min Requirements |
| Wavelength range |
| 1100 – 1700 nm |
| Wavelength Linearity |
| +/- 0.02 nm |
| Wavelength Accuracy |
| +/- 0.02 nm from 1520-1620nm |
| Dynamic Range |
| >55 dB at +/- 0.5 nm from peak |
| Min Resolution |
| 0.05 nm |
| Level Sensitivity |
| -90 dB (1520-1620 nm) |
| Level Linearity |
| +/- 0.05 dB |
| Level Accuracy |
| +/- 0.4 dB |
| Level Flatness |
| +/- 0.2 dB |
| Max Input power |
| At least 10 dBm |
| Optical input |
| SMF |
| Connectors |
| FC |
| Data Logging |
| Internal storage & external data transfer |
| Condition |
| fully tested and assembled |
| Calibration |
| Certificate of calibration required |
| Warranty |
| 1 year factory warranty (parts and labor) |