GENERAL PURPOSE vector network analyzer CALIBRATION STANDARD SPECIFICATION

PURCHASE DESCRIPTION REPORTCSS 03-3101

GENERAL PURPOSE

vector Network Analyzer calibration standard specification

DISTRIBUTION STATEMENT C. DISTRIBUTION AUTHORIZED TO U. S. GOVERNMENT AGENCIES AND THEIR CONTRACTORS TO PROTECT PUBLICATIONS REQUIRED FOR OFFICIAL USE OR FOR ADMINISTRATIVE OR OPERATIONAL PURPOSES ONLY. OTHER REQUESTS FOR THIS DOCUMENT SHALL BE REFERRED TO MEASUREMENT SCIENCE DEPARTMENT, NAVAL SURFACE WARFARE CENTER, CORONA DIVISION, P.O. BOX 5000, CORONA, CA, 91718-5000.

DEPARTMENT OF THE NAVYjanuary 2018

GENERAL PURPOSE Power meter CALIBRATION STANDARD SPECIFICATION

METROLOGY AND CALIBRATION PROGRAM

Measurement Science Department, Naval Surface Warfare Center, Corona Division Page 7 purchase description REPORT

GENERAL PURPOSE

vector network analyzer calibration standard specification

DEPARTMENT OF THE NAVY

METROLOGY AND CALIBRATION PROGRAM

TABLE OF CONTENT

1.0SCOPE………...6
1.1General6
1.2Usage6
1.3Classification6
2.0APPLICABLE DOCUMENTS6
2.1General6
2.2Government Documents7
2.3International Electrotechnical Commission (IEC)7
2.4Institute of Electrical and Electronics Engineers (IEEE)7
3.0EQUIPMENT REQUIREMENTS8
3.1Equipment Components8
3.1.1 Restricted Material8
3.1.2 Frequency Range8
3.2 Design and Construction8
3.2.1 Operator Controls8
3.2.2 Connectors, Electric8
3.2.3 Wiring, Internal8
3.2.4 Internal Cooling8
3.3 Electrical Power Sources and Connections8
3.3.1 Electrical Power Sources8
3.3.1.1 Power Consumption8
3.3.1.2 Lithium Batteries9
3.3.2 Electrical Power Connections9
3.4 Environmental Requirements9
3.4.1 Warm-up9
3.4.2 Temperature and Humidity9
3.4.3 Vibration9
3.4.4 Bounce, Loose Cargo9
3.4.5 Fungus Resistance9
3.4.6 Bench Handling9
3.4.7 Salt Atmosphere9
3.5 Enclosure Physical Characteristics10
3.5.1 Dimensions10
3.5.2 Weight10
3.5.3 Stacking Provisions10
3.5.4 Handles10
3.5.5 Connections and Controls10
3.6 Electromagnetic Compatibility (EMC)10
3.7 Transit Case10
3.8 Digital Interface10
3.9 Safety10
3.10 Marking and Identificaiton10
3.10.1 Equipment Identification10
3.10.2 Firmware and Option Identification11
3.10.3 Calibration Label11
3.11 Calibration Interval11
3.12 Reliability11
3.13 Maintainability11
3.14 Configuration11
3.14.1 Main Box11
3.14.2 Measurement Reference Receivers11
3.14.3 Calibration Kit12
3.14.4 Test Port Cables12
3.14.5 Test Port Cables and Adapters12
3.14.6 RF Connector Requirements12
3.14.7 Accessories12
3.15 Performance12
3.15.1 Frequency Range13
3.15.2 Sweep Type13
3.15.3 Time Base13
3.15.4 Aging Rate13
3.15.5 Dynamic Range13
3.15.6 Source Power Range13
3.15.7 Absolute Power13
3.15.8 Transmission Magnitude Measurement Accuracy13
3.15.9 Reflection Magnitude Measurement Accuracy14
3.15.10 Calibration Methodology15
3.15.11 Calibration Coefficients15
3.15.12 De-embedding Fixture15
3.15.13 VNA Firmware Updates15
3.15.14 Display Screenshot15
3.15.15 Instrument States16
3.15.16 Trace Data16
3.15.17 Marker Analysis16
3.15.18 Trace Statistics16
3.15.19 Limit Lines16
3.15.20 Remote Programming16
3.16 Support Applications16
3.16.1 Uncertainty Calculator16
3.16.2 System Verification Test16
3.17 Operating and Service Manuals17
3.17.1 Performance Verification Tests17
3.17.2 Adjustment Procedures17
3.17.3 Maintenance17
3.17.4 Transmission Measurement Specifications17
3.17.5 Reflection Measurement Specifications17
3.18 Service Provider17

1.0 SCOPE

1.1 General: This calibration standard specification (CSS) defines the minimum mechanical and electrical requirements regarding a Vector Network Analyzer (VNA) system. The primary functionality and responsibility of the equipment is to calibrate passive RF and microwave components such as attenuators, directional couplers, filters, power splitters/dividers, and terminations. The VNA will have the capability to measure forward and reverse transmission/reflection (magnitude and phase) response of a device under test with the range and accuracy requirements provided in this purchase description. The Vector Network Analyzer requires a calibration kit for measurement calibration and test port cables to connect to the device under test. All three components comprise a Vector Network Analyzer system and impact associated measurement capabilities and uncertainties. Stated specification requirements apply to either discrete VNA performance or VNA system level requirements, as explicitly referenced. The VNA system will be used by Navy personnel in shipboard and shore-based laboratories.

1.2 Usage: Test equipment discussed in this purchase description is intended for use in operational conditions within a controlled environment where military equipment is operated. While in transportation or storage, the test equipment may be subjected to conditions aboard military and commercial vehicles on land, sea, and air. The purchase description outlines the usage requirements of test equipment. Specific requirements of MIL-PRF-28800F are applied to test equipment according to defined class of equipment.

1.3 Classification: The Vector Network Analyzer shall conform to Class 4 requirements in accordance with MIL-PRF-28800F for Navy shipboard and shore-based laboratory use as specified in this purchase description.

2.0APPLICABLE DOCUMENTS
2.1General: The documents recorded in section 2 are referenced in this VNA purchase description document. The following specifications and standards form a part of this purchase description to the extent specified herein. Unless otherwise indicated, the issues of these documents shall be those in effect on the date of this solicitation. Document users are cautioned that they must meet all specified requirement documents cited in this purchase description, or as applicable in MIL-PRF-28800F, whether or not they are listed.
2.2Government Documents: The following specifications, standards, andhandbooks form a part of this document to the extent specified herein. Unlessotherwise specified, the issues of these documents are those listed in theissue of the Department of Defense Index of Specifications and Standards(DoDISS) and supplemented thereto.
MIL-PRF-28800F-Test Equipment for use with Electrical and Electronic Equipment, General Specification for
MIL-STD-130-Standard Practice of Identification Marking of U.S. Military Property
MIL-PRF-38793C-Manuals, Technical: Calibration Procedures Preparation
MIL-HDBK-781-Reliability Test Methods, Plans, and environments for Engineering Development, Qualification and Production
MIL-PRF-39012E-Performance Specification, Connectors, Coaxial, Radio Frequency, General Specification For
MIL-STD-348A-Interface Standard, Radio Frequency Connector Interfaces

(Copies of Federal and military specifications and standards are available from the Department of Defense Single Source Stock Point, Building 4D, 700 Robbins Avenue, Philadelphia, PA 19111-5094 [Telephone 215-697-2179]. They are also available at www.everyspec.com.)

2.3International Electrotechnical Commission (IEC):
IEC 61010-1-Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use (General Requirements)

(Copies of IEC documents are available online at http://www.iec.ch/ or EC publications are available in the US from the Sales Department. American National Standards Institute, 11 West 42nd Street,113 Floor, New York, NY 10036, USA.)

2.4Institute of Electrical and Electronics Engineers (IEEE):
IEEE Std 287 (2007)-IEEE Standard for Precision Coaxial Connectors (DC to 110 GHz)

(Copies of IEEE documents are available online at http://www.ieee.org/)

3.0 EQUIPMENT REQUIREMENTS

3.1Equipment Components: The equipment components of the Vector Network Analyzer system shall conform to the requirements stated in paragraph 3.2 of document MIL-PRF-28800F. Manufacturers shall designate a model number for units provided. All units and the replaceable assemblies, subassemblies and parts comprising the unit shall be interchangeable on a non-interference basis with all such designated models provided. The contractor shall immediately inform the procuring activity of any conflict between the interchangeability requirements of this specification and the purchase description.
3.1.1 Restricted Material: Mercury, including mercury batteries; radioactive material;lithium electrochemical cells; organic material that can decay; flammable orexplosive material at operating or storage temperatures; pressurized heliumgas; combinations of materials that can cause decay as a result of the effectsof out-gassing.

3.2 Design and Construction: The design and construction of the Vector Network Analyzer system shall conform to the requirements of paragraph 3.3 of document MIL-PRF-28800F.

3.2.1 Operator Controls: Operating controls and indicators shall be readily accessible to the operator from the front of the VNA. The readings presented on the controls and indicators shall be available to the operator without conversion or calculation.

3.2.2 Connectors, Electric: Provisions shall be made to ensure that connectors will be mated only with the appropriate counterparts. The live or hot side of unmated connectors shall be protected against shorting.

3.2.3 Wiring, Internal: All internal wiring shall be implemented such that the wiring shall not be damaged during normal maintenance and calibration.

3.2.4 Internal Cooling: The VNA shall be capable of operating without interruption for at least 24 hours at the maximum operating temperature for Class 4 equipment (10 to 40 degrees Celsius). Cooling devices used in the equipment shall be an integral part of the equipment and not degrade any aspect of performance or safety.

3.3Electrical Power Sources and Connections: The VNA requirements for powersourceconnections shall be as specified in paragraph 3.3.1 through 3.3.2.
3.3.1 Electrical Power Sources: The VNA shall operate from a source of 103.5Vto126.5V at 50 Hz and 60 Hz 5% single-phase input power.

3.3.1.1 Power Consumption: The VNA maximum power for operation shall not exceed 450 W.

3.3.1.2 Lithium Batteries: Commercial UL approved equipment that contains lithium batteries (size double A or button cell type) used for memory backup is approved for use under a general blanket approval waiver. Lithium or re- chargeable batteries larger than the type specified above would require a separate approval. Tadiran liquid cathode batteries, SO2 (sulfur dioxide) batteries or lithium vinyl chloride “liquid cathode” batteries are strictly prohibited. For the purpose of this specification, a lithium button-type battery is approved.

3.3.2 Electrical Power Connections: The VNA requirements for power sourceconnections shall be in accordance with MIL-PRF-28800 paragraph 3.5.2.
3.4 Environmental Requirements: The VNA, designated Class 4 test equipment,shall meet the environmentalrequirements specified in MIL-PRF-28800paragraph 3.8 with the deviations specified in 3.8.1 through 3.8.7.
3.4.1 Warm-up: After a warm-up period not to exceed 90 minutes, the VNA shallmeet all electrical specifications contained in this specification.

3.4.2 Temperature and Humidity: The VNA shall meet the conditions below:

Temperature (Celsius) Relative Humidity

Operating
10 to 30
5 to 95
Operating
30 to 40
5 to 75
Non-Operating
-20 to 60
Not Applicable

3.4.3 Vibration: The VNA shall meet the random and sinusoidal vibration requirements of MIL-PRF-28800 paragraph 3.8.4.

3.4.4 Bounce, Loose Cargo: The VNA shall conform to the specified performance and accuracy requirements after being tested in accordance with MIL-PRF-28800 paragraph 4.5.5.3.3. In the case of the VNA, the loose cargo bounce test is executed in the Vertical linear mode (straight up and down in vertical plane).

3.4.5 Fungus Resistance: The VNA shall not contain materials that provide nutrients for the growth of fungus as stated in paragraph 3.8.7 of document MIL-PRF- 28800.

3.4.6 Bench Handling: The VNA shall conform to the specified performance and accuracy requirements and there shall be no damage to any controls, indicators, or fuseholders, after being tested in accordance with 4.5.5.4.3. The mechanical calibration kit components shall be tested individually, not within the protective case.

3.4.7 Salt Atmosphere: The VNA has no requirement for salt atmosphere exposure.

3.5 Enclosure Physical Characteristics: The physical characteristics of the VNA equipment enclosure shall be in accordance with MIL-PRF-28800 paragraph 3.6 with the deviations specified in paragraphs 3.5.1 through 3.5.5 below.

3.5.1 Dimensions: The VNA maximum dimensions shall not exceed 18.5 inches (470 mm) in width with handles, 11.5 inches (292 mm) in height with feet, and 23.5 inches (597 mm) in depth including front and rear panel hardware and handles.

3.5.2 Weight: The maximum weight of the VNA shall not exceed 70 pounds (31.82kg).
3.5.3 Stacking Provisions: The VNA enclosure shall have a geometric configurationthat permits stacking without harm to the enclosure or its contents.

3.5.4 Handles: The VNA has no requirement for handles if less than 2U (3.5 inches,

89mm) in height. If the VNA height exceeds 3.5 inches in height, handlesshall be provided. The numberand location ofhandles shall be such that theload distribution per handle shall not exceed 20kg.
3.5.5 Connections and Controls: The operating controls and test ports shall belocated on theinstrument front panel;electrical power connections shall be onthe rear panel; GPIB and LAN connections shall be on the rear panel; otherconnectors canbe located on the front or rear panel. The VNA shall have aminimum of two USB ports to support communication with an external CD/DVDdrive and/or electronic calibration module.
3.6 Electromagnetic Compatibility (EMC): The VNA shall comply to EMC control asindicated in paragraph 3.9 of MIL-PRF-28800.
3.7Transit Case: Transit case is not required for the VNA.
3.8Digital Interface: The VNA shall provide an ANSI/IEEE-STD-488 interface thatconforms to the requirements of MIL-PRF-28800 paragraph 3.4. A LANinterface that conforms to LXI (LAN eXtensions for Instrumentation) standardshall be provided on VNA.
3.9 Safety: For safety ground, all accessible surfaces of the VNA shall be atgroundpotential. The power cable shall include a safety ground conductor. Thisis alsotrue of drawer assemblies; disconnecting the safety ground shall alsodisconnectthe power conductors.
3.10 Marking and Identification: The VNA shall be marked and identified inaccordance with MIL-PRF-28800 paragraph 3.11 and as indicated below.
3.10.1 Equipment Identification: The VNA instrument and paired calibration kit shallinclude at a minimum; manufacturer name, equipment model number,equipment name, place of manufacturer, installed options, and serial numberon identification label/plate.
3.10.2 Firmware and Option Identification: The VNA shall have the ability to displayon the front panel the installed hardware and software options, and firmwareversion upon operator command. If an electronic calibration module isincluded, the VNA shall display the model, serial number, and options uponoperator command.
3.10.3 Calibration Label: The contractor/manufacturer of the VNA shall certify andmaintain substantiating evidence that the instruments meet the operational andgeneral requirements specified herein, and shall supply a calibration labeltothe VNA mainbox and the calibration kit. The label shall contain the datethatthe instrumentwas calibrated. The government reserves the right to requireproof of such conformance and to conduct such tests as necessary to ensure thatall operational and general requirements are satisfied.
3.11 Calibration Interval: Calibration interval requirements shall be in accordancewith MIL-PRF 28800 paragraph 3.12. The manufacturer shall substantiate thecalibration interval by providing technical justification data. The calibrationinterval of the VNA shall ensure a minimum of 85% of the instruments remain intolerance at the end of this interval.
3.12 Reliability: The VNA shall meet the reliability requirements of MIL-PRF-28800paragraph 3.13. The contractor/manufacturer shall provide a summary ofmaintenance or warranty records to establish the validity of the statisticalassertion.
3.13 Maintainability: The design of the VNA shall permit isolation of faults, andrepair down to the modular level (circuit board and sub-assembly level). TheVNA shall be designed to ensure that maintenance access can beaccomplished utilizing ordinary tools. The VNA shall beconstructed so that nodamage to any component shall occur and nopermanent distortion to anystructural member shall be caused during normalmaintenance and calibration.
3.14 Configuration: The VNA system shall consist of a single box solution,twocalibrationkits or electronic calibration modules (Type N and 3.5 mm), and twotest port cables.
3.14.1 Main Box: The VNA main box shall be a benchtop configuration with testportconnectors mounted on the front panel of instrument. VNA computerprocessor/controller shall be internal to main box, an external controller shallnot be required. The VNA shall havetwotest ports. The testport connectorsmust beruggedized(e.g., NMD) and shall consist of 3.5mm male or 2.92 mmmaleconnectors.
3.14.2 Measurement Reference Receivers: The two port VNA shall have twoindependentreference receivers, one for each test port.
3.14.3 Calibration Kit: Either a mechanical calibration kit or an electronic calibrationmodule shall be provided for Type N and 3.5 mm connector types. Themechanical calibration kit shall have female and male components, consistingof a short, open, and load(s). The electronic calibration module shall have aninsertable configuration, one port with a female connector and one port with amaleconnector. The electronic calibration module shall have the capability tosupport a user characterization and storage of such characterization to non-volatile memory. Both a female to female and male to male through adapter isrequired as part of the Type N and 3.5 mm calibration kit to support non-insertablecalibration/measurement configurations. A hard case storage boxshall be provided for calibration kit components.
3.14.4 Test Port Cables: A precision magnitude and phase stable flexible VNA testport cable set shall be provided, which support direct connection to theruggedized test port connectors of the VNA. The device under test (DUT) sideof one test port cable shall be 3.5 mm male, while the other test port cable shallhave a 3.5 mm female connector on the device side. The VNAtest port cablesshall have a length of 36 inches,andsupport a minimum bendradius of 3inches. A flatparallel ridge on thebaseoffemale connector on testport cableis required to support a fixedwrench toprovide a stable referencewhentorquing male connector nutofadapter.
3.14.5 Test Port Cables and Adapters: The VNA test port cable set shall support theconnectivity to precision adapters at the DUT side of cable, providing Type N,
7 mm, 3.5 mm, and 2.92 mm calibration/measurement reference planes.
TypeN, 7 mm,3.5 mm, and 2.92 mmadapters shall meet IEEE STD 287generalprecisionconnector (GPC) specifications.
3.14.6 RF Connector Requirements: RF connectors provided shall meet the criticalinterface dimensions of MIL-PRF-39012 and allow the use of torque wrenchesfor proper connection. RF connector interfaces shall meet critical tolerances inMIL-STD-348A or IEEE STD 287 as applicable to connector type and precisiongrade.
3.14.7 Accessories: Fixed and Torque wrenches shall be provided for properlytorquing ruggedized test port connectors, test port cable connectors, andcalibration components. An ESD antistatic wrist strap is required.
3.15 Performance: The VNA shall meet the specifications and performancecapabilities as listed below under the following conditions; 90 minute warm-upperiod; operating temperature of 20C to 30C; and relative humidity of 20% to60%; unless otherwise stated. Corrected (residual) system performancecriteria is thecombined system accuracy of VNA with associated calibration kitand test portcables, expressed with a coverage factor ofk = 2 (confidencelevelof 95%). Correctedspecifications must apply to minimum IFBW of
10 Hz.Uncorrected (raw data without error correction) performance criteria isexpressedwith a coverage factor of k = 2 (confidence level of95%).
3.15.1 Frequency Range: The VNA system shall have a minimum frequency range of10 MHz to 26.5 GHz.
3.15.2 Sweep Type: The VNA shall support Linear, Log, Frequency Segments,CW,and Power sweeps.
3.15.3 Time Base: The VNA shall have an internal 10 MHz oscillator with aREFERENCE OUTPUT connector, and a REFERENCEINPUT connector tolock onto an external time base.
3.15.4 Aging Rate: The aging rate of VNA internal 10 MHz time base shall not exceed1 x 10-6 per year.
3.15.5 Dynamic Range: The VNA shall have a minimum dynamic range of 90 dB from10 MHz to 26.5 GHz, with an IFBW setting of 10 Hz.
3.15.6 Source Power Range: The VNA source shall have a minimum of 25 dB ofpower range at the instrument test ports, and support an automated test portpower characterization utilizing an external power meter/sensor.
3.15.7 Absolute Power: The VNA shall support absolute power measurements(unratioed) at the measurement reference plane. Calibrating the VNAsourcepower and receivers with external power meter/sensor shall be automatedthroughVNA control.
3.15.8 Transmission Magnitude Measurement Accuracy: The following accuracyrequirements are the combined effects of VNA and applicable Calibration Kit(mechanical or electronic calibration module) with the following criteria;
S11 =S22 = 0 (Linear)forDUT, IFBW = 10 Hz.
Refer to Transmission MagnitudeAccuracy graph.
The accuracy for atransmissioncoefficient (TdB) of0 dB to -10 dB shallbewithin;
0.06 dB, 0.01 to 8 GHz
0.10 dB, 8 to 18 GHz
0.15 dB, 18 to 26.5 GHz
The accuracy for atransmissioncoefficient (TdB) of-10 dB to -60 dB shallbewithin;
0.06 dB + abs[0.001 * (TdB +10dB)], 0.01 to 8 GHz
0.10 dB + abs[0.001 * (TdB + 10 dB)], 8 to 18 GHz
0.15 dB + abs[0.001 * (TdB + 10 dB)], 18 to 26.5 GHz
3.15.9 Reflection Magnitude Measurement Accuracy: The following accuracyrequirements are the combined effects of VNA and applicable Calibration Kit(mechanical or electronic calibration module) with the following criteria;
S21 = S12 = 0 (Linear)forDUT, IFBW = 10 Hz.
Refer to Reflection Magnitude Accuracy graph.
The accuracy for areflectioncoefficient () of0.0 to 0.3 shallbe within;
0.008, 0.01 to 26.5 GHz
The accuracy for areflectioncoefficient () of0.3 to 1.0 shallbe within;
0.008 + [0.04 * ( - 0.3)], 0.01 to 26.5 GHz
3.15.10 Calibration Methodology: The VNA shall support two port SOLT, SOLR, TRL,and LRLcalibration techniques.Non-insertable measurementcalibrationconfigurationsrequiring anadapter(uncharacterized) forthe throughconnection shallbesupportedthrough anUNKNOWN THROUGH(SOLR)calibration algorithm.Unknown through option shall be available when utilizinga mechanicalcalibration kit or electronic calibration module. One portShort/Open/Load and Short/Offset Short/Load calibrations shall be supported.Calibration options for load standard shall include fixed load and/or sliding load.The VNA shall support data-based calibration standards.
3.15.11 Calibration Coefficients: The VNA shall have the capability to store calibrationkit coefficient terms from an external data medium, and support the entry ofcalibration coefficientterms via front panel to support user defined calibrationkits.
3.15.12 De-embedding Fixture: The VNA shall have the capability to de-embed afixture S2P file from Port 1 and/or Port 2, shifting the calibration referenceplane towards the DUT.
3.15.13 VNA Firmware Updates: VNA firmware updates shall be supported on-site,downloadablefrom an external data medium. A summary of critical bug fixesandnew featuresshall be provided with new firmware release.
3.15.14 Display Screenshot: The VNA shall be capable of capturing screen display asan image file format.
3.15.15 Instrument States: The VNA shall have the function to save and recall aninstrument state with active calibration error terms applied.
3.15.16 Trace Data: The VNA shall include trace functions and math analysisincluding;enter Data Memory; view DATA and Memory; math functionsData – Memory, Data + Memory, Data * Memory, Data/Memory. The VNAshall support saving measurement data as a ‘*.SnP’ and ‘*.csv’ file.

3.15.17 Marker Analysis: The VNA marker functions shall include; Max, Min, ∆.

3.15.18 Trace Statistics: The VNA trace statistics shall include; Mean, StandardDeviation, and Peak to Peak values.
3.15.19 Limit Lines: Limit lines shall be available to define PASS or FAIL measurementcriteria.
3.15.20 Remote Programming: The VNA application programming interface (API) shallinclude all major functions supported via front panel operation. All operationmodes, calibration/measurement/analysis functions, stimulus settings, andsave/recall features of the VNA shall be remotely programmable over thedigital interface specified in 3.4, using the command syntax in compliance withthe specifications of Standard Commands for Programmable Instruments(SCPI). While the VNA is in remote mode the front panel shall be disabled untillocal mode is enabled.
3.16Support Applications: An application tool to calculate VNA measurementuncertainties and automate system level accuracy verification shall beprovidedbymanufacturer.
3.16.1 Uncertainty Calculator: The manufacturer shall provide an application programcapable of calculating Transmission and Reflection measurement uncertainties(Magnitude and Phase) for a user defined DUT response and VNA stimulus.Theapplication shall include the following input parameters; VNA hardware(Modeland Options), Calibration Kit (mechanical or electroniccalibrationmodules), Test Port Cable Set, IFBW/Averaging, Calibration Type,Calibrationand Measurement Power Level, and DUT S-parameters (Magnitudecomponent). Uncertainties displayed shall include S11 Magnitude and Phase,S21 Magnitude and Phase, Dynamic Accuracy (Magnitude and Phase) andGroup Delay. The application shall run on a PC with Windows operatingsystem.
3.16.2 System Verification Test: The manufacturer shall provide an applicationprogram capable of running an automated VNA System Level Verification Testusingan applicable Verification Kit. The application shall determine totaleffectivetolerance of VNA system; dependent upon VNA hardware andoptions,Calibration Kit,Test PortCable Set, VNA stimulus, and Verification Kit(with report of calibration). The application shall indicate a PASS or Fail statusafter measuringtheverification kit components and provide a report of testindicating measured value and VNA system tolerance for each measurementpoint. The application shall run directly on the VNA instrument.
3.17 Operating and Service Manuals: The manufacturer shall provide an electroniccopy (PDF format on CD-ROM) of test equipment manuals; covering generalinformation, installation, electrical specifications, operation,calibration,maintenance, service, storage, physical characteristics, and environmentalrequirements. The Government reserves the right to reproduceanyportion ofthe manual(s), including any copyrighted material containedwithin forgovernment use only.
3.17.1 Performance Verification Tests: The manual(s) shall provide supportdocumentation and step by step instructions for the performance verificationtests; including required test equipment and applicable specifications, setupconfigurations, test points and tolerances.
3.17.2 Adjustment Procedures: The manufacturer shall support automatedadjustmentprocedures ofTI using externaltest equipment.Detailedinstructions, setup diagrams, and a list of requiredtest equipmentshall beprovided.
3.17.3 Maintenance: The manual(s) shall include a maintenance section addressingthe philosophy regarding modularcomponent level (circuit card/sub-assembly)repair strategies. Instructions and documents shallbe provided fortroubleshooting, fault isolation, inspection and cleaning,disassembly andreassembly instructions, assembly drawings and diagrams formodular andsub-assemblies, recommended repair components, and any otherpreventativeor corrective maintenance.
3.17.4 Transmission Measurement Specifications: Test equipment specificationsshall include tables depicting the system level accuracy (VNA andapplicablecalibration kit) for transmission coefficients of 10 dB to -90 dB(minimumrange)for frequencies of 10 MHz through 26.5 GHz. Anassumption of
S11 =S22 = 0 Linear is acceptable for specification table.
3.17.5 Reflection Measurement Specifications: Test equipment specifications shallinclude tables depicting the system level accuracy (VNA and applicablecalibration kit) for reflection coefficients of 0.01 to 1 (Linear) for frequencies of10 MHz through 26.5 GHz. An assumption of S21 = S12 = 0 Linear isacceptable for specification table.
3.18 Service Provider: The manufacturer calibration and repair support facilitiesshallbe located within the continental United States.

Transmission Magnitude Accuracy

0.01 to 8 GHz 0 -10 -20 -30 -40 -50 -60 0.06 0.06 7.0000000000000007E-2 0.08 0.09 0.1 0.11 8 to 18 GHz 0 -10 -20 -30 -40 -50 -60 0.1 0.1 0.11 0.12 0.13 0.14000000000000001 0.15 18 to 26.5 GHz 0 -10 -20 -30 -40 -50 -60 0.15 0.15 0.16 0.17 0.18 0.19 0.2 Transmission Coefficient (dB) Accuracy (dB) Reflection Magnitude Accuracy

0.01 to 26.5 GHz 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 8.0000000000000002E-3 8.0000000000000002E-3 8.0000000000000002E-3 8.0000000000000002E-3 1.2E-2 1.6E-2 0.02 2.4E-2 2.8000000000000001E-2 3.2000000000000001E-2 3.5999999999999997E-2 Reflection Coefficient (Linear) Accuracy (Linear) image1.wmf

File details come from the government source that posted it.