Mass Spectrometry/Microbiology Susceptibility Analyzer
Solicitation number
N6264520RCT0001
Issued by

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

The Naval Medical Logistics Command (NMLC) intends to negotiate on a sole source basis (IAW FAR 13.501(a)), Only One Responsible Source. The proposed source is BioMerieux, Inc. (Biomerieux) of 100 Rodolphe St, Durham, NC 27712.

The Naval Medical Logistics Command has a requirement for a contract for one (1) analyzer, laboratory, microbiology, susceptibility, automated [15-091], one (1) analyzer, laboratory, mass spectrometry, microorganism identification [34-062], reagents, quality control, maintenance, and consumables for Naval Medical Center (NMC) San Diego.

The vendor solution shall include one (1) microorganism identification mass spectrometry analyzer, one (1) automated microbiology susceptibility analyzer, and an electronic database.

a) The mass spectrometry analyzer shall be capable of matrix-assisted laser-desorption-ionization time of flight mass spectrometry (MALDI-TOF-MS) for the identification of microorganisms.

b) The mass spectrometry analyzer shall be capable of identification of a minimum of 20,000 microorganisms per year.

c) The mass spectrometry analyzer shall be able to identify microorganisms directly from culture media through the generation of MALDI-TOF mass spectra and comparison with a vendor provided database of clinically relevant microorganisms.

d) The mass spectrometry analyzer shall be FDA approved for identification of gram positive and gram negative bacteria, yeast, nocardia, mycobacterium, and mold: the most commonly encountered clinical isolates to the species level; and the less common isolates to at least the genus level.

e) The mass spectrometry analyzer shall use single-use disposable materials to avoid identification errors due to cross-contamination.

f) The mass spectrometry analyzer shall include a computer and in-vitro diagnostics (IVD database that provides interpretation of mass spectra leading to an identification of the most common test organisms with statistical analysis of confidence in the identification.

g) The mass spectrometry analyzer shall include a research use only (RUO) database from which off-label validations may be performed for less common test organisms.

h) The mass spectrometry analyzer shall require minimal daily maintenance of less than 60 minutes downtime per day.

i) The mass spectrometry analyzer shall have the power requirements of 208V 60Hz

j) The vendor provided solution shall be capable of automated antibiotic susceptibility testing (AST) for appropriate bacterial isolates.

k) The microbiology susceptibility analyzer shall be capable of performing a minimum of 9,000 AST per year.

l) The microbiology susceptibility analyzer shall be able to determine in vitro antibiotic susceptibilities of clinically relevant bacteria isolated in the laboratory.

m) The microbiology susceptibility analyzer shall be automated, capable of processing isolated colonies directly from microbiology culture media, with minimal requirements for manual set up procedures.

n) The microbiology susceptibility analyzer shall include a computer and IVD database that provides interpretation of antibiotic susceptibility breakpoints in accordance with Clinical Laboratory Standards Institute (CLSI) and/or FDA standards.

o) The microbiology susceptibility analyzer shall have the power requirement of 110V 60Hz.

p) The vendor solution shall include an electronic database capable of documenting and tracking the work-up of each microbiological specimen tested in the laboratory.

q) The database shall be able to track the following information for each specimen:

a.

CHCS accession number b.

Type and site of specimen c.

Genus and species designations of all microorganisms isolated d.

Antimicrobial susceptibility results for each isolate tested e.

Additional testing utilized in identification or supplemental AST (e.g. oxidase test, PYR, bile solubility, ceftaroline e-test) f.

Reference to additional ordered tests on the same specimen

r) The database shall be capable of connecting to up to ten (10) workstations currently installed at individual benches or offices in the NMC San Diego laboratory from maintenance of current and queries of past specimen work-ups.

s) The database shall be capable of archiving testing information for all specimens for future access.

t) The database shall be capable of generating antibiograms based upon various specimen identifiers.

u) The analyzers shall be capable of specimen storage stability of at least seven (7) days in a transport container for urine and swab samples stored at a minimum range of 2ºC to 30ºC.

v) Samples shall be able to be kept at -20ºC for up to 90 days if the time of transport/storage exceeds 30 days.

w) The analyzer units shall have integrated software that allows for on-board maintenance/error code generation and automated troubleshooting procedures.

x) The software shall document procedures.

y) The vendor solution shall bi-directionally interface with the Composite Health Care System (CHCS) and with the laboratory information system (LIS), as specified in the Statement of Work (SOW).

z) The analyzer units shall meet the performance characteristics for accuracy and precision, as defined by the 1988 Clinical Laboratory Improvement Act (CLIA). The analyzer equipment shall include sufficient safety features to avoid unnecessary exposure to biohazardous and chemical material.

aa) The units shall include an uninterruptable power supply (UPS) capable of providing all necessary electrical power to each analyzer.

The vendor shall provide installation, validation testing, user training, and preventive maintenance coverage. The vendor shall provide validation of the analyzers, including normal range verification, demonstration of linearity, reportable range, replication studies, recovery studies, and establishment of initial control ranges using control materials. Reagent and supply delivery shall be provided by the vendor, as specified in the SOW. System maintenance coverage and support shall include response times, as specified in the SOW. Training shall be provided by the vendor, as specified in the SOW. The vendor shall provide two (2) operational and two (2) service manuals, each in English. The government anticipates a contract with a Five (5) year ordering period.

The system, including claims made for the product, shall be compliant with Food and Drug Administration (FDA) regulations, with respect to marketing and delivering medical products for use in the United States of America. These requirements shall apply even if delivery is requested outside of the United States of America. The system shall be installed in compliance with the Occupational Safety and Health Administration (OSHA) requirements.

Contractor shall be an Original Equipment Manufacturer (OEM) authorized dealer, authorized distributor, or authorized reseller for the proposed equipment/system such that OEM warranty and service are provided and maintained by the OEM. All software licensing, warranty, and service associated with the equipment/system shall be in accordance with the OEM terms and conditions.

Upon delivery, the contractor shall be responsible for uncrating the unit/system, transporting it through the facility to the location of intended use for installation, and removing of all trash created in this process. If interim storage is required, the vendor shall make arrangements for the storage. The vendor is also responsible for the initial instrument operational and performance validations.

There are not set-aside restrictions for this requirement. The intended procurement will be classified under North America Industry Classification System (NAICS) 334516. This notice of intent is not a request for competitive proposals and no solicitation document exists for this requirement. However, parties interested in responding to this notice will need to submit technical data sufficient to determine capability in providing the same product. All capability statements received by the closing date of this synopsis will be considered by the Government. A determination by the Government not to compete based on responses to this notice is solely within the discretion of the Government. Information received will be considered solely for the purpose of determining whether to conduct a competitive procurement.

If a vendor challenges the basis of this requirement, please email product capability statements (formats for submission: PDF, MS Word, or MS Excel) to Derek Bell at derek.j.bell.civ@mail.mil. Closing date for challenges is no later than 04:00 PM ET on 24 July 2020. No phone calls will be accepted.