NOI_SoleSource Bridge Action Molecular Assay-Analyzer.pdf

PDF 106 KB Posted

Attached to
Notice of Intent to Sole Source to Extend Services for N6264520P0020 - Molecular Assay-Analyzers
Solicitation number
N6264520P0020
Issued by

About this file

This is a Notice of Intent (NOI) from the Defense Health Agency (DHA) to award a sole source bridge contract to BioFire Defense, LLC for molecular diagnostic analyzer systems at Naval Hospital Camp Pendleton, CA and Naval Medical Center Portsmouth, VA. The contract period is April 16, 2025 through October 15, 2025, with one 6-month option period through April 15, 2026.

The contract requires BioFire Defense to provide one PCR analyzer system at each location, along with associated assays, reagents, quality control, maintenance, and consumables. The analyzer must be capable of testing respiratory, blood culture, gastrointestinal, and meningitis/encephalitis targets through integrated sample preparation and nested multiplex PCR analysis. Specific requirements include minimum detection capabilities of 20 respiratory targets, 24 blood culture organisms, 3 antimicrobial resistance markers, 22 GI targets, and 14 ME targets. The system must handle 12 samples minimum, be FDA compliant, and include installation, training, and maintenance. Capability statements from interested vendors are due by 4:30 PM EST on February 28, 2025, limited to 5 pages, and must be submitted via email to margaret.a.suever.civ@health.mil with subject line "Molecular Assay-Analyzer."

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

NOI

The Defense Health Agency (DHA) intends to award a sole source bridge contract under the authority of 41 U.S.C.

1901, to BioFire Defense, LLC, 79 West 4500 South, Suite 14, Salt Lake City, UT 84107 for the provision of all services, materials, and equipment necessary to provide required analyzer, assays, reagents and consumables to Naval Hospital Camp Pendleton, CA (NHCP) and Naval Medical Center Portsmouth, VA (NMCP). The period of performance of the bridge contract is 16 April 2025 through 15 October 2025 with 1 (one) 6-month option period.

BioFire Defense will provide non-personal services for a cost per test (CPT) contract for one (1) analyzer, laboratory, molecular assay, infectious microorganism [device code: 20-802] reagents, quality control, annual maintenance coverage, and consumables for Naval Hospital (NH) Camp Pendleton and Naval Medical Center Portsmouth, VA (NMCP). The automated molecular diagnostic laboratory analyzer shall be able to analyze polymerase chain reaction (PCR) samples to identify molecules of microorganisms within various pathogens. The analyzer shall be capable of testing specifically for respiratory targets, blood culture targets, gastrointestinal (GI) targets and meningitis/encephalitis (ME) targets. The automated molecular diagnostics laboratory analyzer shall be capable of performing at least, but not be limited to, the tests listed below.

Estimated Period of Performance: 16 April 2025 through 15 October 2025

Estimated Period of Performance: 16 October 2025 through 15 April 2026

TEST FOR NH Camp Pendleton TESTS/Per 6 mths

Respiratory Panel Assay 90 Blood Culture Identification (BCID2) 10 Meningitis/Encephalitis Panel Assay (ME) 10 Gastrointestinal Panel Assay (GI) 0 Pneumonia Panel Assay 0

TEST FOR NMC Portsmouth TESTS/Per 6 mths

Respiratory Panel Assay 25 Blood Culture Identification (BCID2) 8 Meningitis/Encephalitis Panel Assay (ME) 3 Gastrointestinal Panel Assay (GI) 10 Pneumonia Panel Assay 0

TEST FOR NH Camp Pendleton TESTS/Per 6 mths

Respiratory Panel Assay 90 Blood Culture Identification (BCID2) 10 Meningitis/Encephalitis Panel Assay (ME) 10 Gastrointestinal Panel Assay (GI) 0 Pneumonia Panel Assay 0

TEST FOR NMC Portsmouth TESTS/Per 6 mths

The PCR analyzer shall be a single-platform walkaway analyzer capable of integrated sample preparation, thermocycle amplification, detection and nested multiplex PCR analysis.. Nested multiplex PCR analysis allows for several test to be conducted at once in one (1) channel. The analyzer shall be capable of deoxyribonucleic acid (DNA) and ribonucleic acid (RNA) extraction and purification. The analyzer shall have a minimum of a twelve (12) sample load capacity. The analyzer shall be able to identify molecules of microorganisms in the following type of samples: swab, positive blood culture, stool, cerebrospinal fluid (CSF), bronchoalveolar lavage (BAL), or sputum.

The analyzer shall be capable of amplifying and detecting a minimum of twenty (20) targets per respiratory panel assay, twenty-four (24) organisms that cause blood stream infections per blood culture identification (BCID) panel assay, three (3) genetic markers that are known to confer antimicrobial resistance per BCID panel assay, twenty-two

(22) targets per GI panel assay, and fourteen (14) targets per ME panel assay. The PCR analyzer shall include bar-code reading capability. The analyzer shall be able to generate, detect and analyze PCR melting curves. The unit shall be able to provide real-time analysis. The analyzer shall perform automatic quality control (QC) and routine operator maintenance without manual intervention. The PCR analyzer shall contain onboard diagnostics and help functions. The dimensions for the molecular assay analyzer shall not exceed 34x24x36 inches (HxWxD). The electrical requirements are 120 VAC and 60 Hz. The analyzer shall be a self-contained unit and shall not require facility hook-ups for water, drainage or heat dissipation.

Corrective maintenance coverage shall be provided and include response times. On-site training shall be provided by the vendor. The vendor shall provide two (2) operational and two (2) service manuals, each in English. Support shall be provided for six (6) months, and one (1) six-month option.

The analyzers, including claims made for the product, shall be compliant with Food and Drug Administration (FDA) regulations, with respect to marketing and delivering medical products for use in the United States of America. These requirements shall apply even if delivery is requested outside of the United States of America. The analyzers shall be installed in compliance with OSHA requirements.

Contractor shall be an Original Equipment Manufacturer (OEM) authorized dealer, authorized distributor, or authorized reseller for the proposed analyzers, such that OEM warranty and service are provided and maintained by the OEM. All software licensing, warranty, and service associated with the analyzers shall be in accordance with the OEM terms and conditions.

Upon delivery, the contractor shall be responsible for uncrating the unit/system, transporting it through the facility to the location of intended use for installation, and removing of all trash created in this process. If interim storage is required, the vendor shall make arrangements for the storage. The vendor is also responsible for the initial instrument operational and performance validations.

There are not set-aside restrictions for this requirement. The intended procurement will be classified under North America Industry Classification System (NAICS) 334516. This notice of intent is not a request for competitive proposals and no solicitation document exists for this requirement. However, parties interested in responding to this notice will need to submit technical data sufficient to determine capability in providing the same product. All capability statements received by the closing date of this synopsis will be considered by the Government. A determination by the Government not to compete based on responses to this notice is solely within the discretion of the Government. Information received will be considered solely for the purpose of determining whether to conduct a competitive procurement.

Sources interested in responding to this notice are required to submit a capability statement that includes management and technical data, in sufficient detail and with convincing evidence that clearly demonstrates the

Respiratory Panel Assay 25 Blood Culture Identification (BCID2) 8 Meningitis/Encephalitis Panel Assay (ME) 3 Gastrointestinal Panel Assay (GI) 10 Pneumonia Panel Assay 0 capability to perform the required work. Capability statements shall not exceed 5 (8.5 x 11 inch) pages using a font size no smaller than 10-point. All capability statements received by the due date of this notice will be considered by the Government. A request for documentation or additional information, or submissions that only ask questions will not be considered as an affirmative response. A determination by the Government not to compete based on responses to this notice is solely within the discretion of the Government. Information received will be considered solely for the purpose of determining whether to conduct a competitive procurement or to proceed with a sole source contract.

Capability statements are due by 4:30 PM EST, 28 February 2025. Capability statements shall be submitted by e-mail ONLY as a Microsoft Word or Adobe PDF attachment to the following address:

margaret.a.suever.civ@health.mil.

IN SUBJECT LINE REFERENCE: "Molecular Assay-Analyzer".

mailto:margaret.a.suever.civ@health.mil