N62645-15-T-1068-00001.pdf
PDF 89 KB Posted
- Attached to
- Brand Name or Equal to Zeiss Lumera 700
- Solicitation number
- N62645-15-T-1068
About this file
This amendment is to respond to questions.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Solicitation_N62645-15-T-1068.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SEE SCHEDULE
AMENDMENT OF SOLICITATION/MODIFICATION OF CONTRACT
Except as provided herein, all terms and conditions of the document referenced in Item 9A or 10A, as heretofore changed, remains unchanged and in full force and effect.
15A. NAME AND TITLE OF SIGNER (Type or print)
30-105-04EXCEPTION TO SF 30
APPROVED BY OIRM 11-84
STANDARD FORM 30 (Rev. 10-83)
Prescribed by GSA
FAR (48 CFR) 53.243
The purpose of this amendment is to respond to questions received as of 24 April 2015 and to revise the solicitation accordingly.
All other terms and conditions remain unchanges.
1. CONTRACT ID CODE PAGE OF PAGES
J 1 5
16A. NAME AND TITLE OF CONTRACTING OFFICER (Type or print)
16C. DATE SIGNED
BY 27-Apr-2015
16B. UNITED STATES OF AMERICA15C. DATE SIGNED15B. CONTRACTOR/OFFEROR
(Signature of Contracting Officer)(Signature of person authorized to sign)
8. NAME AND ADDRESS OF CONTRACTOR (No., Street, County, State and Zip Code) X N62645-15-T-1068
X 9B. DATED (SEE ITEM 11)
20-Apr-2015
10B. DATED (SEE ITEM 13)
9A. AMENDMENT OF SOLICITATION NO.
11. THIS ITEM ONLY APPLIES TO AMENDMENTS OF SOLICITATIONS
X The above numbered solicitation is amended as set forth in Item 14. The hour and date specified for receipt of Offer is extended, X is not extended.
Offer must acknowledge receipt of this amendment prior to the hour and date specified in the solicitation or as amended by one of the following methods:
(a) By completing Items 8 and 15, and returning 1 copies of the amendment; (b) By acknowledging receipt of this amendment on each copy of the offer submitted;
or (c) By separate letter or telegram which includes a reference to the solicitation and amendment numbers. FAILURE OF YOUR ACKNOWLEDGMENT TO BE
RECEIVED AT THE PLACE DESIGNATED FOR THE RECEIPT OF OFFERS PRIOR TO THE HOUR AND DATE SPECIFIED MAY RESULT IN
REJECTION OF YOUR OFFER. If by virtue of this amendment you desire to change an offer already submitted, such change may be made by telegram or letter, provided each telegram or letter makes reference to the solicitation and this amendment, and is received prior to the opening hour and date specified.
12. ACCOUNTING AND APPROPRIATION DATA (If required)
13. THIS ITEM APPLIES ONLY TO MODIFICATIONS OF CONTRACTS/ORDERS.
IT MODIFIES THE CONTRACT/ORDER NO. AS DESCRIBED IN ITEM 14.
A. THIS CHANGE ORDER IS ISSUED PURSUANT TO: (Specify authority) THE CHANGES SET FORTH IN ITEM 14 ARE MADE IN THE
CONTRACT ORDER NO. IN ITEM 10A.
B. THE ABOVE NUMBERED CONTRACT/ORDER IS MODIFIED TO REFLECT THE ADMINISTRATIVE CHANGES (such as changes in paying office, appropriation date, etc.) SET FORTH IN ITEM 14, PURSUANT TO THE AUTHORITY OF FAR 43.103(B).
C. THIS SUPPLEMENTAL AGREEMENT IS ENTERED INTO PURSUANT TO AUTHORITY OF:
D. OTHER (Specify type of modification and authority)
E. IMPORTANT: Contractor is not, is required to sign this document and return copies to the issuing office.
14. DESCRIPTION OF AMENDMENT/MODIFICATION (Organized by UCF section headings, including solicitation/contract subject matter where feasible.)
10A. MOD. OF CONTRACT/ORDER NO.
2. AMENDMENT/MODIFICATION NO. 5. PROJECT NO.(If applicable)
6. ISSUED BY
3. EFFECTIVE DATE
27-Apr-2015
CODE
NAVAL MEDICAL LOGISTICS COMMAND
693 NEIMAN STREET
FORT DETRICK MD 21702-9239
N62645 7. ADMINISTERED BY (If other than item 6)
4. REQUISITION/PURCHASE REQ. NO.
CODE
See Item 6
FACILITY CODECODE
EMAIL:TEL:
N62645-15-T-1068
SECTION SF 30 BLOCK 14 CONTINUATION PAGE
SUMMARY OF CHANGES
SECTION SF 1449 - CONTINUATION SHEET
SUPPLIES OR SERVICES AND PRICES
CLIN 0001
The CLIN extended description has changed from:
Brand Name or Equal
To:
Brand Name or Equal for EGLIN AFB.
CLIN 0002
Brand Name or Equal for KESSLER AFB.
CLIN 0003
Brand Name or Equal for NELLIS AFB.
DELIVERIES AND PERFORMANCE
The following Delivery Schedule item for CLIN 0002 has been changed from:
DELIVERY DATE QUANTITY SHIP TO ADDRESS DODAAC
25-JUN-2015 1 BLDG 0468 5TH ST
KEESLER AIR FORCE BASE MS 39534
FOB: Destination
FM3010
DELIVERY DATE QUANTITY SHIP TO ADDRESS DODAAC
25-JUN-2015 1 KEESLER AFB
BUILDING 0468 5TH STREET
KEESLER AFB
BILOXI MS 39534-2519
FOB: Destination
FM3010
The following have been added by full text:
AMEMENT #1
Answers to questions from Industry as of 24 April 2015.
Q1. Is there a need for a High definition camera on the microscope?
A1. Yes
Q.2 Is there a need for a co-observation tube on microscope?
A2. Yes
Note: Offerors providing Brand Name equipment shall meet the requirement by specific part number for each site.
Offerors providing or Equal equipment shall meet the Essential Characteristics.
The following have been modified:
ESSENTIAL CHARACTERISTICS
BRAND NAME
CLIN 0001 Eglin Air Force Base:
Part Number: 305953-9900-000 Zeiss Lumera 700, Quantity 1
Part Number: 302681-9100-0000 Assistant Microscope with Electrical Zoom, Quantity 1
Part Number: 302681-9024-000 Assistant illumination with slit, Quantity 1
Part Number: 302681-9160-000 Keratoscope, Quantity 1
Part Number: 304977-9023-000 Superlux 180 eye future, Quantity 1
Part Number: 304970-9020-000 Foot Control Panel wireless, Quantity 1
Part Number: 304970-8760-000 Can Bus Cable for Footswitch, Quantity 1
Part Number: 303791-0000-000 Tiltable Binocular Tube, Quantity 2
Part Number: 305810-9015-000 2 Grips for 180 Degrees Tiltable tube, Quantity 2
Part Number: 305810-9003-000 6 pack asespsis Caps, Quantity 2
Part Number: 305542-0000-000 Widefield, Push-in Eyepiece, Quantity 2
Part Number: 302651-9905-000 Objective F=175 APO, Quantity 1
Part Number: 305953-9038-000 Tray with monitor bracket and 22" monitor, Quantity 1
Part Number: 305953-9039-000 Adapter Set 22" monitor/sd video, Quantity 1
Part Number: 305810-9001-000 22mm Asepsis Caps, Quantity 3
Part Number: 305810-9017-000 Asepsis Caps Lumera, Quantity 2
Part Number: 306070-0000-000 Sterile Drapes No 70, Quantity 1
Part Number: 000000-1055-278 Dust Cover for Opmia Blue, Quantity 1
Part Number: 302681-9780-000 Integrated 3CCd Camera, Quantity 1
Part Number: 302681-9760-000 Integrated Medialink, Quantity 1
Part Number: 309438-9002-000 Transport Case Opmi Lumera 700 Flood Stand, Quantity 1
Part Number: 000000-1873-818 SM30-0043-E Opmi Lumera 700, Quantity 2
CLIN 0002 Kessler Air Force Base:
Part Number: 305953-9900-000, Zeiss Lumera 700, Quantity: 1
Part Number: 302681-9100-000, Zeiss assistant microscope with electrical zoom, Quantity: 1
Part Number: 302681-9023-000 SCI Illumination for main and assistant, Quantity: 1
Part Number: 302681-9160-000 Keratoscope, Quantity: 1
Part Number: 304977-9023-000 Superlux 180 Eye Future, Quantity: 1
Part Number: 304970-9020-000 Foot Control Panel Wireless FCP, Quantity:1
Part Number: 304970-8760-000 Can Bus Cable for Footswitch, 6M Long, Quantity: 1
Part Number: 303032-9002-000 Adapter Dovetail 11MM Assembled, Quantity: 2
Part Number: 303791-0000-000 Tiltable Binocular Tube F=170MM, 180 Degrees, Quantity: 2
Part Number: 305542-0000-000 Widefield, Push-In Eye Piece 10x, Quantity: 4
Part Number: 302652-9905-00 Objective, F=200 APO, Quantity: 1
Part Number: 305810-9015-000 2 Grips for 180 Degrees Tiltable Tube, Quantity: 2
Part Number: 305953-9082-000 Additional Display, Quantity: 1
Part Number: 305810-9001-000 22MM Asepsis Caps, Quantity: 2
Part Number: 305810-9017-000 Asepsis Caps Lumera 700, Quantity: 1
Part Number: 000000-1055-278 Dust Cover for Opmia Blue, Quantity: 1
Part Number: 302681-9740-000 Integrated HD Camera, Quantity: 1
Part Number: 309438-9002-000 Transport Case OPMI Lumera 700 Floor, Quantity: 1
CLIN 0003 Nellis Air Force Base:
Part Number: 305953-9900-000, Zeiss Lumera 700, Quantity: 1
Part Number: 302681-9100-000, Zeiss assistant microscope with electrical zoom, Quantity: 1
Part Number: 302681-9023-000 SCI Illumination for main and assistant, Quantity: 1
Part Number: 302681-9160-000 Keratoscope, Quantity: 1
Part Number: 304977-9023-000 Superlux 180 Eye Future, Quantity: 1
Part Number: 304970-9020-000 Foot Control Panel Wireless FCP, Quantity:1
Part Number: 304970-8760-000 Can Bus Cable for Footswitch, 6M Long, Quantity: 1
Part Number: 303032-9002-000 Adapter Dovetail 11MM Assembled, Quantity: 2
Part Number: 303791-0000-000 Tiltable Binocular Tube F=170MM, 180 Degrees, Quantity: 2
Part Number: 305542-0000-000 Widefield, Push-In Eye Piece 10x, Quantity: 4
Part Number: 302652-9905-00 Objective, F=200 APO, Quantity: 1
Part Number: 305810-9015-000 2 Grips for 180 Degrees Tiltable Tube, Quantity: 2
Part Number: 305953-9082-000 Additional Display, Quantity: 1
Part Number: 305810-9001-000 22MM Asepsis Caps, Quantity: 2
Part Number: 305810-9017-000 Asepsis Caps Lumera 700, Quantity: 1
Part Number: 000000-1055-278 Dust Cover for Opmia Blue, Quantity: 1
Part Number: 302681-9740-000 Integrated HD Camera, Quantity: 1
Part Number: 309438-9002-000 Transport Case OPMI Lumera 700 Floor, Quantity: 1
Part Number: 305810-9003-000 6 Pack Asepsis Caps for PD Adjustment, Quantity: 2
OR EQUAL ESSENTIAL CHARACTERISTICS:
Brand name or equal Zeiss Lumera 700 shall be a standard floor stand operating microscope, shall run on a 120V outlet. It shall provide a clear red reflex, provide 100% stereoscopic stereopsis and image quality, shall provide coaxial illumination. It shall have automated microscope translational movement in the x, y and z axes, travel range of max.61mmx61mm It shall have a zoom ratio of 1:6, 10x Wide-angle eyepiece and a focus range of 70mm, shall have translational focus and zoom movement, and shall be controlled entirely through a foot pedal. It shall have automated tilt capability and have several inches of clearance between the operative filed and the scope. It shall have a bright LED fiberoptic light illumination that provides natural appearing colors, have stereo coaxial illumination capabilities, and have an integral UV filter. It shall be a high-definition camera, and have a co-observation tube. It shall have an assistant scope with stereo vision capabilities and independent focusing with no light loss for the main surgeon. It shall have video recording ability via USB, integrated 22’ HD monitor and shall have a backup lamp and it shall have an integrated keratoscope. Floor mounted is preferable, but not required so that the scope is portable.
If the system/device uses a Microsoft Operating System (OS), it shall be Windows 7 or higher. If required to install software on end users computers, it shall run on an Air Force Standard Desktop Configuration (SDC) computer. All servers that require Windows OS shall run Windows Server 2008 or higher.
All equipment and accessories shall be new from the manufacturer. Used or refurbished equipment, or equipment obtained from a third party shall not be accepted. Quoter shall be an Original Equipment Manufacturer (OEM) authorized dealer, authorized distributor or authorized reseller for the proposed equipment/system such that OEM warranty and service are provided and maintained by the OEM. All software licensing, warranty and service associated with the equipment/system shall be in accordance with the OEM terms and conditions.
Regulatory Requirements
The equipment shall be compliant with the Food and Drug Administration (FDA) requirements to market and deliver medical products for use in the United States of America, even should delivery be requested outside of the
United States. The claims made for the product shall comply with the regulations of the FDA with respect to products for marketing and delivery of a medical product for use in the United States of America, even should delivery be requested outside of the United States. The system shall be installed in compliance with OSHA requirements.
(End of Summary of Changes)