The following questions were proposed:

Answer: Government response to your questions:

1) 4.2.3.4 - Camera Housing: change of pipe size from 4" with two 1/2" ports to 1-1/2" pipe and no ports is authorized.

2) 4.2.3.5 - LED Light Housing: change of pipe size from 4" with two 1/2" ports to 1-1/2" pipe and no ports is authorized. It is PSNS&IMF intention to use the suggested model or Model 4173KHB34A

3) 4.2.3.9 - Test Port: Test port may be omitted.

Questions:

The questions I have are regarding SOW items 4.2.3.4, 4.2.3.5, and 4.2.3.9. We can make these in any configuration that you would like, however we recommend the following:

1. 4.2.3.4 - Camera Housing: We recommend not using 4” pipe and instead using 1-1/2” pipe. This 1-1/2” pipe will accommodate the Mirion DOT CAM. If you use 1-1/2” pipe instead of the 4” pipe, you will not have the two ½” ports that would’ve been in the 4”pipe.

1. 4.2.3.5 – LED Light Housing: We recommend not using 4” pipe and instead using 1-1/2” pipe. Please provide information on the LED light that you intend to use so that we can make sure it is properly housed. We recommend using a IST Model 4173LHB24A, or similar LED that can be housed in the 1-1/2” pipe.

1. 4.2.3.9 – Test Port: Do you require the Test Port or can it be omitted?