Statement of Work APN7-16 Requirement A. Background The use of NAVAIR METCAL equipment funding is to procure Calibration Standards (CALSTDs) and ancillary equipment for Aviation Fleet Intermediate Calibration Activities, Fleet Training Activities and Aviation Navy Calibration Laboratories (NCLs) to support Aviation Organizational and Intermediate (I) level maintenance assets. All Cal Standards listed for procurement must first go through an approval process. This process includes: initial submission by the fleet; review and prioritization by the Fleet TYCOM’s; laboratory workload justification and standardization review by a NAVAIR METCAL technical review team; overall procurement strategy review to assess overall program objectives, priorities and funding.
REQUIREMENTS:
1. Mfr. one year warranty with sticker affixed to each unit plus a warranty certificate.
2. Calibration Certificate and calibration sticker (as applicable for each unit)
3. Units shall be serialized
4. Units (over $5,000) shall have an IUID Label applied per DFAR 252.211-7003.
5. Manufacturer’s Standard Operating and servicing manuals (CD version acceptable) shall be provided with each unit. The DFARS clause listed below must be referenced in all contracts and the manufacturer made aware that the use/reproduction of tech data/manuals (for government use only) will follow these use guidelines.
DFARS SUBPART 227.71 - Rights in Technical Data 7.
The US government shall have limited technical data rights for all technical manuals and documents furnished with the procured commercial items, for the purpose of Government use only in duplicating or posting technical documentation to limited access web sites for use by US Government Personnel and Contractors engaged in maintenance, calibration or related logistics activities associated with the procured commercial items in accordance with DFARS 252.227-7015 Technical Data--Commercial Items
B. Tasking:
The contractor shall provide the following NEW:
1) P/N CW801P Continuous Wave AC Power Source Qty 8 To include the following:
1. Input:
i. Power : 800VA
ii. Voltage: 90 - 264 VAC
iii. Current: 13 ARMS max
iv. Frequency: 47 to 63 Hz
v. Phases: single-phase
vi. Power Factor: >0.99 typical at full load nominal line
vii. Efficiency: >73% typical at full load
2. Output:
i. Power: 800 VA
ii. Voltage ranges: 0 to 135 Vrms, 0 to 270 Vrms, user selectable
iii. Accuracy (>5VAC): ±0.1% of range <100 Hz, ± 0.2% of range >100 Hz
iv. Resolution: 0.1 Vrms
v. Total harmonic distortion: 0.65% typical <100Hz add 0.5%/100 Hz above 100 Hz
vi. AC noise level (typical): <50 mVRMS
vii. Amplitude stability¹: ±0.05% of full scale
viii. Load regulation: ±0.1% of full scale voltage for a full resistive load to no load (<10 mVRMS typical, measured at point of sense)
ix. Line regulation: ±0.1% of full scale voltage for a ±10% line change from nominal line voltage (<5 mVRMS typical, measured at point of sense)
x. Remote voltage sense: 5 Vrms total lead voltage drop
xi. Current:
1. 135VAC Range: 6.0 ARMS
2. 270VAC Range: 3.0 ARMS
3. Accuracy: ± 0.5% max
4. Resolution: 0.01 ARMS
xii. Frequency Range:
1. Range: 45 to 500 Hz, 45 to 1000 Hz
2. Accuracy: ±0.02% max
3. Resolution: 0.1 Hz, 0.01 Hz for remote programming
4. Phase: single phase output. Multi-phase system configuration with Digital Expansion Cable
5. Power factor of load: 0 lag to 0 lead
3. Physical:
i. Height: 3.5 in.
ii. Width: 19 in.
iii. Depth: 20.07 in.
iv. Weight: 48 lbs. (22 kg)
4. Environmental:
i. Operating Temperature: 0 to 40°C
ii. Storage Temperature: -40 to +70°C
iii. Humidity Range: 0 to 85% at 25°C derate to 50% at 40°C (non condensing)
iv. Altitude: Operating full power available up to 6,000 feet, non-operating to 40,000 feet
v. Cooling: Dual fan speed with side air intake, exhaust to rear
5. Measurements:
i. Power: 800VA
ii. Voltage:
1. Range: 0 to 270 Vrms, 0 to 310VRMS
2. Accuracy² (VAC >5V): ±0.1% of range <100 Hz, ± 0.2% of range>100 Hz, ± 0.3% of range>500 Hz
3. Resolution: 0.1 Vrms
iii. Current³:
1. Range: 0 - 6.0 ARMS
2. Accuracy: ±0.5% of range for linear loads
3. Resolution: 0.01 ARMS
iv. Peak Current³:
1. Range: 0 to 25 A
2. Accuracy: ±1% of range
3. Resolution: 0.1 A
v. Frequency:
1. Range: 45 to 500 Hz, 45 to 1000 Hz
2. Accuracy: ±0.02% max
3. Resolution of display: 0.1 Hz
vi. Power³:
1. Range: 0 - 800 W
2. Accuracy: ±2% of range for linear loads
3. Resolution: 1 W
vii. Apparent Power³:
1. Range: 0 to 800 VA
2. Accuracy: ±2% of range for linear loads
3. Resolution: 1 VA
viii. Power Factor³:
1. Range: 0 to 1
2. Accuracy: ±4% of range for linear loads
3. Resolution: 0.01
ix. Crest Factor:
1. Range: 0 to 3.5
2. Accuracy: ±5% of range
3. Resolution: 0.01
x. Phase:
1. Range: -359 to +359 degrees. Positive indicates time lag from reference
2. Accuracy: Within 100 microseconds of equivalent angle
3. Resolution: 1 degree C. Deliverables Delivery of the equipment shall be provided at the following address:
NAVAIR Calibration Standards (UIC N48535)
BLDG 612, Bay 9 MCAS
Beaufort, SC 29904
D. Delivery Date Delivery shall be no later than 4 weeks after award of contract, and shall be absolutely no later than December 16, 2016.
***PARTICAL SHIPMENT ACCECPTED***
E. Invoicing All invoicing shall be submitting into Wide Area Workflow for payment after ALL material is received in the condition as stated in section B.
VENDOR – IF ELECTRONIC EQUIPMENT, PLEASE MARK BOX/PACKAGE CONTAINING ELECTRONIC EQUIPMENT WITH THE “DO NOT SCAN” STICKER