Attachment 1
Salient Characteristics 5kV Pulsers
1) 5kV pulser unit
2) Arming and disarming the unit from the front panel. The unit may be manually fired from the front panel, or from a low voltage (+5V) rising edge trigger supplied to the front panel, or a high voltage (+25V to +185V) rising edge trigger supplied to the back panel.
Load testing can be done right up until immediately before firing to verify continuity across a bridge wire or initiator from the front panel. Voltage across the output can be monitored from a BNC connector or the front panel. The unit can also be safely and quickly disarmed from the front panel.
Operation of the unit is safely controlled with a switch closure interlock through a back panel connector. Power to the unit is switched with a keyed switch.
Firing Voltage: Adjustable up to Max KV (5kV)
Response Time: 850 nanoseconds
Jitter: + or – 15ns
Polarity: Positive, Ground Neutral
Trigger Input: +5 rising edge (front) +25V to +185V rising edge (rear)
Monitoring: Voltage monitoring from front BNC connector
Optional current monitoring from rear BNC connector
Power Input: 90-260 VAC 47-63Hz, 300 Watts