Salient_Characteristics_-_4_Port_Non-Linear_Network_Analyzer.pdf
PDF 100 KB Posted
- Attached to
- Notice of Intent to Award a Sole Source Contract
- Solicitation number
- N00173-19-Q-KN01
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Four-Port Non-Linear Network Analyzer Purchase Request No.: 57-4016-19 Salient Characteristics
Salient Characteristics for Four Port, Non-Linear Vector Network Analyzer (N-VNA)
The Naval Research Laboratory (NRL) requires the capability to analyze the behavior of non-linear radiofrequency (RF) receiver front-end components such as RF mixers, and Low-Noise amplifiers (LNAs) as well as diode-based metamaterials and other non-linear electronic components when they are subjected to both small and large-signal excitations. In order to directly import measurements into existing NRL software/equipment, small-signal microwave network analysis measurement outputs shall be in the form of S-parameters and large-signal, nonlinear measurement result outputs shall be equivalent to X-parameter measurement outputs. NRL currently uses the Advanced Design System (ADS) X-parameter circuit simulation software module manufactured by Keysight Technologies for measurement analysis and interactions with other laboratories. As such, the required non-linear analysis method is the X-parameter method as this method is proven to be fully compatible with existing NRL assets.
NRL requires a maximum frequency measurement capability of 67 GHz. Additionally, in order to measure 2-, 3-, and 4-port linear and non-linear microwave networks with pulsed waveform excitations, NRL requires that the non-linear vector network analyzer have four ports with pulse modulation capability on, at least, the first port. NRL requires the ability to perform noise figure measurements on both RF mixers, and LNAs. NRL requires that the four port N-VNA offering come with all of the necessary components for calibration including, but not limited to, an electronic calibration kit, comb generator for phase calibration, and power meter for amplitude calibration.
The following table provides the parameters and specifications for the required Four-Port Non-linear Vector Network Analyzer with Large Signal Non-linear Component Analysis Capability.
Item Parameter Specification 1 Large-Signal Non-Linear Component Characterization yes 2 Small-Signal Linear and Non-Linear Component Characterization yes 3 Maximum Frequency 67 GHz 4 Minimum Dynamic Range 129 dB 5 Maximum Output Power 13 dBm 6 Number of Ports 4 7 Time Domain Analysis up to 67 GHz yes 8 Noise Figure Measurements up to 50 GHz with vector correction yes 9 Internal pulsed RF measurements yes 10 Minimum pulse-width 20 ns 11 Minimum timing resolution for pulse profile measurements 10 ns 12 Pulse I/O adapter for connecting external pulse generators yes 13 50 GHz noise receiver yes 14 Flexible Test-port cables for all for ports yes 15 Thermocouple power sensor up to 67 GHz with appropriate adapters/connectors yes 16 Remotely programmable using MATLAB Equipment Control yes 17 Electronic calibration module up to 67 GHz with appropriate connectors/adapters yes 18 10 MHz to 67 GHz Comb Generator for phase calibration yes 19 Flexible test port cable set yes 20 10 MHz to 67 GHz Power Sensor yes 21 DC to 50 GHz 2.4 power divider with 2.4 mm connectors yes 22 Laboratory DC power supply, dual output 0-25 V, 0-1 A yes
Delivery of the N-VNA shall include a minimum of one (1) day of in-house training on the delivered equipment.