24Q0016_SOL_0001.pdf
PDF 520 KB Posted
- Attached to
- TEMPERATURE CHAMBER Federal contract opportunity
- Solicitation number
- N0016424Q0016
About this file
This document is an amendment to a solicitation for a temperature chamber. The amendment serves to correct an error in the salient characteristics and extend the solicitation response date by one day to October 24, 2023 at 12:00 PM EST.
The amendment updates the salient characteristics to clarify temperature range and change rate requirements, including a required minimum temperature range of 20C to 60C and target change rates of greater than 1C/min and achieving temperature limits within 30-60 minutes for an empty chamber. Preferred characteristics include programmable alarm settings, Ethernet network compatibility, and password protected security levels for the controller. The solicitation also outlines technical requirements for the temperature chamber such as interior sizes and weight limits, temperature tolerances, safety features, and documentation requirements. The related federal contract opportunity is for a temperature chamber issued by the Department of the Navy Naval Sea Systems Command under solicitation number N0016424Q0016.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 24Q0016_SOL.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
AMENDMENT OF SOLICITATION/MODIFICATION OF CONTRACT
Except as provided herein, all terms and conditions of the document referenced in Item 9A or 10A, as heretofore changed, remains unchanged and in full force and effect.
15A. NAME AND TITLE OF SIGNER (Type or print)
30-105-04EXCEPTION TO SF 30
APPROVED BY OIRM 11-84
STANDARD FORM 30 (Rev. 10-83) Prescribed by GSA
FAR (48 CFR) 53.243
This amendment serves to update the included Salient Characteristics to correct a clerical error. Additionally, this amendment w ill extend the solicitation by one day, updating to end on Tuesday, 24 OCT 2023 at 1200 EST.
1. CONTRACT ID CODE PAGE OF PAGES
J 1 4
16A. NAME AND TITLE OF CONTRACTING OFFICER (Type or print)
16C. DATE SIGNED
BY 17-Oct-2023
16B. UNITED STATES OF AMERICA15C. DATE SIGNED15B. CONTRACTOR/OFFEROR
(Signature of Contracting Officer)(Signature of person authorized to sign)
8. NAME AND ADDRESS OF CONTRACTOR (No., Street , County, State and Zip Code) X N0016424Q0016
X 9B. DATED (SEE ITEM 11)
16-Oct-2023
10B. DATED (SEE ITEM 13)
9A. AMENDMENT OF SOLICITATION NO.
11. THIS ITEM ONLY APPLIES TO AMENDMENTS OF SOLICITATIONS
X The above numbered solicitation is amended as set forth in Item 14. The hour and date specified for receipt of Offer X is extended, is not extended.
Offer must acknowledge receipt of this amendment prior to the hour and date specified in the solicitation or as amended by one of the following methods:
(a) By completing Items 8 and 15, and returning copies of the amendment; (b) By acknowledging receipt of this amendment on each copy of the offer submitted;
or (c) By separate letter or telegram which includes a reference to the solicitation and amendment numbers. FAILURE OF YOUR ACKNOWLEDGMENT TO BE RECEIVED AT THE PLACE DESIGNATED FOR THE RECEIPT OF OFFERS PRIOR TO THE HOUR AND DATE SPECIFIED MAY RESULT IN
REJECTION OF YOUR OFFER. If by virtue of this amendment you desire to change an offer already submitted, such change may be made by telegram or letter, provided each telegram or letter makes reference to the solicitation and this amendment, and is received prior to the opening hour and date specified.
12. ACCOUNTING AND APPROPRIATION DATA (If required)
13. THIS ITEM APPLIES ONLY TO MODIFICATIONS OF CONTRACTS/ORDERS.
IT MODIFIES THE CONTRACT/ORDER NO. AS DESCRIBED IN ITEM 14.
A. THIS CHANGE ORDER IS ISSUED PURSUANT TO: (Specify authority) THE CHANGES SET FORTH IN ITEM 14 ARE MADE IN THE
CONTRACT ORDER NO. IN ITEM 10A.
B. THE ABOVE NUMBERED CONTRACT/ORDER IS MODIFIED TO REFLECT THE ADMINISTRATIVE CHANGES (such as changes in paying office, appropriation date, etc.) SET FORTH IN ITEM 14, PURSUANT TO THE AUTHORITY OF FAR 43.103(B).
C. THIS SUPPLEMENTAL AGREEMENT IS ENTERED INTO PURSUANT TO AUTHORITY OF:
D. OTHER (Specify type of modification and authority)
E. IMPORTANT: Contractor is not, is required to sign this document and return copies to the issuing office.
14. DESCRIPTION OF AMENDMENT/MODIFICATION (Organized by UCF section headings, including solicitation/contract subject matter where feasible.)
10A. MOD. OF CONTRACT/ORDER NO.
2. AMENDMENT/MODIFICATION NO. 5. PROJECT NO.(If applicable)
6. ISSUED BY
3. EFFECTIVE DATE
17-Oct-2023
CODE
NAVAL SURFACE WARFARE CENTER CRANE DIV
B. TAYLOR (812-854-1024)
BLDG 3373
300 HWY 361
CRANE IN 47522-5001
N00164 7. ADMINISTERED BY (If other than item 6)
4. REQUISITION/PURCHASE REQ. NO.
CODE
See Item 6
FACILITY CODECODE
EMAIL:TEL:
N0016424Q0016
SECTION SF 30 BLOCK 14 CONTINUATION PAGE
SUMMARY OF CHANGES
SECTION SAP - SAP CONTINUATION SHEET
The required response date/time has changed from 23-Oct-2023 12:00 PM to 24-Oct-2023 12:00 PM.
The following have been modified:
SALIENT CHARACTERISTICS
SALIENT CHARACTERISTICS
NOTE: The “preferred” characteristics are not Government requirements, but are desired. If quoted they will not be evaluated for technical acceptance under technical evaluation factor number 1, Technical Approach. If quoted, the cost of the items will not be excluded from the overall cost of the total effort. Again, these are desired, but not required. There will be no penalty if not quoted.
⦁ Required: Interior Chamber Sizes: 38” Minimum Width X 38” Minimum Depth X 38” Minimum Height and 32 Cubic foot Minimum Volume.
⦁ Required: Temperature Range: ‐20°C to 60°C required. Chamber cannot under any conditions exceed 170°C.
⦁ Required: Temperature change rate greater than 1°C/min preferred.
⦁ Required: Target 20°C to ‐20°C or 20°C to 60°C in 30 minutes or less and 20°C to upper or lower limit in under 60 minutes with the chamber empty.
⦁ Required: Temperature Limited Sheathed Heaters (200°C) for spark suppression.
⦁ Required: Temperature tolerance after stabilization: ± 1°C required, ±0.5°C or better preferred.
⦁ Required: Touch Screen Programmable Controller with 2 USB connections minimum to allow for a mouse and keyboard connection for programming and operation. Must also able to run programs in excess of two years without restarting.
⦁ Required: The controller must have programmable alarm/shut off settings for maximum and minimum safety limits at both hot and cold temperatures with an emergency shut off.
⦁ Preferred: The controller must provide for a power outage recovery mode (resume current program either automatically or manually as instructed).
⦁ Preferred: The controller will have 5 digital input ports minimum to receive signals from other test equipment signaling it to execute the next programming step.
⦁ Preferred: The controller will have 5 digital output modes minimum to signal to other test equipment that a step was completed so the other system knows to complete its next step.
⦁ Preferred: The controller must be Ethernet Network compatible so it can be controlled and monitored by a PC using the chamber manufacturer’s software program.
⦁ Preferred: The controller should be able to monitor the system upon startup and operation for normal performance and provide a message or alarm when service is required and help direct maintenance and repairs to a particular system if not the major component.
⦁ Preferred: The controller shall allow for two level password system security levels. A basic password that allows users to start/stop chamber, set temperature and humidity points manually, and start/stop profiles; and a maintenance/programming level password that allows for the programming and any diagnostic functions along with the basic operation.
⦁ Required: The system will have redundant and separate temperature Hi & Low Temperature limit controls that will sound an alarm and shut down the chamber if the controller fails. This can be manually set on the machine or incorporated as additional limits in the controller.
⦁ Required: 1 Shelf included that can be located in multiple positions supporting 200 lbs. min.
⦁ Required: Floor of the chamber to support 1000 lbs min.
⦁ 208V – 230V 3 Phase. Should not be phase dependent and if it is, it will not be damaged and provide feedback indicating power phase error and have a quick access and wiring correction feature.
⦁ Required: Air Cooled Condenser – No water connection allowed. Must be electrical hook up and operate only.
⦁ Required: Fresh Air Purge system to remove any excess hydrogen gas buildup in the chamber.
⦁ Required: Calibration procedures, Calibration Equipment List, and Calibration specialty items included.
⦁ Required: Chamber to be used for Battery testing and designed to meet EUCAR‐ 3 battery events.
⦁ Preferred: Heaters and cooling to be separated from test space to prevent gas ignition.
⦁ Required: Explosive pressure relief port for equipment safety that operates at all temperatures.
⦁ Required: Spark prevention on internal wiring as much as possible to avoid gas ignition (understood cannot be 100% protected).
⦁ Required: System must use latest Government regulations for Refrigerant and forbidden materials per environmental regulations. Guidance and reference websites from NSWC Environmental and safety groups are provided here:
⦁ As the EPA is considering the Global Warming Potential (GWP) in SNAP determinations regarding acceptability of refrigerants, I strongly recommend selecting a refrigerant with the lowest possible GWP (and of course no Ozone Depleting Potential (ODP)) from the SNAP lists for the humidifier chamber.
https://www.epa.gov/snap/substitutes‐industrial‐process‐refrigeration https://www.epa.gov/snap/substitutes‐non‐mechanical‐heat‐transfer‐systems https://www.epa.gov/snap/substitutes‐industrial‐process‐air‐conditioning https://www.epa.gov/sites/production/files/2015‐ 08/documents/snap_regulatory_factsheet_july20_2015.pdf https://www.epa.gov/snap/reducing‐hydrofluorocarbon‐hfc‐use‐and‐emissions‐federal‐sector‐through‐snap#actions https://www.epa.gov/snap/compositions‐refrigerant‐blends https://www.epa.gov/snap/substitutes‐very‐low‐temperature‐refrigeration
⦁ Required: Quantity 2 (1 on each side) – 4” ports with solid plugs and plugs with openings to allow wire connections into the chamber.
⦁ Required: The chamber shall be provided with both self‐leveling legs / castors.
⦁ Required: On wheels / casters that have locks or lifts (on at least 2 wheels) for mobility and security. Wheels to be sized to allow for thresholds or ramps with a 0.5” bump or to cover a 1.5” gap in surfaces without getting stuck or getting damaged (momentum should allow it to roll over).
⦁ Required: Chamber not to exceed 65” width by 90” depth by 105” height in transportation configuration.
⦁ Preferred: Easy removal of accessories with wingnuts or screws into threaded nuts (not through sheet metal that can be easily stripped) and electrical connectors or quick connectors by NSWC Crane technicians such as Water tanks, Exhaust blowers, etc is allowed to meet the transportation configuration size limits.
⦁ Required: Interior to be able to be lit with door open or closed in a manner consistent with the battery testing and possible gas build up.
⦁ Required: Fog Free Viewing Window.
⦁ Preferred: The chamber compressors shall be equipped with liquid injection thermostat to aid in maintaining the life of the compressor.
⦁ Preferred: The chamber shall have refrigeration service taps and liquid filled refrigeration gauges.
⦁ Required: The chamber shall have removable panels that permit easy access to all systems for maintenance purposes.
⦁ Required: Document Warranty terms, duration, and conditions.