MR CMP22 Data Sheet 145034-25-0021.pdf

PDF 239 KB Posted

Attached to
GML OTT Hydromet d. b. a. Kipp & Zonen Pyranometers & Accessories
Solicitation number
145034250021IV
Issued by

About this file

This document is a technical data sheet for Kipp & Zonen pyranometers (CMP21, CMP22, and SMP22) used for scientific solar radiation measurement. The pyranometers are ISO 9060:2018 Spectrally Flat Class A certified, offering highest level accuracy across a broad spectral range with analog and digital output options. Key specifications include sensitivity of 7 to 14 µV/W/m², operating temperature range of -40°C to +80°C, IP67 rating, and a 5-year warranty.

The related federal contract opportunity is a Sources Sought Synopsis by NOAA's Global Monitoring Laboratory (GML) seeking these specific pyranometers to expand their Surface Radiation Budget Network (SURFRAD) and portable radiation systems. The agency requires CMP21 and CMP22 models to measure total and diffuse solar radiation components, with delivery requested by July 1, 2025. The synopsis targets U.S. domestic businesses, including small businesses, HUBZone, 8(a), Service Disabled Veteran Owned, and Women Owned Small Business firms, with a NAICS code of 334513 and a size standard of 750 employees.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

ISO 9060 Spectrally Flat Class A Highest level accuracy Broadest spectral range Analog and digital outputs 5 years warranty

Applications

CMP21 | CMP22 | SMP22

ISO 9060 & IEC 61724 Class A Fully compliant with ISO 9060:2018 spectrally flat Class A specifications, the CMP21, CMP22, and SMP22 provide solar irradiance measurements with highest accuracy and reliability. They are the sensors of choice for renowned research institutes around the world. Adding highest-quality quartz domes for a wider spectral range makes the CMP22 and the SMP22 the most accurate pyranometers available.

Minimized maintenance The CMP21 and CMP22 as well as the smart SMP22 are extremely robust pyranometers that have been designed to fulfill the highest requirements of scientific research.

The CMP21 and CMP22 have an easy-to-remove drying cartridge filled with easy-to-replace desiccant that is supplied in convenient refill packets. In contrast, the SMP is fitted with a maintenance-free internal desiccant that lasts at least 10 years.

Analog or digital outputs The CMP21 and CMP22 do not require any power. Incoming solar radiation generates a continuous millivolt output, which is converted in a data logger to irradiance in W/m² using the calibrated sensitivity. For easy integration into SCADA systems the SMP22 has Modbus® RTU RS-485 serial communication, plus an amplified analog output. A specific, individual polynomial correction function is stored in each SMP22.

5 years warranty All pyranometers from Kipp & Zonen come with a 5-year warranty and we have service and calibration centers around the world.

Scientific Research Baseline Surface Radiation Network

Pyranometer Premium pyranometers for highest requirements an OTT HydroMet brand

OTT HydroMet GmbH | euinfo@otthydromet.com | www.otthydromet.com

Technical Specifications Classification to ISO 9060:2018

Sensitivity

Impedance

Expected output range (0 to 1500 W/m²)

Maximum operational irradiance

Analogue output • V-version

Analogue output range • V-version*

Analogue output • A-version

Analogue output range • A-version*

Serial output

Serial output range

Response time (63 %)

Response time (95 %)

Spectral range (20 % points)

Spectral range (50 % points)

Zero offsets (unventilated)

(a) thermal radiation (at 200 W/m²)

(b) temperature change (5 K/h)

(c) total zero offset

Non-stability (change/year)

Non-linearity (100 to 1000 W/m²)

Directional response

(up to 80° with 1000 W/m² beam)

Spectral selectivity (350 to 1500 nm)

Tilt response (0° to 180° at 1000 W/m²)

Temperature response

Field of view

Accuracy of bubble level

Power consumption (at 12 VDC)

Supply voltage

Software, Windows™

Detector type

Operating temperature range

Storage temperature range

Humidity range

MTBF (Mean Time Between Failures)

Ingress Protection (IP) rating

Recommended applications

Spectrally Flat Class A

7 to 14 µV/W/m²

10 to 100 Ω

0 to 21 mV

4000 W/m²

< 1.66 s

< 5 s

270 to 3000 nm

285 to 2800 nm

< ±7 W/m²

< ±2 W/m²

< ±9 W/m²

< ±0.5%

< ±0.2%

< ±10 W/m²

< ±3%

< ±0.2%

< ±1% (-20 to +50°C)

180°

±0.1°

Thermopile

-40 °C to +80 °C

-40 °C to +80 °C

0 to 100%

> 15 years

IP67

CMP21 CMP22

Diameter x height

Diffusor height

Cable length

150 x 92.5 mm

68 mm

10, 25, or 50 m

CMP21

150 x 92.5 mm

68 mm

10, 25, or 50 m

CMP22Dimensions

* adjustable with SmartExplorer Software | Note: The performance specifications quoted are worst-case and/or maximum values

SMP22

Spectrally Flat Class A

4000 W/m²

0 to 1 V

-200 to 2000 W/m²

4 to 20 mA

0 to 1600 W/m²

RS-485 Modbus® RTU

-400 to 4000 W/m²

< 0.7 s

< 2 s

210 to 3600 nm

350 to 3500 nm

< ±3 W/m²

< ±1 W/m²

< ±4 W/m²

< ±0,5%

< ±0.2%

< ±5 W/m²

< ±3%

< ±0.2%

< ±0.3% (-20 to +70°C)

< ±0.3% (-40 to +70°C)

180°

±0.1°

V-version: 55 mW

A-version: 100 mW

5 to 30 VDC

SmartExplorer Software, for configuration, test and data logging

Thermopile

-40 °C to +70 °C

-40 °C to +80 °C

0 to 100%

> 10 years

IP67

Spectrally Flat Class A

7 to 14 µV/W/m²

10 to 100 Ω

0 to 21 mV

4000 W/m²

< 1.66 s

< 5 s

210 to 3600 nm

210 to 3600 nm

< ±3 W/m²

< ±1 W/m²

< ±4 W/m²

< ±0.5%

< ±0.2%

< ±5 W/m²

< ±3%

< ±0.2%

< ±0.5% (-20 to +50°C)

180°

±0.1°

Thermopile

-40 °C to +80 °C

-40 °C to +80 °C

0 to 100%

> 15 years

IP67

Scientific research requiring the highest level of measurement accuracy and reliability under all conditions.

150 x 92.5 mm

68 mm

10, 25, or 50 m

SMP22

M E-

KZ

-R

E-ds

-C M

P_

SM

P2 2-

EN

-2 an OTT HydroMet brand

File details come from the government source that posted it. Updated .