SCOPE OF WORK
Mobile Toshiba Aquilion 64, or Equivilant CT Unit Lawton Indian Hospital Lawton, Oklahoma June 3, 2021
A. Overview of Project
1. The intent and objective of this Scope of Work is to lease a mobile Toshiba Aquilion 64, or equivalent CT Unit. The Lawton Service Unit is in the process of purchasing a new 128 slice CT System. The new system is scheduled to be onsite on Aug 29, 2021 with anticipated removal 10/15/2021. The existing system will be scheduled to be decommissioned and removed from the existing building before the new system can be installed. To ensure continuous patient CT services the Lawton Service Unit will lease a mobile CT system. The mobile CT system will need to be fully functional before the decommissioning and removal of existing CT system.
B. Services to be Performed by the Contractor
1. The contractor will lease to the Lawton Indian Hospital a mobile Toshiba Aquilion 64, or equivalent whole-body Computed Tomography (CT) system.
2. Contractor will provide one site visit and provide recommended site locations for the mobile CT unit.
3. The leased mobile unit will consist of the following items:
i. Mobile CT unit will be housed in a mobile trailer with minimum dimensions of 48’L x 8’W
ii. Toshiba Aquilion 64, or equivalent whole-body Computed Tomography (CT) system
iii. MedRad Stellant Dual Power Injector
iv. Vital Vitrea 3D Workstation
v. Bone Density, QCT
4. Contractor will provide the following setup and installation services:
i. Deliver CT mobile unit
ii. Complete setup of CT mobile unit
1. Setting CT mobile unit trailer including leveling and anchoring
2. Connect utility service to the Lawton Indian Hospital’s source
3. Connect phone and data to the Lawton Indian Hospital network
4. Handicap accessible entrance
5. Startup and testing of Mobile CT unit
5. System Requirements
i. The system generates 64 slices per second using the selectable slice-thickness multirow detector (SSMD).
ii. XR-29 Compliant
iii. uncompromised image quality
iv. Image data transfer link
v. High-speed scanning
vi. Long scanning range
vii. Excellent image quality
viii. Advanced image reconstruction
ix. Tilt helical scanning
x. Large-aperture, slip-ring gantry and extra-wide couch
xi. Ergonomic operator console
xii. X-ray generator kV range: 80, 100, 120 and 135 kVp
xiii. X-ray tube heat capacity 4.0 MHU
xiv. Maximum Scan Ranger (cm): 1800 mm cm
xv. Minimum Rotation Speed: 0.4
xvi. Gantry Diameter: 72 cm
xvii. Patient Couch: 180cm long by 47cm wide, 660lb weight limit, metal free
xviii. Volume image processor
xix. V4 software with workflow enhancements, on console Easy 3-D, advanced Quantum de-noising adaptive noise reduction filter and easy to use interface
xx. SUREExposure-Dose modulation based on scanogram
xxi. SUREStart-Real-time contrast detections 12 fps., timing bolus not needed
xxii. SUREScan-Real-time helical scan display
xxiii. Five scan fields of view: 18, 24, 32, 40 and 50 cm
xxiv. Patient positioning lights
xxv. Scan times: 0.4, 0.5, 0.75, 1, 1.5, 2 and seconds full rotation
xxvi. Remote control of couch-top movement
xxvii. Gantry tilt control xxviii. Three preset window can be stored in the eXam plans
xxix. Displays images in 512x512 or 1024x1024
xxx. CT number display ranges from -10536 to +8191
xxxi. High-capacity hard disk
xxxii. Mobile CT Power Plug must plug into a RUSSELL STOLL # DF2504FRAB wall mounted Receptacle
6. Contractor will provide the following support
i. On-line help
ii. Technical support
iii. Customized Equipment to support our clinical needs
iv. Applications Training On-Site x 3 Days beginning 08/30-09/01/2021 (10 days if not Toshiba Aquilion 64 beginning 08/30-09/10/2021)
v. Customer Support 24/7
vi. Site Assistance, Pre-planning and Set-up
vii. Transportation, On-Time Delivery
viii. Systems meet OEM Specifications or greater
ix. Equipment and Trailer Service Maintenance Agreement
x. Current Physicist Surveys provided (within 12 months of set up, not to expire during contract)
xi. All Trailers are licensed through DOH
xii. Equipment is fully insured to meet our facility requirement
xiii. The Joint Commission (TJC) Accreditation
C. Final Inspection:
Final Inspection – completion date is the date established in the contract documents when the project is to be fully complete and is ready to turn over to the Government. All deliverables shall be complete and submitted to the Government at that time.
D. Points of Contact:
Contractor Officer:
Judy Perkins, Chief Contracting Officer 701 Market Drive Oklahoma City, OK 73114 judy.perkins@ihs.gov
Contracting Officer Technical Representative:
Laquita Cormany, Radiology Supervisor Lawton Indian Hospital 1515 Lawrie Tatum Road Lawton, Oklahoma 73507 laquita.cormany@ihs.gov image1.png image2.png
File details come from the government source that posted it. Updated .