This document is a specifications sheet for a high throughput automated microscope system needed by the Naval Research Laboratory (NRL). The system is comprised of six components: a robotic microscope platform, microscope objectives, light sources and optical filters, cameras, an imaging workstation, and image acquisition and analysis software.
The key requirements for the microscope system include: adaptive phase gradient contrast optics for interference contrast imagery, water autoimmersion objectives of at least 50X magnification and 1.2 numerical aperture, an adaptive lens guard to prevent collisions with sample plates, and an automated UV disinfection system. The system must be compatible with a typical laboratory environment. This microscope system is required to support the NRL's biomanufacturing research efforts funded by the Office of the Undersecretary of Defense.
The related federal contract opportunity is a sole source notice for the ZEISS Celldiscoverer 7 optical microscope, which the NRL has determined can meet the unique requirements outlined in the specifications. The NRL intends to award a firm-fixed-price contract to Carl Zeiss Microscopy, LLC for this microscope system. Any firms that believe they can meet the requirements may notify the Contracting Officer within 7 days, but the government retains sole discretion on whether to compete the procurement.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
63-9022-24/1301173156 SPECIFICATIONS FOR HIGH THROUGHPUT AUTOMATED MICROSCOPE
I. SPECIFICATIONS FOR A HIGH THROUGHPUT AUTOMATED MICROSCOPE
II. OVERVIEW: The NRL has a critical need for a high throughput light microscope for automated live cell microscopy workflows capable of automatically supplying and removing immersion medium for complex automated workflows.
III. BACKGROUND: The high throughput automated microscope capable of complex extended workflows is a critical tool for biomanufacturing quality control and phenotyping discovery in live cell microscopy. The current capabilities in Section 6392 consist of routine microscopes that are largely manual in their operation and therefore not appropriate for high throughput automated live cell screens. Additionally, these capabilities lack the capability of adaptive phase gradient contrast for interference contrast imagery and lack the ability to incorporate a water auto-immersion objective of at least 50X magnification and 1.2 numerical aperture. Having a high throughput light microscope in the section will allow us to meet the critical milestones of our biomanufacturing research efforts currently funded externally by the Office of the Undersecretary of Defense (OUSD).
IV. SCOPE: The high throughput automated microscope system is comprised of six individual components: (a) a robotic microscope platform capable of automatically loading samples and changing optics, (b) microscope objectives, (c) light sources and optical filters, (d) cameras, (e) an imaging workstation and, (f) Image acquisition and analysis software. The system shall provide automated live cell imagery on high throughput well plates for periods of hours to days while ensuring optimal cell health as required by the currently OUSD funded biomanufacturing program entitled: “Rapid, In-Line Biomaterial Quality Control Based on Machine Vision & High-Dimensional Analytics” V. SPECIFICATIONS AND SALIENT CHARACTERISTICS FOR HIGH THROUGHPUT AUTOMATED MICROSCOPE The high throughput automated microscope, hereinafter referred to as “The System” shall be comprised of six (6) individual components: (a) a robotic microscope platform capable of automatically loading samples and changing optics, (b) microscope objectives, (c) light sources and optical filters, (d) cameras, (e) an imaging workstation and, (f) Image acquisition and analysis software. The system SHALL meet all of the following minimum technical, physical, and performance requirements:
A. ADAPTIVE PHASE GRADIENT CONTRAST: Adaptive phase gradient contrast optics for interference contrast imagery compatible with plastic or glass substrates and independent of supernatant nutrient medium B. WATER AUTOIMMERSION OBJECTIVES: One or more water autoimmersion objectives, of at least 50X magnification and 1.2 numerical aperture, in which the immersion medium is automatically supplied and removed for complex automated workflows that extend for hours to days.
C. ADAPTIVE LENS GUARD: An adaptive lens guard for avoiding carrier plate collisions for preventing collisions with carrier plate as determined from each inserted plate.
D. UV DISINFECTION: Automated UV disinfection system to prevent cross contamination of samples.
E. The overall system and all components shall be compatible with a typical instrument laboratory setting in terms of power and utility needs, and safety considerations.