MECs - FORMULATRIX F.A.S.T. Liquid Handler.docx
DOCX document 18 KB Posted
- Attached to
- FORMULATRIX F. A. S. T. Liquid Handler
- Solicitation number
- HT9426-26-RFI-3175
- Issued by
- Not on record
About this file
This document establishes the Minimum Essential Characteristics (MECs) for a FORMULATRIX Flow Axial Seal Tip (F.A.S.T.) Liquid Handler procurement. The equipment must meet specific physical and operational parameters: dimensions not exceeding 532mm (W) x 403mm (H) x 596mm (D), weight under 18kg, and a compact desktop footprint compatible with larger automated systems integration. The liquid handler shall perform rapid and efficient liquid transfers across 96-well and 384-well plate formats with dispensing volumes ranging from 0.1µL to 13µL (low volume) and 0.5µL to 50µL (high volume), achieving 5% coefficient of variation precision at 0.1µL. The system must include tip position sensors, automatic tip arrangement, collision detection, positive displacement disposable tips, and contact dispensing capabilities with wide viscosity range support.
The F.A.S.T. Liquid Handler shall feature an open API for third-party system integration, interchangeable decks for flexible labware configuration, semi or fully automated labware definitions and calibration, and automated audit trails. System control requires a Trusted Platform Module (TPM) 2.0 compatible Debian Linux onboard computer with web browser-based operational controls (Google Chrome 91 or higher compatible). The equipment must support Integrated Fluidic Circuits (IFCs) 192.24, 96.96, and 48.84 and serve as a validated solution for Olink IFCs and Proximity Extension Assay (PEA) workflows including Target 48/96, Explore 3072, and Explore-High Throughput kits, supporting both Olink pre- and post-workflows. The procurement includes Olink Protocol bundles, Installation Qualification (IQ), Operational Qualification (OQ), and new install training. Delivery is required to Robert Grant Ave, BLDG 503-Ground West 47, Silver Spring, MD 20910-7500. This RFI/Sources Sought (HT9426-26-RFI-3175) represents a pre-solicitation notification requesting industry feedback.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| A.02 Request for Information FAST 2MAR26.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Minimum Essential Characteristics (MECs) FORMULATRIX Flow Axial Seal Tip (F.A.S.T.) Liquid Handler
· Shall be no more than 532mm (W) x 403mm (H) x 596mm (D) in dimensions.
· Shall not weight over 18kg.
· Shall have a small footprint suitable for desktop use but is also open for integration into larger automated systems.
· Allows for rapid and efficient transfer of liquids in 96-well or 384-well plates, and other formats.
· Capable of dispensing volumes from 0.1 µL to 13 µL with high precision and accuracy, minimizing reagent usage. Transfer precision of 5% coefficient of variation (CV) at 0.1 µL.
· Shall offer both low volume (0.1-13 µL) and high volume (0.5-50 µL) tips, compatible with various plate types on a single head unit.
· Shall have multiple transfer modes allowing to support tasks like simple transfers, plate copies, cherry-picking, serial dilutions, and more.
· Shall include tip position sensors, automatic tip arrangement, and collision detection for reliable and accurate transfers.
· Shall offer an open Application Programming Interface (API) for easy integration with other laboratory automation systems.
· Shall utilize positive displacement disposable tips to ensure the integrity of samples without any risk of cross-contamination.
· Shall provide contact dispensing and have wide viscosity range.
· Shall allow for interchangeable decks allow for set up of tips, plates, and reagents prior to working on the instrument.
· Shall allow for semi or fully automated labware definitions and calibration.
· Shall allow for automated audit trails.
· Shall include Trusted Platform Module (TPM) 2.0 compatible Debian Linux on board computer.
· Shall allow for web browser-based operational controls such as Google Chrome version 91 or higher or other compatible internet browser interface.
· Shall be compatible with Integrated Fluidic Circuits (IFCs) 192.24, 96.96, 48.84.
· Shall be a validated solution for Olink IFCs and Proximity Extension Assay (PEA) Target and Explore workflows including Target 48 and 96, Explore 3072, and Explore-High Throughput (HT) kits.
· Shall allow for both Olink pre and post workflows.
· Shall include Olink Protocol bundles, Installation Qualification (IQ), Operational Qualification (OQ), and new install training.
· Shall be shipped and delivered to Robert Grant Ave, BLDG 503-Ground West 47, Silver Spring, MD 20910-7500