This document is a Minimum Essential Characteristics (MEC) specification for an air-liquid interface (ALI) exposure system acquisition by the U.S. Army Medical Research Institute of Chemical Defense (USAMRICD). The government requires a single air-liquid interface exposure chamber with associated equipment and support services, including installation, training, and one year of warranty. The system must provide controlled environmental conditions that replicate mammalian respiratory tract exposure by delivering airborne test articles (gases, vapors, aerosols, or dry particles) to cell and tissue models while maintaining physiological temperature and humidity levels. Critical functional requirements include a heated, airtight mini-incubator with vacuum operation; an electrostatic deposition system to maximize particle deposition; multi-well plate compatibility supporting 6, 12, or 24-well transwell formats for 6 to 24 simultaneous replicate samples; and real-time monitoring capabilities for airborne test article concentrations and airflow parameters. The system must integrate easily with upstream aerosol generation systems and downstream cleanup filters, operate within a standard 48" x 36" x 59" chemical fume hood footprint, and provide high-quality, reproducible data comparable to published scientific literature.

The specification also requires quantity one each of an airflow humidifier (delivering in-line humidification above 75%), an airflow dryer (preventing condensation downstream of the chamber), and an internal battery-operated vacuum pump with small footprint for situations where house-provided vacuum is inadequate. The contract was awarded to MEDTEC BIOLAB, INC. for their CelTox Sampler system, identified as the only ALI exposure system capable of meeting the stated minimal essential characteristics. The award was issued on July 16, 2026, under solicitation number HT942626PE028 for $82,500.00.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Minimum Essential Characteristics Air-liquid interface exposure system, installation, training, and warranty with shipping for U.S. Army Medical Research Institute of Chemical Defense (USAMRICD)

Quantity 1 = Air-liquid Interface Exposure Chamber

· Air-liquid interface (ALI) exposure

· Provides control over airborne test articles

· Holds and maintains cells and tissue models at physiological ALI conditions consistent with human respiratory tract or dermis

· More closely replicates mammalian lung inhalation exposure conditions by delivering airborne gas, vapor, aerosols, or dry particle test articles than traditional submerged culture models

· Exposed ALI models can be easily moved to incubators post-exposure via single insert or entire exposed plate

· Provides high quality and repeatable data comparable to published scientific literature

· Associated mini-incubator housing for controlled environment

· Heated enclosure to adjust, maintain, and record temperature and humidity close to human physiological condition

· Airtight seal to maintain physiological relevant humidity

· Operates under a vacuum to reduce potential exposure to researchers while maintaining emission source

· Internally facing components are made of material to minimize reactivity, damage to system components, and loss of toxicant to internal wall surface and make cleaning easy

· All air flow system components and sensors are housed within the system for stable reproducibility of mammalian physiological conditions.

· Modular and versatile design to evaluate different toxicant phases, experimental designs, and in vitro model formats

· All-in-One system to evaluate any type of airborne toxicant or drug delivery regardless of its physical form (gas vs. vapor vs. mixed aerosol vs particle).

· User-friendly design for someone with minimal to moderate experience with using ALI systems to conduct airborne toxicant exposures

· Modules are highly compatible and allow easy switch out capabilities

· Flexible design to handle user-defined multi-well ALI model formats

· Allows acquisition of one system under a single funded equipment ceiling budget

· Electrostatic deposition system

· Uses electrostatic unipolar charging source to push particles onto ALI culture models to maximize particle deposition and reduce the use of costly particle samples

· Can be turned off and on, or completely removed from the chamber depending on experimental goals

· Multi-well plate format

· Allows for simultaneous exposures to the same test atmosphere, for example:

· Comparison of with/without electrostatic deposition

· Comparison of two different ALI model systems

· Assess mixed aerosols (particles + gas) by allowing particles to deposit on one plate and not the other

· Dynamic exposure platform capability

· Options to maximize deposition of either gases, vapors, aerosols or particulate

· Ability to hold single “transwell” inserts or entire cell culture plate in 6, 12, or 24-well “Transwell” formats in exposure system

· Throughput design allowing between 6 to 24 replicate samples per exposure run

· Easy integration with other generation and monitoring systems

· Systems that provide liquid aerosols, dry particle, gas, vapors, and mixed aerosols can be incorporated into the system

· Real-time monitoring systems can be deployed and up- and down-stream of ALI exposure system to monitor and quantify airborne test article concentrations and air flow parameters

· Cleanup and filters can easily be incorporated into the downstream airflow to remove toxicants before discharge into laboratory exhaust

· Small physical footprint in lab

· Generation, ALI exposure system, and vacuum pump must fit within a standard (48” wide x 36” deep x 59” high) chemical fume hood

· Low maintenance cost

· Components are self-maintained by user

· Minimal maintenance services for only repairs required by company field technicians Quantity 1 = Airflow Humidifier

· Humidification of test atmospheres

· In-line humidification of test atmosphere above 75% humidity

· Easily charged and reused

· Customizable for other fittings for different tube sizes and integration with several different upstream aerosol generation systems and downstream ALI exposure designs, modules, and chambers Quantity 1 = Airflow Dryer

· Drying of test atmosphere

· Placed downstream of ALI exposure chamber to dry test atmospheres before exhausting to waste bottles or chemical fume hood

· Prevents condensation and blockage of air flow downstream of ALI exposure chambers Quantity 1 = Vacuum pump

· Provides an internal, battery-operated vacuum pump when house provided vacuum is inadequate to maintain low air flow conditions

· Small footprint and sits inside chemical fume hood

· Alternative vacuum source other than house provided

· Allows slower and controlled airflows (<0.5 L/min), especially for nozzle-based systems that require push-pull airflows Quantity 1 = Digital In-line Temperature/ Relative Humidity Sensor

· Monitor and record both temperature and humidity real-time prior to and during experimental exposure runs

· data logging capability with a user provided computer Quantity 1 = Mass flow controller

· Controls air flow upstream of chamber to create a steady and consistent airflow to deliver test atmospheres to exposure chamber

· Visual digital readout to control and monitor airflow during exposures Quantity 1 = 24 nozzle module to accommodate 24-well insert plate design

· Modular

· Compatible with temperature, humidity, and air flow control systems and read outs

· Effective gas and vapor delivery for exposures to 24-well inserts

· Push-pull vacuum control over test atmospheres

· Upstream airflow is ~ 1 L/ min while downstream airflow is < 0.5 L /min

· Each transwell insert receives its own nozzle to ensure and maximize exposure

· Control multiple doses in same experimental run

· Air flow to the module can be divided in halves or thirds to conduct clean air plus one or two target doses of test atmosphere

· Capture of toxic gases and fumes

· Provides more efficient capture of toxic gases and vapors compared to static and passive aerosol delivery systems thereby minimizing accidental exposure to internal surfaces and researchers Quantity 1 = Quartz Crystal Sensor / Microbalance

· Real-time mass quantification of aerosol and particle deposition

· Capable of fitting within exposure chamber, insert holder, or test plate

· Easy placement within chamber

· Interfaces with established software program running on a standard laptop computer

· Reusable if appropriately cleaned and maintained Quantity 1 = Software for microbalance sensor

· Evaluation of real-time mass deposition

· Ensures dry particle or aerosol exposures are occurring to ALI tissue models during exposure runs

· Helps to identify airflow anomalies or bad exposure runs prior to long downstream sample processes

· Compatible with sensor/microbalance

· Receives and logs data in acceptable format, such as text file, comma delimited (.csv), or tab delimited (.tsv)

· Export capability to other data management programs such as Microsoft Excel Freight and shipping to USAMRICD Installation and training to be completed at USAMRICD by 30 September 2026 OPTIONAL 1 year warranty