Liquid Nitrogen Generator SOW.docx

DOCX document 15 KB Posted

Attached to
6680--36C25924Q0159| Nitrogen Generator
Solicitation number
36C25924Q0159
Issued by

About this file

This technical specification outlines requirements for a standalone liquid nitrogen generator to support a mass spectrometer instrument. The generator must be capable of producing up to 35 L/min of analytical-grade nitrogen gas at a maximum pressure of 116 psi, with outlet fittings and redundant moisture removal systems. It will operate continuously to support research and analysis at the Rocky Mountain Regional VA Medical Center in Aurora, Colorado. Responses are due by January 11, 2024 at 4:00 PM for this solicitation issued by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 19. The specification provides detailed performance and design requirements for purifying, compressing, and delivering nitrogen gas to analytical instrumentation.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Liquid Nitrogen Generator Technical Specifications

Standalone nitrogen generator capable of producing analytical grade N2 to support mass spectrometer instrument.

A) Max Flow Rate: up to 35 L/min B) Max Pressure: 116 psi (8 bar) C) Gas Outlets 1 x ¼” BSPP 1 D) Internal Compressors: 2

a. Capable of running 24/7 E) Purity: 95-99.5%

a. Factory certified purity with respect to O2 content. Actual purity varies in relation to flow (min. 95% at maximum flow output) F) Max Relative Humidity: 80% RH @ 31°C G) Redundant moisture removal technology H) Non-methane hydrocarbon filtration removes ambient long-chain hydrocarbons to <1ppm I) Max Altitude: 3000m J) Particles: < 0.01 µm K) Phthalate & BHT Free L) Suspended Liquids: None M) Non-Methane Hydrocarbon Content: < 1 ppm NMHC N) Operating Temperature: 5°C (41°F) to 35°C (95°F) O) Electrical Requirements: 120VAC, 60Hz, 12A P) Power Consumption: 960 VA Q) Heat Output: 4118 BTU R) Noise Level: 56 dB S) Generator Dimensions (H x W x D)

a. 640 x 570 x 710 mm

b. 25.2 x 22.5 x 28’’ T) Generator Weight: 92 kg / 202 lbs U) High-resolution interface with integrated system control, performance monitoring, start-up status check, servicing, and system diagnostics