About this file
This is an Item Description document from the Department of the Air Force, Oklahoma City Air Logistics Complex (OC-ALC), Tinker Air Force Base, for procurement M26-016. The 76 MXSG is requesting one (1) Bruker Invenio Fourier Transform Infrared Spectrometer (FTIR) with a 20-meter fixed path-length gas cell optimized for quantitative gaseous analysis, and one (1) Bruker Hyperion II FT-IR Microscope for advanced materials characterization. The FTIR must include a benchtop system with integrated laser alignment capability (±0.1 m accuracy), high-sensitivity DTGS detector, spectral resolution of 0.4 cm⁻¹ or better, KBr beamsplitter covering 8,000–300 cm⁻¹ (with optional extended range 11,000–350 cm⁻¹), wavenumber accuracy of 0.01 cm⁻¹ or better at 2000 cm⁻¹, and a permanently aligned interferometer. The fixed path-length gas cell must have an aluminum body with corrosion-resistant coating, operate up to 3 ATM with calibrated pressure gauge, easily accessible mirrors, and heating capability to ≥100°C.
Both instruments require comprehensive user manuals (electronic PDF and printed), latest manufacturer's analytical software with full instrument control, automatic water vapor and CO₂ compensation, robust quantitative analysis methods (CLS, PLS algorithms), standard spectral processing functions, Microsoft Windows 11 compatibility, and Microsoft Office Small Business Edition or equivalent. Each delivery includes a dedicated computer and monitor meeting manufacturer's specifications, all unique operation supplies plus 60-day consumables (desiccant, cleaning materials), and special tools for routine maintenance and alignment. Both instruments require one day of on-site setup and one day of on-site training (minimum 8 hours) by a qualified service engineer with performance verification testing, plus an additional 8 hours of training for Air Force chemists covering basic operation, custom quantitative method development, and programming, with documentation of installation and verification procedures provided.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Contract - FA812626P0016 (Body).pdf | ||
| Fully Signed Front Page FA912626P0016.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
DEPARTMENT OF THE AIR FORCE
OKLAHOMA CITY AIR LOGISTICS COMPLEX (OC-ALC)
TINKER AIR FORCE BASE OKLAHOMA
Item Description
Bruker Invenio Fourier Transform Infrared Spectrometer (FTIR) and Bruker Hyperion II FT-IR Microscope
M26-016
The 76 MXSG is requesting Bruker Invenio Fourier Transform Infrared Spectrometer (FTIR) and Bruker Hyperion II FT-IR Microscope. This requirement must have the following specifications:
Bruker Invenio Fourier Transform Infrared Spectrometer (FTIR) – Quantity: 1
• Benchtop FT-IR system with 20-meter fixed path-length gas cell optimized for quantitative gaseous analysis.
• Integrated laser system for precise gas cell alignment and path length verification (±0.1 m accuracy).
• High-sensitivity Deuterated Triglycine Sulfate (DTGS) detector.
• Spectral resolution of 0.4 cm⁻¹ or better; specify signal-to-noise ratio per ASTM E1421.
• Mid-infrared KBr beamsplitter covering 8,000–300 cm⁻¹ standard range; optional extended range
11,000–350 cm⁻¹ available at additional cost.
• Permanently aligned interferometer with documented alignment stability and lifespan.
• Electronically stabilized infrared source with specified stability and lifetime.
• Wavenumber accuracy of 0.01 cm⁻¹ or better at 2000 cm⁻¹, verified by standard method (e.g., polystyrene film).
Fixed Path-Length Gas Cell:
• Aluminum body coated with corrosion-resistant material; chemical resistance details required.
• Operates up to 3 ATM with calibrated pressure gauge and overpressure relief valve.
• Easily accessible mirrors with specified reflectivity and included realignment tools.
• Heated with precise temperature control, capable of reaching ≥100°C.
User Manual:
• Comprehensive electronic (PDF) and printed manuals including safety notices, operating instructions, maintenance schedules, software operation, calibration procedures, troubleshooting, and technical support contacts.
Analytical Software:
• Latest manufacturer’s version with full instrument control.
• Automatic compensation for water vapor and CO₂ interference without separate reference spectrum; specify algorithm.
• Supports robust quantitative gas analysis methods compatible with existing spectral libraries;
detail algorithms (e.g., CLS, PLS).
• Displays sample metadata alongside spectra (8000–300 cm⁻¹).
• Allows simultaneous data acquisition and post-processing.
• Includes standard spectral processing functions (normalization, subtraction, overlay, baseline correction, quantitative analysis).
• Compatible with Microsoft Windows 11 and includes Microsoft Office Small Business Edition or equivalent.
• Peak find capability for easy spectral peak identification.
Computer Hardware:
• Dedicated computer and monitor meeting or exceeding manufacturer’s recommended specifications; specify minimum requirements (processor, RAM, storage, graphics, display resolution).
Supplies and Special Tools:
• All unique supplies for operation plus 60-day consumables supply (e.g., desiccant, cleaning materials).
• Special tools for routine maintenance and gas cell alignment are included with detailed list.
On-Site Installation, Verification, and Training:
• One (1) day set-up and One (1) day training minimum 8 hours on-site installation and performance verification by qualified service engineer, including testing of resolution, wavenumber accuracy, and signal-to-noise ratio.
• Additional 8 hours training for Air Force chemists covering basic operation, custom quantitative method development, and programming.
• Documentation of installation and verification procedures provided.
Bruker Hyperion II FT-IR Microscope – Quantity: 1
• Advanced materials characterization combining FT-IR and QCL imaging with transmission, reflection, and ATR modes for polymers, composites, coatings, and other non-metallic materials.
• High spatial resolution down to diffraction limit for microscopic feature and defect analysis.
• Rapid imaging capabilities with ILIM for single wavenumber measurements to accelerate quality control.
• TE-MCT detectors eliminate need for liquid nitrogen, reducing maintenance and operational costs.
• Selectable apertures enhance isolation and analysis of specific regions.
• Ideal for polymer blends, laminates, coatings, and surface/interfacial studies via ATR imaging.
• Enables failure analysis through microscopic chemical change detection.
• Flexible accessories and optional cooling/heating stages support diverse R&D experiments.
• Combines spectral and visual images for simplified data interpretation and deeper material insights.
On-Site Installation, Verification, and Training:
• One (1) day set-up and One (1) day training minimum 8 hours on-site installation and performance verification by qualified service engineer, including testing of resolution, wavenumber accuracy, and signal-to-noise ratio.
• Additional 8 hours training for Air Force chemists covering basic operation, custom quantitative method development, and programming.
• Documentation of installation and verification procedures provided.