Statement of Work

Diagnostic Requirements - Must comply with the ANSI/AAMI/ISO 81060-2:2013 standard for blood pressure measurement accuracy. Class II medical device manufactured in FDA-regulated NYBP facility Accuracy +/- 5 mmHg (millimeters of mercury)

• Measurement Range - Systolic 50 - 255 mmHg and Diastolic 25 - 220 mmHg

• Measurement Method - Oscillometric method employing stepwise pressure deflation technique

• Overpressure Limit- 290 mmHg maximum

• Pulse Rate – 40 BPM – 250 BPM

• Patented Cuff Arm size circumference of 7 in - 17 in with a test time typically less than 1 minute

Operating Requirements

• Operating Temperature - 0ºC to 40ºC (32ºF to 104ºF)

• Storage Temperature - -20ºC to +65ºC (-4ºF to 149ºF)

• Operating Humidity - to 90% Non-condensing

• Storage Humidity to 90% Non-condensing

• Input Power 100 - 240 V, 50 - 60 Hz, 1.3 amp, 3 Wire Ground Outlet

Features

• Blood Pressure and Pulse Rate

• Proven Accuracy

• Fully-Automated Upper-Arm Pressure Cuff

• LED Display of Readings

• Blood Pressure Guidelines Display

• Oscillometric Methodology

• Four Information Display Panels

• User Record Card Holder