Statement of Work
Diagnostic Requirements - Must comply with the ANSI/AAMI/ISO 81060-2:2013 standard for blood pressure measurement accuracy. Class II medical device manufactured in FDA-regulated NYBP facility Accuracy +/- 5 mmHg (millimeters of mercury)
• Measurement Range - Systolic 50 - 255 mmHg and Diastolic 25 - 220 mmHg
• Measurement Method - Oscillometric method employing stepwise pressure deflation technique
• Overpressure Limit- 290 mmHg maximum
• Pulse Rate – 40 BPM – 250 BPM
• Patented Cuff Arm size circumference of 7 in - 17 in with a test time typically less than 1 minute
Operating Requirements
• Operating Temperature - 0ºC to 40ºC (32ºF to 104ºF)
• Storage Temperature - -20ºC to +65ºC (-4ºF to 149ºF)
• Operating Humidity - to 90% Non-condensing
• Storage Humidity to 90% Non-condensing
• Input Power 100 - 240 V, 50 - 60 Hz, 1.3 amp, 3 Wire Ground Outlet
Features
• Blood Pressure and Pulse Rate
• Proven Accuracy
• Fully-Automated Upper-Arm Pressure Cuff
• LED Display of Readings
• Blood Pressure Guidelines Display
• Oscillometric Methodology
• Four Information Display Panels