Inspection Report.pdf
PDF 890 KB Posted
- Attached to
- 6835--Bulk Oxygen Federal contract opportunity
- Solicitation number
- Not on record
About this file
This is a Preventive Maintenance (PM) Inspection Report for a medical bulk oxygen system. The inspection was conducted on October 14, 2025, at the George H. Obrien VA Medical Center in Big Spring, Texas, by technician Ben Hill. The system consists of a 1500-gallon bulk oxygen tank (Model E-240400, manufactured by Linde) with a Maximum Allowable Working Pressure (MAWP) of 250 psig, equipped with pressure safety valves, burst disks, and dual ambient vaporizers (Thermax/Chart TMF048A-HF-SG). The system includes a Generant regulator (Model 2GDR-1000B-V-BM) set to deliver pressure at 55 psig with a final line pressure safety valve set at 75 psig, along with cylinder reserve supply consisting of 9 x 51-inch cylinders with 2100 psig pressure.
All 15 inspection categories were evaluated and passed, including overall site condition, vessel condition, equipment condition, and medical system requirements. Critical safety measures verified include adequate site access and ventilation, vessel anchoring and concrete pad integrity, pressure relief device installation, piping leak checks, and compliance with NFPA 55 standards. The medical alarm system was tested and confirmed functional with set points for main vessel low level (30 inH₂O), reserve-in-use (47 inH₂O), and reserve low pressure (1600 psig). Notably, during the inspection, technician Ben Hill replaced pigtails and discovered six isolated empty cylinders on the manifold, which were replaced with full cylinders available on site, and two additional missing cylinders that were also replaced. No other equipment issues were identified, and the system was verified to automatically switch to reserve supply with no unauthorized modifications.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C25726Q0738 0001.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Work Order Submitted Date Submitted By
4000667318 Oct 16, 2025 BENJHILL
CUSTOMER INFORMATION
Work Order # 4000667318
Inspection Date 10/14/2025
Customer GEORGE H OBRIEN VA MEDICAL CENTER
Address
300 W VETERANS BLVD, BIG SPRING, TX 79720-
Account 3462728
Contact
Phone
Product OXYGEN USP BULK
Location Type Outdoor
Customer Owned NO
Type of Inspection PM
MAIN VESSEL INFORMATION
Tank Size 1500 GA
F5010 - AIR-119 - Bulk - PM Inspection - Medical LOX/LIN/N2O/CO2 System
AIR-119 - Bulk - PM Inspection - Medical LOX/LIN/N2O/CO2 System
Tank Code E-240400
NB# 2488
S/N E-240400
MFR Name LINDE
MFG Year
MAWP 250
PSV Inlet Size .75 in
PSV Outlet Size 1 in
PSV Set Point 250 psig
BD Inlet Size .75 in
BD Outlet Size .75 in
BD Set Point 339 psig
BAM# or FF# B0501600002473
Telemetry Type BAM
BAM Phone # 5278638651
Level Gauge Full 140 inH2O
Level Gauge Reorder 35 inH2O
RESERVE SUPPLY INFORMATION
Type Cylinders
For Reserve Cylinders, Qty
Cylinder Size (cylinder height excludes valve) 9 x 51
1. OVERALL SITE CONDITION
1.1 Adequate trailer access, clear for safe delivery?
(clear line of sight, less than 25’ from trailer to fill valves)?
Yes
1.2 Adequate lighting for night time deliveries? Yes
1.3 Site secured, fenced, and gates are locked? Yes
1.4 Vessel and system is protected where exposed
to vehicle impact by bollards or other approved means?
Yes
1.5 Site has adequate ventilation or mitigation (vents
captured, etc.) if indoor or courtyard installation?
Yes
1.6 Adequate and safe access to service
equipment? (3' around vessel and other equipment)
Yes
1.7 Pipe supports anchored, pipe clamps tight, and
unistrut stand plastic caps installed?
Yes
1.8 Vessel, vaporizers, and equipment anchored? Yes
1.9 Vessel and equipment concrete pad is in good
condition (no spalling, cracks, frost heaves, etc.)
Yes
1.10 Site has adequate drainage? Yes
1.11 Pressure relief devices installed between block
valves and discharge in a safe direction?
Yes
1.12 Electrical equipment and conduit in good
condition (no signs of damage, labeled, etc.)?
Yes
1.13 System piping and components leak checked
and free of leaks?
Yes
1.14 Overall site condition is free of debris, chemical
storage, combustibles, or other hazards?
Yes
1.15 Exposure distances comply with NFPA 55 and Airga s Standards?
Yes
NFPA Distances Table
LINK:
https://docs.google.com/spreadsheets/d/1vuovRlq8 X7jCJ9Qcrfn7jGR7JRkOI_iGpPdDt44i_ec/edit?
usp=sharing
1.16 Concrete off-load pad is installed, in good
condition, and free of combustibles for O2 or N2O installations?
Yes
1.17 'No Smoking' signs posted for O2 or N2O installations?
Yes
1.18 Grounding system (if installed for O2 and N2O
tanks, lightning protection, etc.) is in good condition and tight?
Yes
1.19 Telemetry system is functioning properly and
solar panel cleaned (if appl.)?
Yes
2. VESSEL CONDITION (MAIN TANK)
2.1 MAIN TANK: Vessel cosmetic condition (paint, dirt, logo, etc.) is adequate?
Yes
2.2 MAIN TANK: All required Airgas, NFPA, and
warning decals are in place, correct, and legible?
Yes
2.3 MAIN TANK: Tank code is legibly stenciled or
labeled on vessel (verify code if instructed)?
Yes
2.4 MAIN TANK: Vessel has pressure relief device
on the fill line?
Yes
2.5 MAIN TANK: Pressure gauges are functional
and legible?
Yes
2.6 MAIN TANK: Liquid level gauge is functional, legible, and zero checked, and has 'reorder' and 'full' arrows?
Yes
2.7 MAIN TANK: Vessel has no signs of poor
vacuum such as venting, cold spots, etc; (take vacuum reading if suspect)?
Yes
2.8 MAIN TANK: Fill connection correct, in good
condition, and equipped with dust or pressure cap?
2.9 MAIN TANK: Fill line check valve is installed &
operationally checked by safely opening top fill valve to check back-flow?
Yes
2.10 MAIN TANK Hose drain and full trycock valves
are installed and working properly?
Yes
2.11 MAIN TANK: Pressure building system tested
and is operating correctly at system set point per data label?
Yes
2.12 MAIN TANK: All valve handles are in good
working order and valves tagged properly?
Yes
2.13 MAIN TANK: Vessel has diverter assembly with
dual PSVs/BDs? Cycle to both sides to check flow and for any leaks.
Yes
2.14 MAIN TANK: Vessel PSVs are in good
condition, legible, seals intact, and have no signs of weeping?
Yes
2.15 MAIN TANK: Vessel PSVs marked test date will
be less than 5 years old at time of next PM (capture test date)?
Yes
Test date 09/14/2022
2.16 MAIN TANK: Vessel PSVs are secure, free
from obstruction, and venting towards ground or safe location?
Yes
2.17 MAIN TANK: Vessel burst disks (BDs) visually
inspected; set point < 150% tank MAWP, good condition, and not weeping?
Yes
EQUIPMENT INFORMATION
Main Vaporizer Type Ambient
Main Vap. Qty. 2
Main Vap. MFG Thermax/Chart
Main Vap. Model # TMF048A-HF-SG
Reserve Vaporizer Type
Reserve Vap. Qty.
Reserve Vap. MFR
Reserve Vap. Model #
PCM Type Liq Main/Cyl Reserve
Reg. MFG Generant
Reg. Model 2GDR-1000B-V-BM
Reg. Size 1
Delivery Pressure Set Point 55
Final Line PSV Set Point: 75
Auxiliary Manifold (Hyperbarics) No
Aux Manifold Set Point
Aux PSV Set Point
Vaporizer Switching Manifold No
Local Alarm Light Panel Yes
Other Equip. Notes
3. EQUIPMENT CONDITION
3.1 Piping material and/or insulation is correct for
the service and in good working condition?
Yes
3.2 Process vaporizers at least 50% free of ice build
up?
Yes
3.3 Vaporizer switching (if installed) verified for
proper operation?
N/A
3.4 Vaporizers are protected with thermal PSV and
inspected for set point, orientation, and condition?
3.5 Line regulators are inspected and checked for
proper operation and set point?
Yes
3.6 Final line PSV inspected and is in good
condition, legible, and has no signs of weeping?
Yes
4. MEDICAL SYSTEM REQUIREMENTS
4.1 Bulk system alarm contacts are wired to the
facility?
Yes
4.2 Discussed with facility their requirements to test
alarms?
Yes
4.3 Main vessel low level alarm signal works
correctly?
Yes
Set Point 30
4.4 Reserve-in-use alarm signal works correctly? Yes
Set Point 47
4.5 Reserve vessel low pressure alarm or reserve
cylinders low pressure alarm works correctly?
Yes
Set Point 1600
4.6 Reserve vessel low level alarm works correctly? N/A
4.7 Local alarm light panel (if installed) functions
properly?
Yes
4.8 Emergency oxygen supply connection (EOSC)
installed? (N/A for stand-alone hyperbarics facility.)
Yes
4.9 Cylinder Reserve: In acceptable condition, with
legible product label and lot number, and properly secured?
Yes
4.10 Cylinder Reserve: Pressure is acceptable (not
depleted)? Consider replacing if ~1900 psig.
Pressure 2100
4.11 System will automatically switch to reserve
supply
Yes
4.12 System verified to be free of unauthorized
changes?
Yes
5. NITROUS OXIDE AND CARBON DIOXIDE
5.1 Each fill connection has blow down valve at the
CGA connection and hose restraint cable in good condition?
N/A
5.2 Electrical disconnects appear to be in good
condition and no visual signs of blown fuses, loose conduit?
N/A
5.3 Refrigeration unit and compressor verified for
proper operation and maintained per MFR recommendations?
N/A
5.4 Vessel pressure is above 250 psig and within
design range on data label? Record pressure in notes.
N/A
5.5 Pressure building system tested and is operating
correctly at system set point per data label?
N/A
5.6 N2O only: 240 V power plug with disconnect and
grounding source for N2O truck are in good condition?
N/A
FORM CLOSURE
Comments
Replaced pigtails. While changing pigtails found that six of the cylinders were isolated off and empty.
Replaced cylinders with full ones that were on site.
Also replaced two missing cylinders that were not on manifold. Found no other issues with equipment.
Technician Name Ben Hill
Technician Signature
Customer Name
Customer Signature