Compliance Checklist: Joint Commission’s Imaging Standards
Use this tool to assess your organization’s compliance with The Joint Commission’s requirements that relate to the provision of imaging services.
* = Frequent area of noncompliance Updated April 2020
Modality: Magnetic Resonance Imaging (MRI) Requirement: Areas to assess: Yes No Comment Initials
EC.02.01.01 EP 14
(Applies only to MRI)
Processes are in place to address the following MRI safety risks (staff can describe processes for):
- patients with claustrophobia, anxiety, emotional distress -urgent/emergent patient care needs -patients with medical implants, devices, embedded metallic objects -preventing entry of ferromagnetic objects into MRI are a (only MRI-safe equipment e.g. fire extinguishers)* -protecting patients from acoustic noise
EC.02.01.01 EP 16
Access to the MRI area is restricted.* -all staff and patients are screened prior to entering the MRI area.* -there are controls in place to prevent unauthorized access to the MRI area* -warning signage is posted at the entrance to the MRI scanner* -signage is posted indicating that the magnet is always on (as applicable)*
EC.02.04.01 EP 4
(Applies to CT, PET, NM, MRI, x-ray and fluoroscopy)
Identifies activities and frequencies in writing for maintaining, inspecting and testing all medical equipment on inventory
EC.02.04.01 EP 10
NM, MRI and fluoroscopy)
Equipment quality control and maintenance activities are identified Timeframes are established for how often they are to be done
EC.02.04.03 EP 18
NM, and MRI)
Equipment quality control and maintenance activities are done QC logs are complete*
EC.02.04.03 EP 22
A performance evaluation is performed annually by an MRI scientist, and includes all required tests.
Evaluation/ testing results and recommendations are documented.
Use this tool to assess your organization’s compliance with The Joint Commission’s requirements that relate to the provision of imaging services.
* = Frequent area of noncompliance Updated April 2020
Modality: Magnetic Resonance Imaging (MRI)
Requirement: Areas to assess: Yes No Comment Initials
EC.02.04.03 EP 25
(Applies to CT, PET, NM, and MRI)
A performance evaluation that includes all required tests and parameters is performed on each image acquisition monitor annually by a medical physicist or MRI scientist (for MRI only)
HR.01.05.03 EP 25
Documentation of staff annual training and ongoing education on all required topics is available
PI.01.01.01 EP 34
Data is collected on any MRI-related patient thermal injuries
PI.01.01.01 EP 35
Data is collected on:
-incidents where have unintentionally entered the MRI scanner room -injuries resulting from the presence of ferromagnetic objects in the MRI scanner room
Use this tool to assess your organization’s compliance with The Joint Commission’s requirements that relate to the provision of imaging services.
* = Frequent area of noncompliance Updated April 2020
Modality: Computed Tomography (CT)
EC.02.02.01 EP 17
(Applies to CT, NM, PET, Fluoroscopy)
Dosimetry monitoring results reviewed quarterly by radiation safety officer, diagnostic medical physicist, health physicist; assess ALARA
EC.02.02.01 EP 18
PET, Fluoroscopy, x-ray)
Radiation workers checked periodically for amount of radiation exposure
NM, MRI, x-ray and fluoroscopy)
Identifies activities and frequencies in writing for maintaining, inspecting and testing all medical equipment on inventory
EC.02.04.01 EP 5
NM, MRI, x-ray and fluoroscopy)
Deemed: Maintenance, inspection testing in accordance with manufacturer’s recommendations
(Applies to CT, PET, NM, MRI and fluoroscopy)
Equipment quality control and maintenance activities are identified Timeframes are established for how often they are to be done
(Applies to CT, PET, NM, and MRI)
Equipment quality control and
EC.02.04.03 EP 20
(Applies only to CT)
At least annually:
- the CTDI is measured for adult brain, adult abdomen, pediatric brain, and pediatric abdomen, or other commonly used protocols and
- the radiation dose for each protocol is verified to be within 20% of dose displayed -the measurements and dose verifications are done by a medical physicist
Use this tool to assess your organization’s compliance with The Joint Commission’s requirements that relate to the provision of imaging services.
* = Frequent area of noncompliance Updated April 2020
Modality: Computed Tomography (CT)
Requirement Areas to assess: Yes No Comment Initials
EC.02.04.03 EP 21
A performance evaluation is performed annually by a medical physicist and includes all required tests. Evaluation/ testing results and
NM, and MRI)
A performance evaluation that includes all required tests and parameters is performed on each image acquisition monitor annually by a medical physicist or MRI scientist (for MRI only)
EC.02.06.05 EP 4
and NM)
A structural radiation shielding design assessment is conducted prior to imaging equipment installation or room modification The survey is conducted by a medical physicist or health physicist
EC.02.06.05 EP 6
and NM)
A radiation protection survey is conducted after installation of imaging equipment or construction The survey is done prior to clinical use of the room and is conducted by a medical physicist or health physicist
HR.01.01.01 EP 32
Documentation is available for verification of specified qualifications for each technologist who performs diagnostic CT exams
HR.01.01.01 EP 33
Documentation is available of verification of specified qualifications for each medical physicist supporting CT services
Use this tool to assess your organization’s compliance with The Joint Commission’s requirements that relate to the provision of imaging services.
* = Frequent area of noncompliance Updated April 2020
Modality Computed Tomography (CT)
Requirement Areas to assess: Yes No Comment Initials
HR.01.05.03 EP 14
Documentation of staff annual training and ongoing education is available.
The training includes:
-radiation dose optimization techniques -safe operation of CT equipment they will use
PC.01.02.15 EP 5
Radiation dose index is documented for on every CT exam.
The dose index is exam specific, summarized by series or anatomic area and retrievable
PC.01.02.15 EP 12
NM, MRI)
Appropriateness of exam (patient age, recent imaging exams, exam type)
PC.01.03.01 EP 26
Imaging protocols are reviewed, kept current. Input is provided by an interpreting MD, medical physicist, and imaging technologist.* Protocols are reviewed per established timeframes
PI.02.01.01 EP 6
Incidents where radiation dose indices exceeded expected dose index range are reviewed and analyzed. These incidents are compared to external benchmarks
Use this tool to assess your organization’s compliance with The Joint Commission’s requirements that relate to the provision of imaging services.
* = Frequent area of noncompliance Updated April 2020
Modality: Nuclear Medicine (NM) & Positron Emission Tomography (PET)
Requirement: Areas to assess: Yes No Comment Initials
EC.02.02.01 EP 6
(Applies to NM, PET)
Minimizes risk associated with selecting, handling, storing, transporting, using and disposing of radioactive materials
EC.02.02.01 EP 17
PET, Fluoroscopy)
Dosimetry monitoring results reviewed quarterly by radiation safety officer, diagnostic medical physicist, health physicist; assess ALARA
EC.02.02.01 EP 18
(Applies to deemed:
CT, NM, PET,
fluoroscopy, x-ray)
Radiation workers checked periodically for amount of radiation exposure
NM, MRI, x-ray and fluoroscopy)
Identifies activities and frequencies in writing for maintaining, inspecting and testing all medical equipment on inventory
NM, MRI, x-ray and fluoroscopy)
Deemed: Maintenance, inspection
NM, MRI and fluoroscopy)
Equipment quality control and maintenance activities are identified
NM, and MRI)
Equipment quality control and
EC.02.04.03 EP 1
(Applies to MRI, CT, NM,PET, x-ray, Fluoroscopy, Performs safety, operational, functional checks Non-deemed: before initial use.
Deemed: Before initial use and after major repairs or upgrades
Use this tool to assess your organization’s compliance with The Joint Commission’s requirements that relate to the provision of imaging services.
Modality: Nuclear Medicine (NM) & Positron Emission Tomography (PET)
EC.02.04.03 EP 16
(Applies to NM, PET)
Qualified staff inspect, test and calibrate NM equipment annually.
Results, completion dates documented.
EC.02.04.03 EP 23
(Applies only to NM)
A performance evaluation is performed annually by a medical physicist or nuclear medicine physicist and includes all required tests.
Evaluation/ testing results and
EC.02.04.03 EP 24
(Applies only to PET)
A performance evaluation that includes all required tests is performed annually by a medical physicist Evaluation/testing results and
NM, and MRI)
A performance evaluation that includes all required tests and parameters is performed on each image acquisition monitor annually by a medical physicist or MRI scientist (for MRI only)
EC.02.06.05 EP 4
NM)
Structural shielding assessment prior to installation of new/replacement of existing imaging equipment; rooms where ionizing radiation emitted or radioactive material stored, medical or health physicist
EC.02.06.05 EP 6
NM)
Radiation protection survey to verify adequacy of shielding after installation of imaging equipment or construction by medical or health physicist
MM.06.01.01 EP 13
(Applies to Nuclear Medicine)
Verified dose is within 20% prescribed dose or range
Use this tool to assess your organization’s compliance with The Joint Commission’s requirements that relate to the provision of imaging services.
Modality: Nuclear Medicine (NM) & Positron Emission Tomography (PET)
MS.03.01.01 EP 16
(Applies to CT, NM.
PET, x-ray, fluoroscopy)
Deemed: Medical staff determines qualifications radiology staff who use equipment & administer procedures
MS.03.01.01 EP 17
(Applies to NM, PET)
Deemed: Approves nuclear services doctor qualifications, training, functions, responsibilities NM staff
Use this tool to assess your organization’s compliance with The Joint Commission’s requirements that relate to the provision of imaging services.
Modality: Fluoroscopy
NM, MRI, x-ray and fluoroscopy)
Deemed: Maintenance, inspection
NM, MRI and fluoroscopy)
Equipment quality control and maintenance activities are identified
EC.02.04.03 EP 34
(Applies to fluoroscopic units)
A performance evaluation that includes all required tests is performed annually by a diagnostic medical physicist. Evaluation/testing results are documented
PC.01.02.15 EP 13
(Applies to fluoroscopy)
Cumulative air kerma, kerma area product (fluoroscopy time, number of spot films) documented in retrievable format
Miscellaneous
Requirement Areas to assess: Yes No Comment Initials
LD.04.01.05 EP 25
(applies if any radiologic services are provided)
Designated radiation safety officer with authority and responsibilities as described