This is an Invitation for Bid (IFBC) from the State of Missouri Department of Natural Resources for Multi-Gas Calibrators (Solicitation Number IFBC30034901900040). The bid seeks three (3) multi-gas calibrators suitable for dynamic calibration of instruments at monitoring sites operated by the State of Missouri. The calibrators must automatically perform zero, precision, span checks, and multi-point calibrations using various gases including carbon monoxide, ozone, sulfur dioxide, nitric oxide, isopropyl nitrate, N-propyl nitrate, and nitrogen dioxide. The bid opened July 5, 2018, and closed July 18, 2018, with delivery required to the Environmental Services Program in Jefferson City. The initial contract period runs from September 20, 2018, through September 19, 2019, with two renewal options available.
The equipment specifications require two thermal mass flow controllers, an ozone generation module, mixing chamber, photometer, three-port glass output manifold, reaction chamber, four gas inlet ports, one dilution inlet port, and control electronics. Technical requirements include flow accuracy of ±1.0% from 10-100% of full scale flow, repeatability of 0.05%, operating temperature of 0-50°C, maximum weight of 30 pounds, and power requirements of 115 VAC at 60 Hertz. Delivery terms are FOB Destination with Net 45 payment terms. The bid documents specify that vendors must provide firm fixed pricing for both the first and second renewal periods.
View the file
Other files
Show all 27
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Solicitation Number: IFBC30034901900040
Multi-Gas Calibrators
Invitation for Bid - Commodity (IFBC)
State of Missouri
July 05, 2018 - July 18, 2018
Page:2
General Header Information
IFBC30034901900040
No. IFBC30034901900040 Title: Multi-Gas Calibrators Start Date: July 05, 2018 at 4:30:00 PM CDT End Date: July 18, 2018 at 3:00:00 PM CDT Vendor Q&A Start Date:
Vendor Q&A End Date:
Estimated Total Value:
Who can respond to this bid? : All Vendors Description: Contract Period: SEPTEMBER 20, 2018 THROUGH SEPTEMBER 19, Delivery Terms: Free On Board Destination Delivery Notes: Department of Natural Resources
Environmental Services Program 2710 W. Main St.
Jefferson City, MO 65101
Payment Terms: Net 45 Days Contact Information: State of Missouri
Tammy Michel 301 West High Street, Room 630 Jefferson City MO, 65101 United States Tel: 573-751-3114 Fax: 573-526-9816 Tammy.Michel@oa.mo.gov
Contact Details: If you have any questions, please contact:
Tammy Michel 301 West High Street, Room 630 Jefferson City MO, 65101 United States Tel: 573-751-3114 Fax: 573-526-9816 Tammy.Michel@oa.mo.gov
Selected Categories:
Page:3
Header Custom Fields:
IFBC30034901900040
NOTICE
Vendor Responsibility
The vendor is solely responsible for ensuring timely submission of their solicitation response, whether submitting an online response or a hard copy response. Failure to allow adequate time prior to the solicitation end date to complete and submit a response to a solicitation, particularly in the event technical support assistance is required, places the vendor and their response at risk of not being accepted on time.
The State of Missouri's Privacy Policy can be accessed here.
http://www.mo.gov/privacy-policy/
Page:4
Solicitation Requirements:
This bid has no questions defined.
IFBC30034901900040
REQUIREMENTS
Requirements:
To access the official solicitation document, which includes all applicable requirements, detailed specifications, and terms and conditions, the vendor must click on the Overview tab to view the solicitation. Under the Overview tab, the vendor must either: 1) print the solicitation document in order to submit a hard copy to Division of Purchasing, or 2) review and accept the solicitation documents under the Overview tab, and submit an electronic response to the solicitation through the MissouriBUYS system.
Page:5
Documents:
IFBC30034901900040.dotx
Page:6
Item Specifications
No. Item Alternative Supplier Part No Mfr.
Name
Mfr. No Del
Date
Unit Unit Bid Qty. Total
1 Multi-Gas Calibrators each 3.00
Additional Item Fields
1. 1st Renewal Period(required):State firm, fixed price for the 1st Renewal Period
2. 2nd Renewal Period(required):State firm, fixed price for the 2nd Renewal Period
3. Delivery:: State delivery days if less than 45 days ARO:
4. Warranty:: State warranty period for parts and labor if vendor's warranty exceeds 2 years:
Item Specification for Multi-Gas Calibrators
Description: The multi-gas calibrator must be suitable for the dynamic calibration of instruments at monitoring sites operated by the State of Missouri. The calibrators must consist of a self-contained instrument capable of automatically performing zero, precision, span checks, multi-point calibrations using carbon monoxide, ozone, sulfur dioxide, nitric oxide, isopropyl nitrate (IPN), N-propyl nitrate (NPN) and nitrogen dioxide gases. The calibrators must include a status board option or equivalent to allow the data acquisition system to operate the calibrators remotely. The calibrators must consist of two (2) thermal mass flow controllers (one air 0 – 20 SLPM and one gas 0 – 100 SCCM), an ozone generation module, a mixing chamber, photometer, three (3) port glass output manifold, a reaction chamber, four (4) gas inlet ports, one (1) dilution inlet port, and control electronics. The calibrators must accept local or remotely entered commands for at least a seven (7) day schedule of calibration activities. The calibrators must produce an accurate concentration of dilution and span gases within +1% of 10 to 100% of full scale flow. The calibrators must accept commands from a display panel, contact closures or an RS-232 serial port, or Ethernet connection. The ozone generator must have the capability to function at the following ranges: 0.02 –
0.5 PPM at 5 – 10 SLPM and 0.05 – 1 PPM at 5 – 10 SLPM and have and accuracy of +/- 2% or +/- 0.005 PPM at 10 SLPM. The ozone generator must be calibrated at the factory with a NIST traceable ozone standard. The ozone generator must compensate automatically for changes in pressure and lamp deterioration. The calibrator must store mass flow controller data and retain it through power outages. The calibrator must have stainless steel and Teflon fittings. Brass fittings shall be unacceptable. The calibrator must include any needed Teflon tubing. The calibrators shall meet or exceed the following requirements: • Flow accuracy: ó 1.0% from 10 to 100% of full scale flow • Repeatability: 0.05% from 10 to 100% of full scale flow • Warm-up Time: 30 minutes • Operating
Temperature: 0 – 50 C • Maximum Weight: 30 pounds • Power Requirements: 115 VAC @ 60 Hertz The calibrators shall include any special tools necessary to adjust, dismantle or service the equipment.
Allow Vendor to Enter Negative Value : No