Attachment H - Down Selection Evaluation Criteria (REVISED) - Amendment 0002.docx

DOCX document 22 KB Posted

Attached to
Ultrasound Field Portable (USFP) Assessment and Development Efforts
Solicitation number
HT9425-23-R-0065
Issued by

About this file

This document outlines down selection evaluation criteria for an ultrasound field portable system solicitation. Key criteria include environmental testing of systems for reliability, operational assessment of usability with military end-users, and price reasonableness evaluation. Systems will undergo testing and be rated on attributes such as FDA clearance for trauma and musculoskeletal exams, reliability with mean time between failures, power source duration, weight, size, temperature range, storage capacity, and number of transducers. Acceptable ratings must meet all thresholds and objectives where specified. Price reasonableness will also be evaluated for fair and competitive pricing of 3,500 systems including support. The Defense Health Agency solicitation seeks these ultrasound field portable assessment and development efforts.

View the file

Other files

File Type Posted
Vendor Questions and Government Responses RPP HT9425-23-R-0065-0003 USFP.pdf PDF
Request for Prototype Proposal (RPP) HT9425-23-R-0065-0003 USFP.docx DOCX document
Attachment E - Statement of Work (revised) - Amendment 0002 (showing changes).pdf PDF
Attachment G - Cost Price Template (REVISED) - Amendment 0002.xlsx XLSX spreadsheet
Request for Prototype Proposal (RPP) HT9425-23-R-0065-0002 USFP (REVISED) (showing changes).pdf PDF
Attachment A - Statement of Objectives (REVISED) - Amendment 0002.docx DOCX document
Request for Prototype Proposal (RPP) HT9425-23-R-0065-0002 USFP (REVISED).docx DOCX document
Attachment F - Agreement (terms considitions) (REVISED) - Amendment 0002 (showing changes).pdf PDF
Attachment E - Statement of Work (REVISED) - Amendment 0002.docx DOCX document
Attachment A - Statement of Objectives (REVISED) - Amendment 0002 (showing changes).pdf PDF
Vendor Questions and Government Responses RPP HT9425-23-R-0065-0002 USFP.pdf PDF
Attachment F - Agreement (terms considitions) (Revised) - Amendment 0002.doc DOC document
Attachment H - Down Selection Evaluation Criteria (REVISED) - Amendment 0002 (showing changes).pdf PDF
B08 Request for Prototype Proposal (RPP) HT9425-23-R-0065-0001 USFP.docx DOCX document
Request for Prototype Proposal (RPP) HT9425-23-R-0065 USFP.docx DOCX document
Attachment B - Warranties and Representatives.docx DOCX document
Attachment C - Biomedical Technology Readiness Levels (TRL).pdf PDF
Attachment H - Down Selection Evaluation Criteria.docx DOCX document
Attachment E - Statement Of Work (SOW) Template.docx DOCX document
Attachment D - Medical Device Evaluation and Risk Assessment (MDERA).docx DOCX document
Attachment A - Statement of Objectives.docx DOCX document
Attachment F - Agreement (terms considitions).doc DOC document
Attachment I - MIL-STD-810H.pdf PDF
Attachment J - Acronyms.docx DOCX document
Attachment G - Cost-Price Template.xlsx XLSX spreadsheet
Show all 25

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Attachment H - Down Selection Evaluation Criteria

The ultrasound system will be evaluated and down selected based on the following evaluation criteria:

For down select purposes, system attributes that are weighted as “high” are more important than system attributes weighted as “medium” and “low,” during environmental and operational testing. Additionally, systems attributes that meet the objective will be evaluated “higher,” in comparison to system attributes only meeting the threshold. Price and overall usability will also be considered. Down selection will be based on a comprehensive review and assessment against the evaluation factors below.

Evaluation Factors:

Environmental testing Operational Assessment/Usability Pricing Reasonableness

· Environmental Testing The Ultrasound Field Portable (USFP) system will undergo environmental testing. The system will be rated according to the data received and how the system performed in each attribute below.

· Operational Assessment/Usability

· The Operational Assessment will contain scenarios in a simulated operational environment where military end-users will use the USFP system and then will provide feedback. The system will be rated according to the information received in the feedback and how the system performed according to the below attributes during the Operational Assessment.

· Overall usability of the entire USFP system (including accessories) will be considered and rated, which may include other attributes in the MIL-STD-810. “Usability” is defined as the effectiveness, efficiency, ease of use and user satisfaction while an end-user interacts with the USFP system in a simulated operational environment.

· Price Reasonableness

· Price reasonableness of the 3,500 USFP systems (Deliverable 12). For a price to be reasonable, it must, in its nature and amount, represent a price to the Government that a prudent person would pay in the conduct of competitive business. Normally, price reasonableness is established through price analysis.

Ratings

Ratings for Environmental Testing attributes
Description
EXCEPTIONAL
The USFP system attributes performed well under military environmental conditions, received no concerns from testing AND MEETS one or more OBJECTIVE attributes. Risk of unsuccessful performance is very low.
GOOD
The USFP system attributes performed well under military environmental conditions and received no concerns from testing. Risk of unsuccessful performance is very low.
ACCEPTABLE
The USFP system attributes performed well under military environmental conditions and received some minor concerns from testing AND meets the threshold. Risk of unsuccessful performance is low.
UNACCEPTABLE
The USFP system attributes received one or more notable concern(s) from testing. USFP systems is not acceptable.
Ratings for Operational Assessment/ Usability
Description
EXCEPTIONAL
The USFP system attributes performed well during operational assessment and received no concerns from military test users. Risk of unsuccessful performance is very low.
GOOD
The USFP system attributes performed well during operational assessment and received some minor concerns from military test users. Risk of unsuccessful performance is low.
ACCEPTABLE
The USFP system attributes received one or more notable concern(s) from military test users. There is risk of unsuccessful performance.
UNACCEPTABLE
The USFP system was unable to perform the operational assessment and/or the end user was unable to use the system. USFP system is not acceptable and/or has critical failures.
Ratings for Price reasonableness
Description
ACCEPTABLE
Pricing for the USFP system (including logistical support and all associated accessories) is fair, reasonable, and is competitively priced. The written price narrative reflects sound, rational judgment in deriving and applying historical data and pricing methodologies.
UNACCEPTABLE
Pricing for the USFP system (including logistical support and all associated accessories) is not fair, reasonable, and/or competitively priced. The written price narrative does not reflect sound, rational judgment in deriving and applying historical data and pricing methodologies.
Attribute:
Evaluation Weight:
Threshold:
Objective:
U.S. FDA Clearance
High
Shall have U.S. FDA Clearance
Shall have U.S. FDA Clearance
Focused Assessment with Sonography for Trauma (FAST) Capable
High
Shall be FDA Cleared for use in the abdomen and capable of performing FAST exams
Shall be FDA Cleared for use in the abdomen and capable of performing FAST exams
Extended Focused Assessment with Sonography for Trauma (e-FAST) Capable
High
Shall be FDA Cleared for us e in the abdomen and for cardiac use, and capable of performing eFAST exams
Shall be FDA Cleared for us e in the abdomen and for cardiac use, and capable of performing eFAST exams
Musculoskeletal (MSK) Capable
High
Shall be FDA Cleared for musculoskeletal use and capable of performing MSK/MSKI exams
Shall be FDA Cleared for musculoskeletal use and capable of performing MSK/MSKI exams
Reliability
Medium
The USFP system should have a 95% probability of completing 24 total run time hours within a 3-day timespan without a System Abort (SA), which equates to 468 hours Mean Time Between System Abort (MTBSA).
The USFP system should have a 95% probability of completing 24 total run time hours within a 3-day timespan without a System Abort (SA), which equates to 468 hours Mean Time Between System Abort (MTBSA).
Power Source Duration
Low
The power source should last for two hours at the highest setting
The power source should last for five hours or more at the highest setting
Power Source -
low
Should have a rechargeable power source
Should have a rechargeable and interchangeable power source
Weight
Low
The complete USFP system (screen & transducer(s)) should not weight more than three pounds
The complete USFP system (screen & transducer(s)) should not weight more than 0.5 pounds
Size
Low
The complete USFP system (screen & transducer(s)) should be a cube of no more than 280 cubic inches
The complete USFP system (screen & transducer(s)) should be a cube of no more than 160 cubic inches
High Temperature
Low
The fully functional USFP system should be operational at all settings at 49°C
The fully functional USFP system should be operational at all settings at 60°C
Low Temperature
Low
The fully functional USFP system should be operational at all settings at -1°C
The fully functional USFP system should be operational at all settings at -26°C
Transducer(s)
Low
The USFP system should have transducers to perform FAST, eFAST and MSK exams
The USFP system should have one transducer to perform FAST, eFAST and MSK exams (can have multiple heads on transducer)
Storage
Low
The USFP system should have at least 8 GB of storage space
The USFP system should have at least 16 GB of storage space

(End of Down Selection Evaluation Criteria)