Radiopharmaceutical Calibrators
Solicitation number
HT940827QE013
Issued by

About this file

This is a Solicitation/Contract/Order for Commercial Products and Commercial Services (Standard Form 1449) issued by the Defense Health Agency for the procurement of radiopharmaceutical calibrators.

The solicitation requires the delivery of three complete radiopharmaceutical dose calibrator systems for the Department of Nuclear Medicine at William Beaumont Army Medical Center (WBAMC) in Fort Bliss, Texas. Each system must include a dose calibrator with an 8-inch VGA color touchscreen, an integrated or compatible well counter, shielding, roll printer, ticket printer, dose calibrator inserts, dipper, and compatible dose-label tickets. Essential specifications include user-selectable curie or becquerel readings, USB and personal-computer communications capability, memory for more than 80 radionuclides, capability to measure fluorine-18 doses up to 2 curies (74 gigabecquerels) and technetium-99m up to 6 curies (222 gigabecquerels), plus enhanced quality-control functions. The ionization chamber must have specific dimensions and performance parameters including accuracy better than ±2%, linearity within ±2%, and repeatability within ±1% over 24 hours. All equipment must be new, fully functional, and delivered by December 31, 2026, between 8:00 a.m. and 3:00 p.m. Mountain Standard Time, Monday through Friday, excluding Federal holidays.

The solicitation number is HT940827QE013, with an offer due date of November 25, 2026, at 11:00 a.m. Eastern Standard Time. The contract is 100% unrestricted and employs firm fixed pricing. Inspection and acceptance will occur at both the contractor's destination and AMC Beaumont Bliss in Fort Bliss. The solicitation incorporates multiple FAR and DFARS clauses, including provisions for electronic payment processing through WAWF (Wide Area Workflow), Buy American-Free Trade Agreements-Israeli Trade Act requirements, tax on certain foreign procurements, and covered defense telecommunications equipment restrictions. The contractor must submit pricing for all line items and provide manufacturer documentation, operating instructions, and warranty information with delivery.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

CUI//PROCURE

Product Description/Salient Characteristics

(Calibrator, Radiopharmaceutical)

1.0 General Information.

1.1 Scope The Contractor shall provide all labor, equipment, materials, packaging, transportation and accessories necessary to deliver three complete, new and fully operational radiopharmaceutical dose calibrator systems that meet all requirements of this Product Description. Each system shall include all required components and accessories, including the dose calibrator, integrated or compatible well counter, shielding, roll printer, ticket printer, dose calibrator inserts, dipper and compatible dose-label tickets. The Contractor shall provide applicable manufacturer documentation, operating instructions, warranty information and standard certifications furnished with the equipment.

1.2 Background The Contractor shall provide, deliver and furnish three (3) complete radiopharmaceutical dose calibrator systems for the Department of Nuclear Medicine at William Beaumont Army Medical Center. Each system shall include a dose calibrator, integrated or compatible well counter, shielding, roll printer, ticket printer, dose calibrator inserts, dipper and compatible dose-label tickets. The systems shall meet all essential/salient characteristics identified in this Statement of Work and support the measurement, verification and documentation of radiopharmaceutical doses used in Nuclear Medicine and radiation therapy procedures.

1.3 Applicable Documents. The Contractor shall perform in accordance with the Specifications/Salient Characteristics, the resulting contract, manufacturer instructions, and all applicable Federal, Department of War, Defense Health Agency (DHA), installation, safety and delivery requirements.

1.4 Restrictions: The Contractor shall provide equipment that meets all required salient characteristics in this SOW. The Contractor shall not substitute equipment, components, or accessories that do not meet the stated requirements without prior written approval from the Contracting Officer.

2.0 Schedule & Coordination.

2.1 Contract Management. The Contractor shall coordinate all contract-related matters with the Contracting Specialist and Contracting Officer’s Representative (COR) as applicable.

2.2 Work Hours and Special Conditions. Delivery shall occur between 8:00 a.m. and 3:00 p.m.

Mountain Standard Time, Monday through Friday, excluding Federal holidays.

2.3 Place of Performance Nuclear Medicine Department, William Beaumont Army Medical Center, 18511 Highlander Medic St, Ft. Bliss, TX 79906. The Contractor shall coordinate delivery and access requirements with the COR before delivery.

Attachment 1

2.4 Site Surveys Site survey is not required.

2.5 Start-Up Meeting. Start-up meeting is not required.

3.0 Requirement. Non-applicable for this requirement.

4.0 Contractor Requirements.

The Contractor shall provide all labor, equipment, materials, packaging, transportation, and accessories necessary to deliver complete, operational radiopharmaceutical dose calibrator systems that meet the requirements of this SOW.

The Contractor shall ensure all delivered equipment is new, fully functional, and includes the required components, accessories, printers, inserts, dipper, shielding, and compatible dose-label tickets.

The Contractor shall provide applicable manufacturer documentation, operating instructions, warranty information, and any standard certifications provided with the equipment.

4.1 Security Requirements N/A

5.0 Attachments. N/A

ESSENTIAL/SALIENT CHARACTERISTICS / OPERATION CONDITIONS

6.0 Essential / Salient Characteristics: The Government requires three (3) radiopharmaceutical dose calibrator systems. Each system shall include a dose calibrator, an integrated or compatible well counter, shielding, roll printer, ticket printer, dose calibrator inserts, dipper, and compatible dose-label tickets.

Each calibrator console shall have the following:

• An 8-inch VGA color touchscreen LCD.

• User-selectable curie or becquerel readings.

• A display showing the selected radionuclide, calibration number, measured activity and units.

• USB and personal-computer communications capability.

• USB printer capability.

• Software upgrade capability through USB or flash drive.

• Plug-and-play chamber capability.

• At least 28 programmable radionuclide keys.

• Memory for more than 80 radionuclides, including calibration number and half-life.

• A preset key capable of measuring fluorine-18 (F-18) doses up to 2 curies or 74 gigabecquerels.

• Built-in dose calibration, quality-control and self-diagnostic functions.

• Automatic chamber quality control, including linearity testing.

• Automated linearity and geometry testing.

• Compatibility with nuclear medicine information management systems through USB.

• Capability to produce hard-copy records through the furnished printer.

The dose calibrator console shall have approximate dimensions and weight of:

• Height: 24 cm (9.5 inches).

• Width: 23 cm (9 inches).

• Length: 27 cm (10.5 inches).

• Weight: 3.4 kg (7.5 pounds).

Each ionization chamber shall meet the following requirements:

• Height: approximately 43.8 cm (17.25 inches).

• Diameter: approximately 17.2 cm (6.76 inches).

• Weight: approximately 13.6 kg (30 pounds).

• Well diameter: approximately 6.1 cm (2.4 inches).

• Well depth: approximately 25.4 cm (10 inches).

• Cable length: approximately 3.7 m (12 feet).

• Fill gas: 12-atmosphere ultra-pure argon.

• Technetium-99m measurement range of at least 6 curies or 222 gigabecquerels.

• F-18 measurement range of at least 2 curies or 74 gigabecquerels.

• Resolution of no more than 0.01 microcurie or 0.001 megabecquerel.

• Response time within 2 seconds for low-activity samples and 4 to 16 seconds for other measurements.

• Electrometer accuracy better than plus or minus 2%.

• Linearity within plus or minus 2%.

• Repeatability within plus or minus 1% during a 24-hour period when continuously powered.

• Daily auto-zero, background adjustment, voltage and data-check tests.

• Enhanced quality-control functions, including linearity, geometry and strip quality-control testing.

Each well counter shall have the following approximate dimensions:

• Height: 23.8 cm (9.38 inches).

• Diameter: 15.2 cm (6 inches).

• Weight: 6.9 kg (15.2 pounds).

• Well diameter: 1.7 cm (0.67 inch).

• Well depth: 3.8 cm (1.5 inches).

• Cable length: 2.7 m (9 feet).

Operation Conditions: The radiopharmaceutical calibrator is used to measure the radioactivity of radiopharmaceuticals and verify that each prepared dose contains the prescribed amount of activity at the intended time of administration. Personnel use the equipment during receipt, preparation, quality control, dispensing, and final verification of radiopharmaceutical doses to account for radioactive decay and confirm that the correct radionuclide and activity are available for the ordered diagnostic or therapeutic procedure.

The calibrator supports accurate patient-specific dosing by enabling staff to assay unit doses, syringes, vials, and other radiopharmaceutical containers before use. It provides documented measurement results needed to detect dosing discrepancies, prevent administration of incorrect activity levels, maintain radiation-safety controls, and meet applicable quality-assurance and regulatory requirements.