PERFORMANCE WORK STATEMENT (PWS)
DMEA 16-6H0
28 June 2016 Project Engineer: Jeremy Bridge, DMEA/MEBD, (916) 999-2728 Contracting Officer Representative: Allen Wong, DMEA/MEBD, (916) 231-1642
1.0 SCOPE:
1.1 TITLE: Temperature and Humidity Chamber Purchase and Install.
1.2 APPLICABILITY: DMEA/MEDB Test Reliability Program.
1.3 BACKGROUND: DMEA/MEBD has a need to acquire a Temperature and Humidity Chamber that capable of performing application specific integrated circuit (ASIC) testing in accordance with Mil-Std-883 Method 1005, Method 1015 , and Method 1004 in support of the reliability test program .
1.4 PURPOSE: The purpose of this PWS is to acquire and install a Temperature and Humidity Chamber with option from Espec North America, Inc. Identified in Paragraph 3.3.1, 3.3.2, and 3.3.3; and the technical capabilities and standards identified in Appendix A & B.
2.0 REFERENCE GOVERNMENT PROVIDED RESOURCES: N/A
2.5 GOVERNMENT FURNISHED FACILITIES/ACCESS:
| Facility/Access |
| Date |
| DMEA Facility, 4234 54th St., McClellan, CA 95652-2100 |
| As scheduled under the contract |
3.0 REQUIREMENTS:
3.1 GENERAL REQUIREMENTS:
3.1.1 RESERVED
3.3 TASK REQUIREMENTS:
3.3.1 MATERIAL REQUIRMENT AND SUPPORT: The contractor shall deliver a Temperature and Humidity Chamber and support as identified below:
| Item |
| QTY |
| Description |
| Part # |
| 1. |
| 1 |
| A factory calibration certificate, traceable to NIST, Espec Chamber with specifications identified in Appendix A |
| BTX-475 with window or Equal. |
| 2 |
| 1 |
| Options identified in Appendix B |
| 3. |
| 1 |
| Chamber shall be delivered to the address listed in paragraph 2.5of this PWS. |
3.3.2 USER TRAINING: The contractor shall provide equipment delivery for the Temperature and Humidity Chamber identified above. The contractor shall provide one training session (12hrs) for up to three people on the operation and maintenance of the system.
3.3.3 Maintenance: The contractor shall provide preventative maintenance functions at the time of setup. This includes inspection, setting and calibration of the machine, and any other tasks needed to properly maintain the machine. All preventative maintenance for the equipment shall be carried out by the contractor during normal DMEA business hours.
3.3.4 WARRANTY & SUPPORT: The contractor shall provide the following:
01. Extended twelve (12) month warranty from ship date including all repair parts and labor.
02. Response time of six (6) hours for any service related matter (via phone or email).
03. Onsite response time of less than twenty-four (24) hours for machine down service calls.
4.0 POINTS OF CONTACT:
DMEA/ MEBD/ Project Engineer
Name Jeremy Bridge
Address: 4234 54th Street
McClellan, CA 95652-2100
Voice: (916) 999-1671
E-Mail: jeremy.bridge@dmea.osd.mil
DMEA/ MEBD/ Contract Officer Representative
Name Allen Wong
Address: 4234 54th Street
McClellan, CA 95652-2100
Voice: (916) 231-1642
E-Mail: allen.wong@dmea.osd.mil
Appendix A (Specification Requirements)
1) High performance and quiet operation (65 dBA).
2) Space-saving footprint, designed for flush installation to wall.
3) Stainless steel exterior and durable thermoformed plastic door.
4) Full thermal break around doorframe.
5) One 2" and one 4” diameter cable port on left, with flexible silicone plug.
6) Three levels of overheat protection, plus overcool protection.
7) ETL listed electrical panel conforming to UL 508A.
8) Accurate wet/dry-bulb humidity measurement system that is easy to maintain.
9) Direct water supply connection for continuous operation.
10) Watlow F4T programmer/controller.
11) Web Controller remote chamber management.
12) Chamber Connect software for Watlow F4T.
13) 6"x12" Viewing window.
14) Adjustable wire shelf.
15) Cart with casters.
16) Low humidity control system.
17) Factory calibration certificate (traceable to NIST).
18) Printed & bound manual set.
19) Start-up & training for temperature/humidity models.
20) Cold Temperature minimum: less than or equal to -70oC
21) Hot Temperature Chamber Range maximum: greater than or equal to 180oC
22) Cool Down Change Rate: 2.5°C/m cool-down
23) Heat Up Change Rate: 2.5°C/m cool-down
24) Temperature Fluctuation: ± 0.5°C
25) Temperature Gradient: ± 1.5°C
26) Live Load Capacity: 300 W to -40°C
27) Humidity Range: 10 to 95% RH
28) Humidity Fluctuation: ±5% RH
29) Interior Volume: 4 cu. ft
30) Test area dimensions greater than or equal to (W×D×H): 19.6" x 15" x 23.6"
31) External dimensions less than or equal to (W×D×H): 30.25" x 33.5" x 46"
32) Heaters: 1000 W
33) Refrigeration: 3/4-hp x2 (cascade)
Appendix B
(Options)
Instrumentation:
Web Controller remote chamber management (in place of RS-232)
Chamber Connect software for Watlow F4T USB data acquisition to CSV for Watlow F4T Cabinet
Viewing window (6"x6" for 1.5 cu. ft., 6"x12" for 4 cu. ft.)
Cart with casters Cable port, 2" (50mm) diameter, with one flexible silicone plug
Cable port, 4" (100mm) diameter, with one flexible silicone plug Humidity Low humidity control system (red area shown on humidity chart
Other
Factory calibration certificate (traceable to NIST) Printed & bound manual set (in addition to standard CD-ROM version) Start-up & training for temperature/humidity models