Text version
Performance Work Statement
Annual service contract for FIAlab Inc., FIAlyzer 2000 flow injection analysis system
Work to be performed during the 12-month service contract:
· One preventative maintenance visit. Work to be performed during the visit: update FIAsoft software to the latest version; change all nuts, ferrules, tubing and back-pressure regulators; clean and inspect the Lab-On-Valve manifold; inspect valve head for leaks; change valve rotor;
· change pump tension spring as needed; clean flow cells and change windows and seals; inspect fibers for light throughput and damage; leave spare light source bulb as needed; perform system leak test; perform flow rate test; perform precision test.
· Provide e-mail technical support as needed.
· Provide telephone technical support as needed.
· Provide remote desktop support as needed.
· Provide additional on-site visits with a 72-hour response time at reduced fees.