FAA Technical Specification Sheet.pdf
PDF 136 KB Posted
- Attached to
- Market Survey Capability Assessment - Micro Gas Chromatograph for Hydrogen Leak Detection
- Solicitation number
- CT2600028A2
About this file
This document is an FAA Technical Specifications Sheet for procuring a micro gas chromatograph (Micro-GC) to detect hydrogen leaks in hydrogen-powered aircraft. The Hydrogen Safety Team within the FAA William J. Hughes Technical Center's Fuels and Energy Section (ANG-E283) requires a highly sensitive, accurate, and lightweight device meeting specific technical requirements.
The Micro-GC must have a hydrogen detection sensitivity below 10 parts per million, capability to measure hydrogen concentrations simultaneously at up to 20 different locations, and complete leak detection measurements in under three minutes. Additional specifications include a second channel for measuring carbon monoxide and nitrogen oxide concentrations, comprehensive operational and application training for FAA lab engineers, telephone technical support, and a one-year warranty covering parts, replacements, and software maintenance updates at no additional cost to the FAA.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
FAA Technical Specifications Sheet
Hydrogen leaks in hydrogen-powered aircraft pose a significant aviation safety concern. To address this issue, the Hydrogen Safety Team within the Fuels and Energy Section (ANG-E283) of the FAA
William J. Hughes Technical Center for Advanced Aerospace is seeking to purchase a highly sensitive, accurate, and lightweight micro gas chromatograph (Micro-GC) for detecting hydrogen leaks. Any interested entity must meet or exceed all of the EXACT minimum specifications listed below for the
Micro-GC to be used in aircraft applications:
1. Sensitivity: The Micro-GC must have high sensitivity for hydrogen detection, with a detection limit of less than 10 parts per million (ppm) level.
2. Multiple Point Measurements: As hydrogen leaks can result from multiple places and accumulate in various locations within hydrogen-powered aircraft, the Micro-GC must be capable of automatically measuring hydrogen leaks in two locations simultaneously. With automatic sampling valves, the Micro-GC must be able to determine hydrogen concentrations at least twenty (20) different locations.
3. Measurement Speed: Hydrogen leak detection using the Micro-GC must be completed in less than three (3) minutes per measurement to meet the safety requirements.
4. Impurity Measurement: Since impurity gases such as carbon monoxide (CO), nitrogen oxides (NOx) are harmful and others may be involved in hydrogen leak detection, the Micro-
GC must have a second channel for measurements of ppm concentrations of carbon monoxide CO) and nitrogen oxides.
5. Training: The product manufacturer must provide both operational and application training for FAA lab engineers. The manufacturer must provide telephone assistance for any product questions the FAA may have.
6. Warranty: A one-year warranty must be provided. The warranty must start immediately after installation and operation training have been completed. The warranty must provide for part repair or replacement and include software maintenance updates at no cost to the FAA.