This document is the Specifications for a Portable Computerized Radiography system required for the Arnold Air Force Base, TN Arnold Engineering Development Complex (AEDC). The key requirements include a Durr HD-CR35NDT Computed Radiography Scanner or equivalent, Durr D-TectX software or equivalent, transport case, imaging plates and sleeves, and installation, training, and warranty support. The system must meet various technical specifications related to resolution, grayscale, imaging plate size, weight, power, and operating temperature. This requirement is the subject of a Request for Quote (RFQ) under Solicitation Number FA9101-24-Q-B052, issued by the Department of the Air Force Materiel Command Test Center.

View the file

Other files for this federal contract opportunity

Other files attached to Portable Computerized Radiography, newest first.
File Type Posted
FA9101-24-Q-B052 - Portable Computerized Radiograhpy- Combined Synopsis Solicitation-Final.pdf PDF
FA910124QB052 - Model Contract.pdf PDF
Attachment 2 - FA9101-24-Q-B052 Redacted - single_source_justification.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications for Portable Computerized Radiography at

Arnold Air Force Base, TN Arnold Engineering Development Complex (AEDC)

Provide all materials, labor, and equipment necessary to provide a new fully functional Portable Computerized Radiography system. Refurbished or remanufactured parts are not authorized for this requirement.

This procurement is to add portable computerized radiography to the AEDC 300 KeV X-Ray generator and AEDC cobalt-60 and iridium-192 sources.

Requirements:

1) Durr HD-CR35NDT Computed Radiography Scanner or equivalent with the following requirements:

− Mechanical Iris focused laser with high resolution, high signal-to-noise ratio, and adjustable spot size between 12.5 μm to 50 μm − Rotational scanning method that maintains equidistant along a 14” by 17” imaging plate − Achieves a 30 μm basic special resolution in accordance with ASTM E2445, ASTM E2446, and ISO 16371-1 − 16-bit grayscale resolution − Compatible Image Plate Sizes: 2 to 35 cm width, 3 to 200 cm length − Maximum Dimensions (H x W x D): 16” x 15” x 19” − Maximum Weight: 40 lbs.

− Input Power: 120 VAC at 60 Hz with the ability to operate on battery − Operating Temperatures: 55°F to 90°F − Ethernet for PC Interface (ability to disable all wireless and still operate) − Standalone operation capable (can operate without a connected PC)

2) Durr D-TectX software or equivalent with the following features − Software must be installed from a physical CD/DVD with offline activation and licensing.

− Native DICONDE file format support and compatibility − GPU acceleration and parallel CPU computing − Unlimited image file size support − Ability to evaluate images while acquiring image data and acquire images from DDA and CR simultaneously − Filtering for image enhancement with image history for tracking all image operations − Ability to meet ASME Section V requirements for radiography using phosphor imaging plate − Included tools: histogram, SNR/SNRN calculation, Automatic SRb determination per ASTM E2446, wall thickness analysis, length measurements, annotations, line profiles − Support for DDA’s with or without upgrade − Report generation

3) Transport Case − Rugged case which holds the scanner, battery pack, and essential hardware − Total weight: less than 55 lbs.

− Includes retractable handle and wheels

4) Imaging Plates and Sleeves − Two (2) Phosphor Imaging Plates, HCR, 4.5” x 10”, 63 μm Maximum Basic Spatial Resolution and 3.3 mGy Minimum Radiation Dose − Two (2) Phosphor Imaging Plates, HCR, 4.5” x 17”, 63 μm Maximum Basic Spatial Resolution and 3.3 mGy Minimum Radiation Dose − One (1) Phosphor Imaging Plates, HCR, 14” x 17”, 63 μm Maximum Basic Spatial Resolution and 3.3 mGy Minimum Radiation Dose − One (1) Phosphor Imaging Plates, UH-IP, 10 cm x 24 cm, 30 μm Maximum Basic Spatial Resolution and 15.5 mGy Minimum Radiation Dose − One (1) Phosphor Imaging Plates, UH-IP, 24 cm x 30 cm, 30 μm Maximum Basic Spatial Resolution and 15.5 mGy Minimum Radiation Dose − Two (2) Phosphor Imaging Plates, UH-IP, 4.5” x 17”, 30 μm Maximum Basic Spatial Resolution and 15.5 mGy Minimum Radiation Dose − One Thousand (1000) 10 cm x 24 cm disposable light proof sleeves − Two (2) flexible film loading sleeves, 24 cm x 30 cm − Two (2) flexible film loading sleeves, 35 cm x 43 cm − Ten (10) boxes of ten (10) cleaning wipes for scanner

5) Installation, Training, and Warranty − Complete on-site installation of scanner and software.

− Scanner and Software Training within 30 days after software installation − Training on PMT Gain and laser power technique development − 5-year warranty for 100% parts and labor on scanner − Three (3) year guaranteed annual service by a factory authorized service center which includes a disassembly and cleaning of all internal components; rollers, drive belts, transport belts, check optical path, basic image test with CR Phantom.