DNA_Sequencing PWS_16NOV.docx
DOCX document 41 KB Posted
- Attached to
- DNA sequencing and bioinformatics services
- Solicitation number
- FA860127Q0026
About this file
This is a Performance Work Statement (PWS) for DNA sequencing and bioinformatics services to be provided to the Air Force Institute of Technology (AFIT) under a Blanket Purchase Agreement (BPA) effective November 2026. The Contractor shall provide comprehensive DNA sequencing and bioinformatics services for Government-furnished extracted environmental DNA samples, including sample intake, storage, pre-sequencing quality control, library preparation, sequencing, bioinformatics analysis, technical reporting, and electronic data delivery. The Government anticipates approximately three shipments annually (one every four months), with estimated quantities of 144 shotgun metagenomic sequencing samples, 9 whole-genome sequencing samples, and 9 targeted amplicon sequencing samples; however, these estimates are for planning and evaluation only and do not constitute a minimum guarantee. The Contractor must maintain capability to perform short-read and long-read DNA sequencing, targeted amplicon sequencing, shotgun metagenomic sequencing, and whole-genome sequencing, with a qualified technical lead identified within five business days of award.
The PWS establishes specific performance standards including baseline sequencing quality requirements (at least 80 percent of bases with Phred score Q30 or greater, cross-sample contamination not exceeding 0.1 percent, and method-specific minimum depths for amplicon, metagenomic, and whole-genome sequencing). The Contractor shall acknowledge receipt of each BPA call/order within two business days, acknowledge physical sample receipt within 48 hours, and deliver a complete electronic deliverable package within eight weeks of sample acceptance, including technical reports, demultiplexed raw sequence files, QC records, bioinformatics outputs, and documentation of all methods and software used. The Contractor must submit a Quality Control Plan within 10 business days after award for written acceptance before beginning sequencing work. For Contractor-attributable failures, the Contractor shall correct deficiencies or repeat affected work at no additional cost to the Government. The Government will inspect and accept or reject deliverables within 10 business days of receipt, with rejected deliverables requiring corrective-action plans within five business days.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment 1 - Instructions.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
PERFORMANCE WORK STATEMENT
DNA Sequencing and Bioinformatics Services Air Force Institute of Technology (AFIT) November 2026
This performance work statement establishes the recurring requirements, standards, and deliverables for DNA sequencing and bioinformatics services ordered under the Blanket Purchase Agreement (BPA). Individual BPA calls/orders establish the specific sample quantity, sequencing method, technical parameters, delivery schedule, funded amount, and any requirement-specific quality standards. The BPA does not obligate Government funds and does not guarantee any minimum quantity, expenditure, or number of purchases. The Government is obligated only for authorized purchases actually made under the BPA by personnel authorized in writing by the Contracting Officer.
1. General Information
1.1 Purpose
The Contractor shall provide DNA sequencing and bioinformatics services for Government-furnished extracted environmental DNA. Services include sample intake, storage, pre-sequencing quality control, library preparation, sequencing, bioinformatics, technical reporting, and electronic delivery of required data products.
1.2 Background
AFIT requires recurring DNA sequencing and bioinformatics services to support research and teaching activities associated with its educational mission. The Government anticipates approximately three shipments annually, normally one shipment every four months. This BPA does not guarantee a minimum quantity or dollar value.
1.3 Scope
The Contractor shall perform all services at its facility using Contractor-furnished personnel, equipment, facilities, supplies, reagents, and quality systems. The Government will provide extracted DNA samples and sample manifests as identified in each BPA call/order.
1.4 BPA Structure
This BPA establishes the terms governing recurring requirements. The Government may issue BPA calls/orders during the ordering period. The Government is obligated only to the extent of authorized calls/orders. The Contractor shall not begin sequencing, library preparation, or bioinformatics work until receipt of an authorized BPA call/order, except for actions necessary to confirm receipt and preserve sample integrity.
1.5 Definitions
| Term |
| Definition |
| BPA call/order |
| An authorized request under this BPA that identifies the specific work to be performed, funded amount, required delivery date, and applicable technical requirements. |
| Sample acceptance |
| Written or electronic confirmation that the sample shipment and associated manifest meet the minimum intake requirements or that identified discrepancies have been resolved by the Government. |
| Defective service |
| Any service output that fails to meet the applicable BPA, BPA call/order, or accepted Quality Control Plan requirement. |
| Government technical point of contact |
| The Government individual designated in the BPA call/order to address technical questions. This individual may not change price, schedule, scope, or other contractual terms. |
| QC |
| Quality control performed by the Contractor to verify that services and deliverables meet contractual requirements. |
2. Government-Furnished Items and Responsibilities
2.1 Government-Furnished Material
The Government will furnish extracted DNA from environmental samples, which may include drinking water, groundwater, and atmospheric water samples. DNA will normally be provided in nuclease-free tubes and suspended in Tris-HCl or TE buffer. The Government will provide a sample manifest for each shipment.
2.2 Estimated Quantities
The Government anticipates approximately three shipments annually, normally one shipment every four months. For solicitation price evaluation, Appendix B uses estimated quantities of 144 shotgun metagenomic sequencing samples, 9 whole-genome sequencing samples or isolates, and 9 targeted amplicon sequencing samples, with corresponding bioinformatics line items. These estimates are for planning and evaluation only; they do not constitute a minimum guarantee and may not represent unique physical samples.
2.3 Government Responsibilities
Issue authorized BPA calls/orders before sequencing or analysis begins.
Provide a complete sample manifest and required technical information for each BPA call/order.
Provide the sequencing method, required read configuration, sequencing depth or coverage, required controls, and required bioinformatics outputs when not established by this BPA.
Provide expected organism type and genome-size estimate for whole-genome sequencing, when applicable.
Review and accept or reject submitted deliverables within 10 business days after receipt.
Provide timely written direction when the Contractor reports an intake discrepancy, QC failure, or need for a Government decision.
3. Contractor-Furnished Items and Responsibilities The Contractor shall furnish all personnel, laboratory facilities, equipment, computing resources, software, licenses, reagents, consumables, shipping materials necessary for return shipments, and services required to perform the work unless a BPA call/order expressly states otherwise.
3.1 Technical Capability
Maintain capability to perform short-read and long-read DNA sequencing, as applicable to the BPA call/order.
Maintain capability to perform targeted amplicon sequencing, shotgun metagenomic sequencing, and whole-genome sequencing, as applicable to the BPA call/order.
Maintain sufficient laboratory and bioinformatics capacity to meet the required delivery schedule.
Provide a technical lead with demonstrated experience in environmental DNA sequencing and bioinformatics. The Contractor shall identify the technical lead in writing within five business days after BPA award and notify the Contracting Officer through the COR of any change before the change takes effect, when practicable.
3.2 Facility and Access Controls
The Contractor shall perform work in an access-controlled facility that limits access to authorized personnel. Contractor personnel are not expected to require access to a Government installation unless a BPA call/order states otherwise.
3.3 Contractor Communications
The Contractor shall acknowledge receipt of each BPA call/order within two business days. The Contractor shall acknowledge physical receipt of each sample shipment within 48 hours. The Contractor shall respond to routine Government technical inquiries within two business days unless a shorter response time is specified in a BPA call/order.
4. BPA Call/Order Requirements Only Government personnel authorized in writing by the Contracting Officer and acting within delegated authority may place calls under this BPA. Each call shall identify the BPA number, authorized purchaser, item or service, quantity, unit price, total amount, delivery/performance period, and delivery/performance location.
Each BPA call/order shall include the elements below. If an element is not applicable, the call/order shall state “not applicable.”
| Required element |
| Minimum content |
| Identification and funding |
| BPA call/order number, authorized purchaser, funded amount, and billing instructions. |
| Samples |
| Number of samples, sample identifiers, manifest version, expected DNA concentration/volume, and any known sample limitations. |
| Sequencing service |
| Amplicon, shotgun metagenomic, isolate whole-genome, short-read, long-read, or combination approach. |
| Technical specifications |
| Read configuration, target depth or coverage, primers/target region if applicable, expected genome size if applicable, and required controls. |
| Bioinformatics |
| Required analyses, reference database(s), pipeline(s), version(s), and reporting requirements, if Government standardization is required. |
| Schedule |
| Required receipt confirmation, initial QC report date, and final delivery date. |
| Price |
| Applicable BPA unit prices, quantity, and total call/order value. |
| Government technical POC |
| Name and contact information for technical questions and sample-disposition decisions. |
4.1 BPA Pricing Schedule
The BPA pricing schedule is provided as Attachment 1 – BPA Pricing Schedule and Price Evaluation Worksheet. Attachment 1 is incorporated into and made part of this BPA.
The Contractor shall provide firm-fixed unit prices for all Contract Line Item Numbers (CLINs) identified in Attachment 1. The Contractor shall not separately bill costs included in an applicable CLIN price.
Each BPA call/order shall identify the applicable Attachment 1 CLINs, quantities, unit prices, and total funded value. The Government will issue BPA calls/orders only as authorized and funded.
The estimated quantities in Attachment 1 are for solicitation evaluation and planning only and do not guarantee a minimum quantity or dollar value.
5. Specific Tasks and Performance Standards
5.1 Sample Receipt, Inspection, and Acceptance
Within 48 hours of physical receipt, the Contractor shall confirm receipt, inspect the shipment for damage or temperature-control failure, verify sample identifiers against the manifest, and document any discrepancy. If containers are damaged, leaking, mislabeled, or otherwise compromised, the Contractor shall segregate the affected sample(s), document the condition, and notify the Government technical POC and COR within one business day.
The Contractor shall perform initial sample QC and report DNA quantity, concentration, integrity or degradation assessment, and any material limitation within seven business days after sample acceptance. The BPA call/order may establish method-specific minimum input quantity, volume, purity, or integrity thresholds.
5.2 Sample Storage and Disposition
Before processing, the Contractor shall store received DNA samples at -30 degrees C to -15 degrees C, unless a BPA call/order specifies another temperature. The Contractor shall record storage location, date received, and date processed. The Contractor shall not discard, transfer, or use samples for any purpose outside the authorized BPA call/order.
The Contractor shall retain residual DNA for 30 calendar days after Government acceptance of the associated final deliverable, unless the BPA call/order states another period. After the retention period, the Contractor shall return or destroy residual samples only as directed by the Government. The Contractor shall document the return or destruction action.
5.3 Library Preparation and Sequencing
The Contractor shall use a next-generation sequencing platform and workflow appropriate to the method specified in the BPA call/order. The Contractor shall document the sequencing platform, library preparation method, reagent lot numbers, run identifiers, read configuration, and demultiplexing results.
The Contractor shall include positive and negative controls appropriate to the sequencing method. The specific controls and their frequency shall be identified in the BPA call/order or the accepted QCP.
5.4 Baseline Sequencing Quality Requirements
Unless a BPA call/order establishes a different method-specific standard, the following performance requirements apply:
At least 80 percent of bases per sample shall have a Phred score of Q30 or greater.
Cross-sample index hopping or contamination shall not exceed 0.1 percent, as measured by index-combination analysis or another documented equivalent method.
Unassigned reads after demultiplexing shall not exceed 5 percent of the sequencing run.
For targeted amplicon sequencing, the minimum depth shall be 50,000 read pairs per sample. Negative-control reads shall not exceed 1 percent of the lowest-yield sample, unless otherwise approved in writing by the Contracting Officer.
For shotgun metagenomic sequencing, the minimum depth shall be 10 million read pairs per sample. Negative-control reads shall not exceed 1 percent of the lowest-yield sample, unless otherwise approved in writing by the Contracting Officer.
For whole-genome sequencing, the minimum depth shall be 30x average coverage using the genome-size estimate identified in the BPA call/order. Contamination metrics shall be appropriate to the organism and analysis workflow. When CheckM is used for applicable microbial genome assemblies, contamination shall not exceed 5 percent.
For a required mock community, recovered composition shall be within plus or minus 20 percent of the expected composition, unless a BPA call/order establishes another method-specific threshold.
5.5 Bioinformatics
The Contractor shall perform the bioinformatics analyses identified in the BPA call/order. At a minimum, the Contractor shall perform read-quality assessment, contamination evaluation, and taxonomic assignment when taxonomy is required.
Document read trimming, filtering, host or non-target screening when applicable, and contamination-screening methods.
Identify all software, pipeline versions, database names and versions, parameter files, and reference files used for the analysis.
Provide taxonomic results at the resolution supported by the data and method, with relative abundance or read-count information, confidence metric where available, and unclassified-read rate.
For whole-genome sequencing, provide assembly statistics and the applicable assembler or completeness-tool QC report.
Provide a clear explanation of material limitations, ambiguous assignments, failed controls, or results that do not meet the applicable performance standard.
5.6 Technical Report and Data Delivery
The Contractor shall submit a complete electronic deliverable package no later than eight weeks after sample acceptance and receipt of an authorized BPA call/order, unless the BPA call/order establishes another schedule. The final delivery package shall include:
Technical report summarizing methods, results, QC outcomes, deviations, limitations, and required interpretation.
Demultiplexed raw sequence-read files in gzipped FASTQ format for all samples and controls; or assembled genomes in FASTA format for isolate sequencing, when specified.
FastQC or equivalent and MultiQC reports.
Reconciled sample manifest identifying all received, accepted, rejected, processed, and reported samples.
Library-preparation, sequencing, and QC records, including reagent lot numbers and instrument run identifiers.
Signed QC summary and any exception, deviation, corrective-action, or re-sequencing reports.
For amplicon sequencing, primer sequences, target region, and read configuration.
For whole-genome sequencing, assembly statistics and applicable completeness/contamination results.
Bioinformatics outputs, taxonomic tables, pipeline documentation, database versions, parameter files, and command logs sufficient to reproduce the analysis.
Checksum file for each delivered data package and a file-naming convention that links all files to the applicable sample identifiers and BPA call/order.
6. Quality Control Plan and Corrective Action
6.1 Quality Control Plan
The Contractor shall submit a Quality Control Plan (QCP) to the Contracting Officer through the COR within 10 business days after BPA award. The QCP shall be accepted in writing before the Contractor performs sequencing under the BPA, unless the Contracting Officer authorizes otherwise in writing.
At a minimum, the QCP shall address sample receipt and acceptance; pre-sequencing QC; control use; sequencing QC; bioinformatics QC; contamination control; nonconforming work; re-sequencing; data integrity; corrective action; records retention; and notification procedures.
6.2 Nonconforming Work
The Contractor shall notify the Government technical POC and COR within one business day after identifying a condition that may materially affect sample integrity, schedule, required sequencing depth, quality, or deliverable completeness. The notification shall identify the affected call/order, samples, issue, preliminary impact, proposed corrective action, and any Government decision needed.
For Contractor-attributable failures, including avoidable sample mishandling, sequencing-run failure, incorrect processing, or incomplete required deliverables, the Contractor shall correct the deficiency or repeat the affected work at no additional cost to the Government, unless the Contracting Officer determines in writing that the failure was caused by Government-furnished material or Government direction.
7. Data Handling, Delivery, and Records The Contractor shall use the Government-approved electronic transfer method identified in the BPA or BPA call/order. The Contractor shall not transmit Government data through an unapproved method. The Contractor shall retain raw data, intermediate analysis files, final deliverables, and associated QC records for at least 90 calendar days after Government acceptance, unless the BPA call/order specifies a different period.
The Contractor shall provide Government data only to personnel with a need to know for contract performance. The Contractor shall not use Government-furnished samples, data, or derived results for research, publication, training, marketing, benchmarking, or any other purpose outside the BPA call/order without prior written authorization from the Contracting Officer.
8. Inspection and Acceptance The Government will inspect each final delivery package for completeness, conformance with the applicable BPA call/order, required QC metrics, and required data formats. The Government will accept or reject the delivery in writing within 10 business days after receipt. If the Government rejects a delivery, the rejection notice will identify the deficiency. The Contractor shall submit a corrective-action plan within five business days and shall correct and resubmit the defective deliverable within the time approved by the Contracting Officer.
9. Quality Assurance Surveillance Plan and Performance Requirements Summary The Government will monitor Contractor performance using the following performance requirements summary. Surveillance may include review of receipt confirmations, QC reports, delivery packages, records, and corrective-action documentation.
| Performance objective |
| Standard |
| Surveillance method |
| Frequency |
| Required action for failure |
| Receipt confirmation |
| Confirm shipment receipt within 48 hours. |
| Review receipt notice and log. |
| Each shipment |
| Notify COR; correct process deficiency. |
| Initial sample QC |
| Provide initial QC report within 7 business days after sample acceptance. |
| Review QC report. |
| Each shipment |
| Provide corrective-action plan if late or incomplete. |
| Sequencing quality |
| Meet applicable baseline or call/order-specific quality thresholds. |
| Review QC and control results. |
| Each delivery |
| Re-sequence or correct as required. |
| Data package completeness |
| Provide all required files, records, checksums, and report elements. |
| Government inspection. |
| Each delivery |
| Correct and resubmit. |
| Final delivery schedule |
| Deliver within eight weeks after sample acceptance unless call/order states otherwise. |
| Compare receipt and delivery dates. |
| Each call/order |
| Provide corrective-action plan. |
| Material risk notification |
| Notify Government within 1 business day. |
| Review notice. |
| As needed |
| Correct notification process. |
Appendix A. BPA Call/Order Template The Government will use the following format, or an equivalent authorized format, for each BPA call/order.
| Field |
| Required entry |
| BPA number / call-order number |
| [Insert] |
| Date issued / authorized purchaser |
| [Insert] |
| Funded amount / billing instructions |
| [Insert] |
| Government technical POC |
| [Name, title, phone, email] |
| Sample shipment |
| [Sample count, identifiers, manifest version, anticipated ship date] |
| Sequencing method |
| [Amplicon / shotgun metagenomic / whole-genome / short-read / long-read] |
| Read configuration and depth |
| [Insert] |
| Primers and target region |
| [Insert or N/A] |
| Expected genome size and organism type |
| [Insert or N/A] |
| Controls and method-specific QC |
| [Insert] |
| Bioinformatics and reference database requirements |
| [Insert] |
| Data package and delivery method |
| [Insert] |
| Required final-delivery date |
| [Insert] |
| Applicable Attachment 1 CLINs, quantities, unit prices, and total funded value |
| [Insert] |
| Special instructions or deviations approved by Contracting Officer |
| [Insert] |
Department of War | Air Force Institute of Technology
File details come from the government source that posted it. Updated .