Attachment 2. SOW - CZT RadioXenonDetector 10 Nov 2026.pdf

PDF 185 KB Posted

Attached to
Custom CdZnTe X-Ray and Gamma-Ray Detector Prototype for Beta-Gamma Radioxenon Education and Research.
Solicitation number
FA860127Q0006
Issued by

About this file

This is a Statement of Work (SOW) for the design, fabrication, characterization, and delivery of one custom pixelated cadmium zinc telluride (CdZnTe) x-ray and gamma-ray detector prototype in a well or top-hat geometry that fits over a Government-furnished beta gas cell for the Air Force Institute of Technology (AFIT).

The detector prototype shall operate over a photon energy range of 25 keV to 525 keV and must meet specific technical requirements. The detector design shall use pixelated CdZnTe detection modules arranged radially around the radioxenon sample volume, providing no less than 18 square centimeters of detection area and 18 cubic centimeters of detection volume. Energy resolution requirements are specified in Table 1 and include a full width at half maximum (FWHM) no greater than 10% at 31 keV, 3.0% at 81 keV, 2.0% at 122 keV, and 1.2% at 356 keV, with a non-binding objective of 4.5% resolution at approximately 30 keV. Intrinsic photon detection efficiency minimums are also specified at these same energies, ranging from 9% to 45%. Additional requirements include demonstrating separation of Ba-133 K-alpha and K-beta x-ray groups near 31 keV and 35 keV, providing coincidence timing interface capability, exporting data in list mode and .n42 format, and delivering comprehensive software documentation and an API or SDK for detector control and data acquisition. The contractor shall provide heritage documentation demonstrating the CdZnTe detector technologies have been environmentally tested per ANSI N42.34 standards, deliver a detailed characterization package with measured performance data at specified energies using Ba-133, Co-57, and Cs-137 sources, provide one year or more original manufacturer warranty, and deliver on-site training by a factory-certified technician for 8 hours covering up to 4 faculty and staff members. Delivery is required within 120 days after contract award to AFIT/ENP, 2950 Hobson Way, Wright Patterson AFB, Ohio 45433-7765, unless the contractor is within a 225-mile radius, in which case AFIT will arrange pickup. The contractor must include all shipping, handling, broker fees, and tariffs in the quote. CMMC Level 1 (Unclassified and Non-Sensitive) applies to this procurement.

View the file

Other files

File Type Posted
ATTACHMENT 1 - Instructions.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Statement of Work

For

Custom CdZnTe X-Ray and Gamma-Ray Detector Prototype for Beta-Gamma Radioxenon Education and Research

Title: Custom CdZnTe X-Ray and Gamma-Ray Detector Prototype for Beta-Gamma Radioxenon Education and Research

Document Date: 10 Nov 2026

Scope: This Statement of Work (SOW) defines the requirements for the design, fabrication, characterization, and delivery of one (1) custom pixelated cadmium zinc telluride (CdZnTe) x-ray and gamma-ray detector prototype in a well or top-hat geometry that fits over a Government-furnished beta gas cell. The deliverable includes the detector modules, cooling, spectroscopy electronics, control and acquisition capability, coincidence timing output, data-export capability, calibration and characterization data, documentation, and required interconnects. The beta gas cell, its electronics, radioxenon samples, and laboratory integration remain Government responsibilities.

Background: The Air Force Institute of Technology (AFIT) conducts graduate education and research in the measurement of radioactive xenon (radioxenon) by beta-gamma coincidence counting, the technique used by the International Monitoring System (IMS) of the Comprehensive Nuclear-Test-Ban Treaty Organization (CTBTO) to detect and attribute underground nuclear tests. A concealed underground test can suppress its seismic signature, but radioxenon escaping the test cavity may still reach a monitoring station; in the most difficult cases it is the principal observable evidence that a test occurred. Distinguishing that signal from civilian reactor and medical-isotope emissions rests on activity ratios of four xenon isotopes measured at very low activity.

In the currently fielded architecture, a plastic scintillator senses beta particles and conversion electrons while a sodium iodide (NaI) detector senses the coincident x-rays and gamma rays.

NaI energy resolution near 30 kiloelectronvolts (keV), roughly 20 percent, cannot separate the closely spaced xenon and cesium x-rays in that region, which blurs the metastable-isotope signatures on which the four-isotope determination depends. The prototype specified here replaces the gamma-detection side of the architecture over the Government-furnished beta gas cell so that AFIT can evaluate whether resolving the 30 keV region rejects background and lowers the minimum detectable concentration (MDC).

The prototype is equally an educational platform in direct support of Department of Defense workforce development. It gives nuclear engineering and applied physics graduate students hands-on experience with semiconductor detector physics, pixelated charge collection, energy calibration, coincidence timing, and the data products of treaty monitoring, and it prepares officers who routinely proceed to operational assignments employing this technology. One quarter of USAF nuclear engineering graduates go on to use the technologies during operational assignments. The capability will support thesis research and laboratory instruction in NENG 612, NENG 650, NENG 680, NENG 791, NENG 799, NENG 812, NENG 880, and NENG 999.

Requirements/Salient Characteristics: The contractor shall design, fabricate, test, and deliver one (1) custom CdZnTe-based x-ray and gamma-ray detector prototype for beta-gamma radioxenon measurement research. The detector shall operate over a photon energy range of 25 keV to 525 keV.

Detector Design.

1. The detector design shall be a pixelated CdZnTe detection modules.

2. The contractor shall arrange the CdZnTe crystals radially around the radioxenon sample volume in reproducible close proximity to the beta-sensitive detection region.

3. The detector shall provide no less than 18 square centimeters of detection area and no less than 18 cubic centimeters of detection volume.

4. The detector shall fit over a plastic beta gas cell that is approximately a cylinder of radius

1 centimeter and height 3 centimeters. The Government will provide the gas cell envelope dimensions after contract award.

Energy Resolution.

• The detector shall achieve a full width at half maximum (FWHM) energy resolution no greater than the values in Table 1 at the indicated photon energies. As a non-binding objective, resolution of 4.5 percent at approximately 30 keV is desired in order to separate the xenon K-alpha x-ray doublets cleanly.

Intrinsic Efficiency.

• The detector shall achieve an intrinsic photon detection efficiency no lower than the values in Table 1 at the indicated photon energies.

Table 1. Required energy resolution and intrinsic efficiency.

Photon energy and test source

Energy resolution (FWHM), maximum

Intrinsic efficiency, minimum

31 keV (Ba-133) 10% 15%

81 keV (Ba-133) 3.0% 45%

122 keV (Co-57) 2.0% 45%

356 keV (Ba-133) 1.2% 9%

Low-Energy Separation.

• The detector shall demonstrate separation, or quantitative fitting, of the Ba-133 K-alpha x-ray group near approximately 31 keV and the K-beta x-ray group near approximately 35 keV.

Coincidence Timing Interface.

• The detector shall provide event timing information and a hardware timing interface sufficient to support coincidence detection during list-mode acquisition.

Data Export.

• The detector shall export measurement data in both list mode and .n42 format.

Exported data shall include all applicable event, timing, detector-channel, energy or pulse-height, metadata, status, and data-quality information.

Software and Interface Documentation.

• The contractor shall provide software, a documented API, or an SDK that enables a Government-furnished computer to configure and control the detector, acquire measurement data, monitor detector status, and export collected data. The contractor shall provide sufficient documentation for Government operation and software integration.

Environmental Testing.

• The contractor shall provide evidence, in the form of test reports or test summaries, that the CdZnTe detector technologies on which the prototype design is based have been environmentally tested in accordance with ANSI N42.34 test methods. For the purpose of this task heritage documentation for similar fielded devices will meet the requirement. This section does not require new environmental testing of the delivered prototype.

Characterization Package.

• The contractor shall report measured centroid, FWHM energy resolution, fitted or measured peak area, acquisition conditions, and detection efficiency or relative efficiency, where statistically supportable, at the following energies and energy regions: the K-alpha x-ray group near 31 keV and the K-beta x-ray group near 35 keV, and 81, 276, 303, and 356 keV, using barium-133; 122 keV, using cobalt-57;

and the Compton continuum of cesium-137. The contractor shall deliver the underlying exported detector data with the characterization report.

Deliverables and Schedule.

1. The contractor shall deliver one (1) detector prototype meeting paragraphs 3.1 through 3.7, the performance demonstration data and analysis of paragraph 3.9 through 3.10, and the operating and data-format documentation of paragraph 3.8 required to operate the detector and export its data.

2. The contractor shall deliver all items no later than twelve (12) months after contract award.

3. Delivery is not required if contractor is within a 225-mile radius from delivery address specified below. AFIT will pick up items if adequate packing is provided to secure safe transport. Delivery is required if contractor is outside of 225-mile radius from delivery address.

4. The Manufacturer must include all shipping and handling and Broker Fees and Tariffs in the Quote.

PERIOD OF PERFORMANCE

• The contractor shall deliver within 120 days after contract award.

• Delivery Address – AFIT / ENP 2950 Hobson Way Wright Patterson AFB, Ohio

45433-7765

Warranty.

• The contractor shall provide an original manufacturer warranty for one (1) year or more.

Training.

• The contractor shall provide on-site training by a factory-certified technician for 8 hours to train up to 4 Faculty and Staff.

CMMC Level

• The CMMC Lever for this procurement is level 1(Unclassified and Non-Sensitive)

Applicable Documents:

1. The following documents are invoked in Section 3 to the extent stated there.

2. MIL-STD-810H, Department of Defense Test Method Standard: Environmental

Engineering Considerations and Laboratory Test Methods (invoked by paragraph 3.10 for heritage test evidence only).

File details come from the government source that posted it. Updated .