This document is a vendor questions and government answers attachment for Solicitation FA860125Q0130, dated 03 July 2025, specifically addressing a Fuel Flashpoint Tester procurement. The document contains a vendor inquiry about the ERAFLASH X unit, which runs method D7094 and offers excellent correlation to the traditional D93 method with improved safety features. The government responded affirmatively, confirming that D7094 is an acceptable correlated method that meets D93 standards, thereby providing technical clarification for potential vendors preparing their quotations for the solicitation.

View the file

Other files

File Type Posted
Atch 3 - TO 42B-1-1 8.1& 8.2.pdf PDF
Atch 1 - Solicitation FA860125Q0130.pdf PDF
Atch 2 - Statement of Work.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

03 July 2025

Solicitation FA860125Q0130 – Fuel Flashpoint Tester

1. Vendor Question: We have a unit called the ERAFLASH X that runs method D7094, which offers excellent correlation to D93, without any of the fire concerns you might have with traditional D93. I attached a brochure for the ERAFLASH X and a small paper on the ERAFLASH explaining how D7094 provides excellent correlation to D93, but is faster and safer. Let me know if you will accept correlated methods.

Government Answer: After reviewing D7094 is an acceptable correlated method that meets D93 standards.