This document is a Question and Answer (Q&A) response related to a solicitation for a Reverse Osmosis System for the Wright-Patterson Medical Center (WPMC).
The key details are:
- The solicitation (FA8601-24-Q-0101) is for an entire reverse osmosis system, including softening equipment, carbon filtration, and deionization tanks, to replace the existing system.
- The required specifications include a 185-gallon tank and 70 GPM reverse osmosis unit to support increased surgical capabilities.
- The system will be used for sterile processing and must comply with ANSI/AAMI ST108:2023 standards.
- The contract will be a firm-fixed-price commercial item procurement evaluated on technical capability and price.
- Quotes are due by August 15, 2024 and should include detailed equipment descriptions, individual and total pricing, delivery schedule, and other required information.
- The solicitation does not require the manufacturer to be licensed in Ohio, and a site visit is not being offered prior to award.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| RO Pic 2.jpg | JPG image | |
| RO Pic 1.jpg | JPG image | |
| Reverse Osmosis Q and A_7.31.2024.pdf | ||
| Solicitation - FA860124Q0101.pdf | ||
| Reverse Osmosis Statement of Work _6.21.24.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
DEPARTMENT OF THE AIR FORCE
BUSINESS AND ENTERPRISE SYSTEMS DIRECTORATE
WRIGHT PATTERSON AIR FORCE BASE, OHIO
31 July 2024
FROM: AFLCMC/PZIBA
1940 Allbrook Drive, Room 109
Wright-Patterson AFB OH 45433
SUBJECT: Response to Inquiries Regarding FA8601-24-Q-0101
QUESTIONS
QUESTION: Are you requesting replacing just the RO unit or the complete system? Why are you asking for a 185 gallon tank? The minimum requirements calls for a 185 gallon tank, but a
70 gallon per minute RO unit.
ANSWER: We are replacing the entire system. The below specifications are required for the department’s sterilization equipment that the reverse osmosis water flows through. The 185
Gallon Tank and 70GPM are the requirements per the SOW as our capabilities increase with our manning. As we plan to increase how many surgeries we do a day, those numbers will increase.
The 185 Gallon tank and 70 GPM serve as a blanket coverall for a projected increase in capes.
QUESTION: Does the RO water feed the hot water supply?
ANSWER: The RO does not feed the hot water supply.
QUESTION: Under the standards, guidelines or approvals, it mentions “for devices that will contact the bloodstream” is this system for Sterile processing or something else?
ANSWER: For Sterile Processing.
QUESTION: One of the guidance documents referenced in the Scope of Work AAMI
TIR34:2014/(R)2017 has been superseded by ANSI/AAMI ST108:2023. Please confirm that
WPMC Sterile Processing wants to use TIR34.
ANSWER: We are using ANSI/AAMI ST108:2023.
QUESTION: Is a walk-through scheduled?
ANSWER: A walk-through is not being offered at this time. The awardee may request a walk-through following contract award.
QUESTION: Can photos of the existing Reverse Osmosis System be provided?
ANSWER: See RO Pic 1 and RO Pic 2 attachments.
QUESTION: I have a question about line item 0001 (Reverse Osmosis System in accordance with Statement of Work) in solicitation FA860124Q0101. Does this price include the removal of the old system?
ANSWER: The SOW does include removal of the old system as a requirement. Installation and maintenance are also listed as requirements in the SOW. The cost of these services may all be combined into CLIN 0001 so long as it is clear that the requirements are being met, or quotes may be submitted that break these services out from the price of the system itself.
QUESTION: Can you please advise who was the supplier of the current reverse osmosis system, and how much was the previous contract worth.
ANSWER: Wright Patterson Medical Center’s current Reverse Osmosis System was manufactured by Siemens Healthcare. The current system was not considered as accountable equipment until 2019. As such, the Medical Group is unsure when the unit was purchased or by whom as there is no historical data in the record system.
QUESTION: Is this solicitation simply for a reverse osmosis unit or would you be looking to also obtain and have installed softening equipment, carbon filtration, and deionization tanks?
ANSWER: Yes, we are also looking to obtain and have installed softening equipment, carbon filtration and deionization tanks.
QUESTION: Does the manufacturer need to be licensed in Ohio? If so, can that licensing be sought after the contract is awarded?
ANSWER: Manufacturer does not need to be licensed in Ohio.
Tracy L. Ferrise
Contracting Officer