This document is a Statement of Work (SOW) for a Fast Digital Oscilloscope to be procured by the Air Force Institute of Technology (AFIT). The key requirements include:
- Oscilloscope with a minimum 2 GHz bandwidth, 10 Gs/s sampling rate, and 4 analog channels
- Capability to measure 8 logic signals simultaneously with the analog inputs
- Minimum 62.5M points record length
- 15.6-inch HD color touchscreen display
- Advanced triggering options and connectivity (USB, Ethernet, Display Port)
- Delivery within 8 weeks of contract award to AFIT in Ohio
The related federal contract opportunity is a Request for Quote (RFQ) under the North American Industry Classification System (NAICS) code 334515 with a Small Business size standard of 750 employees. Quotes are due by May 2, 2024 at 3:00 PM EST. Award will be made on a lowest price technically acceptable (LPTA) basis to the responsible offeror whose quote conforms to the SOW requirements.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Solicitation - FA860124Q0089.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work for Fast Digital Oscilloscope Document Date: 13 Mar 2024
1. Scope: The measurement of short-lived nuclear events requires instrumentation with fast acquisition rates. Oscilloscopes currently available in the teaching and research laboratories in B470 do not offer the necessary bandwidth or sampling rate to capture the key characteristics associated with the output pulses of fast scintillation detectors such as BaF2, thus limiting our experimental and troubleshooting capabilities.
1.1 Background: The instrumentation in AFIT/ENP’s Nuclear Spectroscopy Laboratory is insufficient to measure key characteristics of our fast detector output pulses. These detectors produce waveforms with rise times of approximately 1 ns with corresponding transients of tens of ps, and this drives the requirement for oscilloscope bandwidth; substantial attenuation of the signal will result when measured with an instrument with a bandwidth less than 2 GHz. The minimum recommended sampling rate is thus 10 Gs/s, with higher sampling rates A fast oscilloscope will serve ENP’s current and future students and faculty across a wide range of teaching and research activities.
2. Applicable Documents
N/A
3. Requirements/Salient Characteristics:
Fast Oscilloscope- quantity: 1
· Bandwidth ≥ 2 GHz (Note: This minimum bandwidth must be available on each channel during synchronous use.)
· Sampling rate ≥ 10 Gs/s
· Number of channels ≥ 4 Analog
· ≥ 4 Analog Channels inputs
· Digital Channels: Up to 32 (optional): Each channel enables eight digital logic inputs with TLP058 logic probe
· Capability to measure ≥ 8 logic signals
· Ability to measure 8 logic signals simultaneously with analog inputs on other channels
· channel enables eight digital logic inputs with TLP058 logic probe
· Waveform capture rate: >500,000 waveforms/s
· Signal Trigger types: Edge, Pulse Width, Runt, Timeout, Window, Logic, Setup & Hold, Rise/Fall Time, Parallel Bus, Sequence, Visual Trigger, Video
· Display: Minimum 15.6-inch (396 mm) High Definition (1,920 x 1,080) color
· Capacitive touchscreen (Pinch/Zoom/Swipe/Drag)
· Record Length Minimum 62.5M points record length
· Connectivity: USB Host (7 ports) or USB 3.2 Gen 2x2 for both the device and host ports, USB 3.0 Device (1 port) or USBC connectors, minimum LAN (10/100/1000 Base-T Ethernet), Display Port, DVI-I, VGA or HDMI
· Basic waveform arithmetic, FFT, and advanced equation editor; option for arbitrary waveform generator (AWG)
· Passive 10X probes for each channel
· Minimum 1 year warranty
4. Delivery/Period of Performance: Equipment will be shipped to WPAFB. Customer pays shipping, import/export fees and taxes. The customer—the Air Force Institute of Technology (AFIT)—will be 8 weeks after contract award.
4.1 Place of Performance: This is an F.o.b. destination requirement.
Air Force Institute of Technology (AFIT/ENP)
ATTN: TBD
2950 Hobson Way Wright-Patterson AFB, OH 45433-7765