SPECIFICATION FOR
Servo-Hydraulic Material Test Frame Controller System Upgrade
1. PURPOSE
· Application: This specification is intended to define the hardware, software, data acquisition and control requirements for multiple Servo-Hydraulic Material Test Frame Controller System Upgrades. System use includes static, fatigue, block or random spectum fatigue, multi-step ramping.
· Objectives: This specification is intended to ensure the following characteristics of the Servo-Hydraulic Material Test Frame Controller System Upgrade:
· Minimum Mission Impact
· Minimum Facility Configuration Impact
· Capability
· Safety
· Maintainability
· Minimum Purchase and Operational Costs
2. EXISTING SYSTEMS DESCRIPTION
· Controller #1 MTS FlexTest GT, 4 channel with PC Per Station Control
· Qty 2 11kip (50kN) capacity MTS Landmark MN: 370.10
· Qty 2 22kip (100kN) capacity MTS Landmark MN: 370.10
· Controller #2 MTS FlexTest GT, 4 channel with PC Per Station Control
· Qty 1 5kip (25kN) capacity Local manufacture with MTS load cell, and actuator with integrated LVDT.
· Qty 1 220kip (1000kN) capacity MTS MN: 311.31
· Qty 1 500kip (2272kN) capacity MTS MN: 311.41
· Qty 1 55kip (250kN) capacity MTS MN: 318.25
· Controller #3 MTS FlexTest GT, 3 channel with PC Per Station Control
· Qty 2 110kip (500kN) capacity, MTS 810 MN: MTS 380.50
· Qty 1 55kip (250kN) capacity MTS Landmark MN: 370.50
· Controller #4 MTS FlexTest GT, 3 channel
· Qty 3 11kip (50kN) capacity Shore Western Actuators
· Integrated LVDTs
· Interface Load Cells
· Mounting: Shore Western Swivel Base
· Controller #5 MTS TestStar IIs
· Qty 1 22kip (100kN) capacity MN: MTS 312.31
· Controller #6 MTS TestStar IIs
· Qty 1 5.5kip (25kN) capacity MTS 858 Tabletop MN: 359 L/F
· Load Cell 1Kip capacity
· Existing Test Frame Computer System (Qty. 14)
· INTEL CORE i7-4770 @3.4GHz
· MEMORY, 8GB (2 x 4GB), 240-PIN, DDR3 SDRAM
· 1TB 7200 RPM 64MB CACHE SATA INTERNAL HARD DRIVE
· ASUS MAXIMUS VI HERO LGA INTEL MOTHERBOARD
· PNY GeFORCE GT 610 1GB 64-BIT DDR3 PCI EXPRESS VIDEO CARD
· INTEL 10/100/1000Mbps PCI-EXPRESS NETWORK ADAPTER
· Alignment System
· Model 709.20D-01 Alignment Data Acquisition and Conditioning System
· Model 709.20D-01 Alignment Data Acquisition and Conditioning System
· Model 609 Alignment Fixtures are only installed on test frames with capacities from 5.5kip through 55kip.
· Software
· Current control/data acquisition software and optional components includes the following:
· MTS FlexTest GT 793 basic software
· MTS 793.10 Multi-Purpose Testware (MPT)
· MTS 793.04 PC per Station (MPC)
· MTS 793.11 Profile Editor
· MTS 793.08 Peak-Valley Phase Control (PVP)
· MTS 793.07 Arbitrary End –Level Control (ALC)
· MTS 793.03&02 Calculations with Outputs & Inputs
3. UPGRADE CONTROLLER ATTRIBUTES
· The existing 6 controllers shall be replaced with than no less than 4 multi-channel controllers. The ideal configuration based upon existing facility configuration is represented below.
· #1 – 4 channels
· #2 – 4 channels
· #3 – 4 channels
· #4 – 4 channels
· The existing wiring to the 4 controller locations shall be used to the greatest extent possible.
· The 2 test frames with MTS TestStar IIs controllers are not currently cabled to the “Ideal” configuration. The 2 test frames required cabling lengths are approximately 25ft (8m).
· Each channel shall be operated from individual computer stations (existing and provided)
· Handset remote controls shall be provided for each channel/test frame.
· The exception of Qty. 3 11kip (50KN) capacity Shore Western Actuators, has only one operator computer station but three handset controls.
· Handset functionality shall include; zeroing/offset signal channels, start/stop tests, hydraulics on/off control, actuator position adjustment.
· Control loop frequency shall be a minimum of 4kHz, 16bit
· Each test frame/channel must have dedicated sensor I/O of load, position, and extensometer (strain)
· Each test frame/channel must have dedicated hydraulic servo and hydraulic service manifold controls.
· Additional Integrated Data Acquisition
· Each controller shall have a minimum of 64 channels distributed among the test frames/channels on each idealized controller. Equivalently, each test frame shall have no less than 16 channels/test frame.
· Sampling rate range shall be at a minimum 1Hz to 6kHz
· Multiple sampling rates shall be definable per test frame.
· The 64 channels/controller or 16 channels /test frame, shall be capable of ¼, ½, full bridge, high level voltage inputs ±10 volts. Thermocouple inputs are desired but not required.
· ¼ bridge completions shall accommodate 120OHM, 350OHM, and user defined.
· Excitation, Filtering, Bridge type, Bridge completion, limit detections programmable on a per channel basis
· Shall be clock syncing with controller
· Alignment System shall follow and adhere to ASTM E1012
· The system shall be integrated with controllers for force control and/or feedback.
· The system shall be portable to each test frame/channel.
· The system shall be compatible with existing MTS 609 Alignment Fixtures
· The system shall be compatible with 8 and 12 gage alignment specimens.
· An instrumented alignment specimen shall be provided.
· Specimen shall be a flat, rectangular cross section rated to a safe stress between 10 KSI and 60 KSI
· Specimen shall have a minimum of 8 strain gages
· Video Extensometery system shall adhere to ASTM E83
· Shall be integrated with controller and provided software.
· Shall follow and adhere to ASTM E83 with a minimum B-2 classification
· Measurements shall include as a minimum, displacement, axial strain & transverse strain (Possion’s ratio), shear strain and rotation.
· Calibration devices processes shall be provided for both axial and transverse directions.
· Shall have necessary lenses to accommodate Working Distances (WD) from 100mm to 500mm
· Shall have necessary lenses to accommodate Fields of Views (FOV) from 50mm to 400mm
4. SOFTWARE ATTRIBUTES
· Software must be Windows 10 compliant.
· Software must run on existing computers. See Section 2 Exising System Description
· Software shall be Commercial Off The Shelf (COTS) with technical and update support.
· The basic core software shall operate the test frames and collect data from the test frames.
· The basic core software and additionally proposed software shall be equivalent and/or more capable than existing controller software.
· The reuse of existing software is desirable as it has the lowest training requirement and impact on mission.
· Any software proposed not in the Existing System Description shall be accompanied with training. See section 8 Installation and Training.
· Preferred software set-up shall follow a graphical test flow.
· Data can be displayed from customizable reports to raw ASCII text.
5. ACCESSORIES
· Provisions shall be provided to mount controllers into standard 19” rack consoles.
· Necessary cabling and components shall be provided.
6. AVAILABILE POWER AND UTILITIES
· 115VAC, 20A (unconditioned)
· 115VAC, 20A (conditioned)
· Compressed air (100psi max)
7. WARRANTY
· Minimum 1 year coverage on parts and workmanship.
· Minimum 1 year of software maintenance, to include upgrades across all packages and optional software components
· Phone support
· On-site support including parts, labor, and travel
8. INSTALLATION AND TRAINING
· Installation
· Vendor technician/representative shall install setup, verify operation and provide NIST traceable calibration of controllers with their respective test frames and components.
· All travel and living expenses shall be included.
· Training
· Training shall be provided in a laboratory/classroom environment at a vendor site for up to 6 personnel.
· Travel and living expenses will be at the expense of the government.