This document is a Memorandum for Record describing the Item Description for a Portable X-Ray System. The Air Force is requesting the purchase of four (4) Portable X-Ray Units with detailed technical specifications outlined, including input voltage requirements, x-ray output, radiation output/leakage, focal spot size, temperature and duty cycle ratings, dimensions, and weight. The related federal contract opportunity is Solicitation Number FA812524Q0034 for the procurement of the four (4) Portable X-Ray Systems, which is a 100% small business set-aside. The solicitation includes instructions for responding, such as completing company information, signing, providing pricing and delivery information, and addressing technical evaluation criteria. The FOB destination is Tinker Air Force Base in Oklahoma.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Solicitation Amendment FA812524Q00340001 SF 30.pdf | ||
| Solicitation - FA812524Q0034.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
DEPARTMENT OF THE AIR FORCE
OKLAHOMA CITY AIR LOGISTICS COMPLEX
TINKER AIR FORCE BASE OKLAHOMA
28 March 2024
MEMORANDUM FOR RECORD
SUBJECT: Item Description A23-100 Portable X-Ray System
Reference: AIF, Dated 1 February 2024
1. The 76th AMXG is requesting utilization of funds for the purchase of Four (4) Portable X-ray Units.
2. The requested equipment item descriptions/specifications are outlined below.
SPECIFICATIONS
Input Line Requirements: Automatically adapts to input line voltage100‐130Vac, 50/60Hz, 20 Amperes maximum 200‐250Vac, 50/60Hz, 10 Amperes maximum. May also be portable‐ generator powered X‐Ray Output 10 to 160kV, 0.5mA to 5.0mA (800 watts max.) Constant potential, end‐grounded anode, air cooled versions available.
X‐Ray Tube Window: Beryllium 0.8mm (directional) 40° directional
Radiation Output: 14R/min at 50cm filtered with 0.5 inches (12.7mm) aluminum at 160kV, 5mA
Radiation Leakage: Less than 2.0 Roentgens per hour at 1 meter from the X‐Ray tube target
Effective Focal Spot: Standard: 0.7mm x 1.0mm (0.03in x 0.04in) EN12543 0.5mm x 0.5mm (0.02in x 0.02in) nominal IEC60336in
Ambient Temp: 100% duty cycle @ 120°F (49°C)
Duty Cycle: 100% ‐ air‐cooled
Storage Temp: ‐30°F to 160°F (‐35°C to 71°C)
Anode Cooling: Fan forced air cooling
Tube Head Dimensions: No greater than 7.25" Diameter, 30" Length
Air Cooled: 33lbs. (14.97kg)
X‐Ray Control Unit: Digital microcomputer based with local control, RS‐232 and Ethernet
X‐Ray Control Unit Weight: 26lbs. (11.8kg) approx.
Safety Devices:
• Tubehead Pressure Relief Valve
• Tubehead Thermal Cut‐Out
• Tubehead Pressure Gauge
• Tubehead Low Pressure Cut‐Out @ 25psi (1.72 Bar)
• Control Unit Safety Keyswitch
• Microcomputer‐based Self Diagnostics
• Continuous Exposure Parameter Display
Standard Accessories:
• Operation manual
• Tubehead carrying case
• Tubehead Cable‐ 100’ with strain relief
• Extra key (1) for Control Unit Safety Lock
• Power cable ‐ 25’ with strain relief
GUI Control Software for the GUI is specifically designed for controlling series systems. As an alternative to the front panel control, the GUI will allow the user to control all necessary functions of the system from a user friendly windows based menu. Additionally the GUI can be used as a diagnostic tool when the system is controlled via the front panel.
3. Point of contact for this memorandum is Kathlyn Stephens at (405) 582-1230, or e-mail at Kathlyn.Stephens@us.af.mil