This document outlines the salient characteristics for a Hemodynamic Monitoring System. The key requirements include compatibility with Swan-Ganz pulmonary artery catheters and Edwards oximetry central venous catheters, the ability to provide a range of hemodynamic measurements such as Central Venous Oxygen Saturation, Continuous Cardiac Output, Right Ventricular Ejection Fraction, and Pulmonary Vascular Resistance, scalable expansion modules, wired and wireless communication, multiple clinical support screens, and a monitor screen size between 12"-18". The related federal contract opportunity is Solicitation Number FA301024QKK05 for a Hemodynamic Monitoring System from the Department of the Air Force Air Education and Training Command, with a description matching the salient characteristics detailed in this document.

View the file

Other files

File Type Posted
FA301024QKK05 - Hemodynamic Monitor.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

March 7, 2024

Salient Characteristics HemoDynamic Monitoring System

1) Hemodynamic monitoring capability must be compatible with Swan-Ganz pulmonary artery catheters.

2) Monitoring capability must be compatible with Edwards oximetry central venous catheters.

3) Must include a range of support screen choices that provide visual-clinical clarity to facilitate decision making during rapidly changing situations.

4) Must be able to provide Central Venous Oxygen Saturation (ScvO2).

5) Must be able to provide Continuous Cardiac Output (COO, CCI).

6) Must be able to provide Continuous Right Ventricular Ejection Fraction (RVEF).

7) Must be able to provide Continuous Right Ventricular End Diastolic Volume (RVEDV, RVEDVI).

8) Must be able to provide Mixed Venous Oxygen Saturation (SvO2).

9) Must be able to provide Pulmonary Vascular Resistance (PVR).

10) Must be able to provide Right Ventricular Stroke Work Index (RVSWI).

11) Must be able to provide Stroke Volume (SV, SVI).

12) Must be able to provide Systemic Vascular Resistance (SVR, SVRI).

13) Must have scalable expansion modules and interchangeable cable ports.

14) Must have wired and wireless communication.

15) Must have multiple clinical support screens.

16) Must have minimum of 2 hot-swappable battery packs and include charger.

17) Monitor must have plug-in option for power.

18) Monitor screen must be between 12”-18”.