This document is an attachment that outlines the salient characteristics for Sterile Storage Cabinets required under Solicitation Number FA301024QKK04. The solicitation is for the Department of the Air Force Air Education and Training Command to procure 5 Operating Room Console Miscellaneous Supply Cabinets and 3 Channel Spreaders.

The cabinets must be stainless steel with glass-windowed doors, at least 4 stainless steel shelves, stainless steel sloped tops, sub-bases of at least 10 inches, stainless steel end panels, a maximum width of 36 inches, 2-inch clearance from walls, and a height between 6.5 and 7 feet with a depth between 17 and 20 inches. The 3 channel spreaders are required for spacing between the connecting cabinets. The solicitation requires installation of all 5 cabinets and accessories per the provided diagram.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

7 May 2024

Salient Characteristics

5 each - Operating Room Console Miscellaneous Supply Cabinets

• Must be stainless steel.

• Must have doors with glass windows to allow visibility of sterile contents.

• Must have at least 4 stainless steel shelves.

• Must have stainless steel slopes for top of each. (to meet sterility standard)

• Must include sub-bases for each at a minimum of 10 inches that will fully enclose space below cabinets.

• Must include stainless steel end panels to fully enclose console units.

• Must not exceed more than 36” wide for each cabinet.

• Much include spacer to provide 2” clearance from wall for each unit.

• Height must be at least 6.5 ft, not exceed 7 ft.

• Depth must be a minimum of 17”, not to exceed 20”.

3 Each – Channel spreaders for three middle Operating Room Console Miscellaneous Supply Cabinets

• Required for spacing between each connecting unit.

Must include installation of all 5 cabinets and accessories. See diagram for basic installation layout.