PURCHASE DESCRIPTION

FOR

AFPSL - Sensor EPROM Burning System

1. SCOPE:

• This Automated and Manual Calibration system shall be a complete turn-key solution (hardware, software, controller and printers) for updating power sensor correction factors for all vendors power sensors listed (See Appendix A), from 9 kHz to 50 GHz.

2. GENERAL REQUIREMENTS:

• This system shall upload EPROM Data, after it performs calibration of an EPROM enabled power sensor. The system shall, after operator concurrence, automatically upload the calibration factors to the power sensor Device Under Test (DUT).

• Power Splitter Support Option: The Sensor EPROM Burning System shall come with additional hardware and software as required to support 2-port power splitter calibrations up to 50 GHz reporting the Full S-Parameters.

• Calibration method for Reflection Coefficient (of DUT): this shall be determined by means of the

"Juroshek" method.

• Source calibration factors: these shall be determined by comparison with two transfer standards (listed in 2.5) from 100 kHz to 50 GHz, calculating power applied shall be determined by the complex reflection coefficients of DUT, power splitter and transfer standard.

• From 100 MHz to 50 GHz the measurement deviations caused by mismatch are to be corrected for under consideration of the complex reflection coefficients of DUT and complex reflection coefficients of the transfer standard (Gamma Correction). Below 100 MHz the small deviations are to be taken into account by the uncertainty budget based on the DUT’s Reflection Specifications.

• Linearity deviation: this shall be determined by this system with direct current in comparison with the power dissipated in a precision 50 Ohms resistor of known value at the output port.

• This system shall measure and adjust linearity deviations of diode sensor operation outside the square law region of the diode and measure power from -36 to + 33dBm relative to the 50 MHz 0dBm cal point with a +/- 1% Uncertainty.

• System will support an additional linearity measurement and adjustment operation utilizing a level source or signal generators such as the Fluke 9610 or 9640.

• This system shall support two reference standards (listed in 3.5.2) during the calibration process, allowing users to select the most accurate standard for a frequency range.

• For calibration factors below 100 kHz, the Sensor EPROM Burning System shall utilize a high accuracy voltage meter, Fluke 5790A/AF, to provide NIST traceable calibrations between 9 to 100 kHz.

• Measurement uncertainty: The measurement uncertainties are to be calculated and provided during calibration and must contain the uncertainties which includes the standards used and matching uncertainty of the DUT. The uncertainty stated shall be the expanded uncertainty (coverage factor k = 2).

• Software must support hardware configurations currently used by Air Force Calibration activities.

These are: (Tegam 1805A/B, 1806A & 1830, Agilent 8360 Series & PSG, Antirsu 68000 & 69000 Series Signal Generators, Agilent PSA-X Series VNA/PNA, N5424A (26.5 GHz), E8364B/C (50 GHz) Fluke 5790A/AF, 9610A/AF / 9640A/AF, and Rohde and Schwarz ZVA50 50 GHz).

• The Air Force reserves the right to distribute the software to Air Force Base PMELS.

3.0 PERFORMANCE REQUIREMENTS:

3.1 Frequency and Power:

Frequency Range: 9 kHz to 50 GHz Power Range: 0.001 to 100 mW (-30 to +20 dBm) RF Power Output >+33 dBm @ 50 MHz >+20 dBm from 9 kHz to 26.5 GHz >+16 dBm from 26.5 GHz to 50 GHz

3.2 Determine Calibration Factor (C.F.) using gamma corrections (Reflection Coefficient): Rho measurement uncertainties shall be directly taken from a VNA/PNA (listed in 3.5), the measurement uncertainty calculations shall be based on the specifications of a VNA/PNA, budget using splitter, mismatch error, source error.

Cal Factor Uncertainties;

< 2.5 % from 9 kHz to 100 kHz < 2.0 % from 100 kHz to 18 GHz < 2.5 % from 18 GHz to 26.5 GHz < 3.0 % from 26.5GHz to 40 GHz < 4.5 % from 40 GHz to 50 GHz

3.3 Equivalent Source Match: Rho measurement uncertainties are a directly taken from a VNA/PNA.

As a result the measurement uncertainty calculations are based on the specifications of a VNA/PNA.

SWR (rho) Uncertainties < 0.01 from 100 kHz to 18 GHz < 0.015 from 18 GHz to 26.5 GHz < 0.020 from 26.5 GHz to 40 GHz < 0.030 from 40 GHz to 50 GHz

3.4 Amplification Specifications:

Cal Factor / High bandwidth Path:

>+20 dBm from 100 kHz to 18 GHz, frequency response < 3dBm;

>+20 dBm from 100 kHz to 26.5 GHz, frequency response < 3dBm;

>+16 dBm from 100 kHz to 50 GHz, frequency response < 5dBm

Linearity / High power 50 MHz Path:

>+33 dBm @ 50 MHz – Linearity Test Path

3.5 Equipment List:

3.5.1 Equipment to be provided by the Contractor: The Power Sensor EPROM Burning System shall be brand name or equal to the following minimum suite of equipment.

(Qty 1) 11901B Agilent 2.4 mm female to APC-3.5 female adapter

(Qty 1) 11903B Agilent 2.4mm (f) to type-N (f) adapter

(Qty 1) N1913A Agilent Single Channel Power Meter With Options 101, 005, ABA, 903 Includes 1 power sensor cable

(Qty 1) N1914A Agilent Single Channel Power Meter With Options 101, 005, ABA, 903 Includes 2 power sensor cables

(Qty 1) N8481A Agilent Power sensor 10MHz to 18 GHz -30 to +20dBm

(Qty 2) N8487A Agilent Power Sensor - Thermocouple, average, 50MHz to 50GHz 8487A-CFT Cal Factor Option, -30dBm to +20dBm 8487A-H84 Avr Pwr,50MHz-50GHz, -30dBm to 20dBm, 2.4mm(male), (Qty 1) 8481D Agilent Power Sensor 10MHz to 18 GHz -70 to -20dBm

(Qty 2) 8487D Agilent Power Sensor, 50 MHz to 50 GHz, -70 to -20 dBm 8487D-ATO Power Sensor, 50 MHz to 50 GHz, -70 to -20 dBm 8487D-H84 Std lab cal at specified frequencies

(Qty 1) 11667A Agilent DC-18 GHz power splitter, type N, 50 ohm 11667A-ATO DC-18 GHz power splitter, type N, 50 ohm 11667A-STD Type-N female input and female output connector 1

(Qty 1) 11667B Agilent Power Splitter, DC To 26.5 GHz, 3.5 mm Female Connectors

(Qty 1) 11667C Agilent DC - 50 GHz power splitter, 50 ohm

(Qty 1) UTB-Rack Assembly UTB System rack including system power, power cabling, rack fans, rails, panels, and GPIB cabling

(Qty 1) Work Shelf

(Qty 1) STORAGE DRAWER, 3.5 INCH-RACK MOUNT

(Qty 1) STORAGE DRAWER, 5.5 INCH-RACK MOUNT

(Qty 2) 1U Custom Feed-thru panel with bulkhead adaptors

(Qty 1) Cal Lab Solutions Z2890A PCS-K150 Amplified Switch Assembly with GPIB Control

3.5.2 Government Equipment: The system must integrate/operate with and support the following Air Force equipment.

N5242A Agilent PNA-L network Analyzer (26.5 GHz)

ZVA50 Rohde and Schwarz (50 GHz)

8753ES Agilent ENA Series Network Analyzer (6 GHz)

5790A/AF Fluke AC Measurement Standard

1830A Tegam RF Power Meter / RF Bridge

F1116 Tegam Low freq 100 k to 10 MHz Type N

F1109 Tegam Mid Freq 10 MHz to 18 GHz Type N

F1117A Tegam High Freq 50 MHz to 26.5 GHz 3.5 mm

ML2438A Anritsu Power Meter

9640A/AF Fluke RF Reference Source

4.0 PHYSICAL/MECHANICAL REQUIREMENTS:

4.1 Custom Leads / Cables: The following custom cables to be provided with the standard configuration of this system:

1ea- Custom flexible 50 GHz, 4 ft., 2.4mm, male-to-male interconnect cable 1ea- 2.4mm, female-to-female patch panel connection with semi-ridged male-to-male loop back cable, custom fit length.

1ea- Custom flexible 18 GHz, 4 ft., N-type, male-to-male interconnect cable 1ea- N-Type, female-to-female patch panel connection with semi-ridged male-to-male loop back cable, custom fit length.

4.2 Power Requirements: The total power requirements of this system configuration shall be less than 1700 Watts (115VAC 50/60 Hz, 15 Amps). All necessary power and electrical connections shall be supplied by the rack.

4.3 Weight: Total system configuration weight shall be less than 400 lbs.

4.4 Rack Dimension Size: In a single rack configuration is 80” high, 24” Wide and 36” deep.

4.5 RF Input Connector: The transfer standard must use a precision 2.4mm RF connector with gold-plated copper center pin and stainless steel outer conductor.

4.6 VNA Output connections: Type 2.4 mm coaxial female, gold-plated copper inner conductor, passivated stainless steel outer conductor. DUT port connector shall be separable from the splitter for inexpensive replacement, if wear occurs.

4.7 Operating Environment: All purchase description specifications shall be met when the unit is used within this operating environment range. The instrument shall perform to all specifications within the following environment:

Temperature Range: 22.8 ºC ±3.3 ºC (73 ±6 ºF).

Humidity Range: 20% to 50% RH

4.8 Storage Environment: Damage to the unit shall not occur as a result of storage in the following environment:

Temperature Range: 0 ºC to 50 ºC (32 ºF to 122 ºF)

Humidity Range: 15% to 80% RH

5.0 SYSTEM CONTROLLER REQUIREMENTS:

When firmware or software upgrade is anticipated by a manufacturer during the life cycle of this contract, the contractor shall contact AFMETCAL for concurrence. If AFMETCAL concurs with the proposed upgrade.

5.1 SCOPE:

A laptop controller is required that can efficiently operate under the Microsoft Windows 7 Professional operating system.

5.2 GENERAL DESCRIPTION:

Controller: A laptop controller is required that can efficiently operate under Microsoft Windows 7 Professional operating system. All major components of the system must be commercially available and fully supportable by commercial sources. Where specific brands are identified in this specification, no substitutes are allowed without prior approval from the Government. The Controller must meet or exceed Microsoft Windows 7 Professional hardware requirements.

5.3 SALIENT CHARACTERISTICS:

5.3.1 Processor: Intel Pentium IV or otherwise equivalent or better generation minimum,1.8 GHz clock speed minimum.

5.3.2 Memory: Minimum 4 GB system Memory.

5.3.3 Input Output (I/O) External Ports, Slots and Connectors:

At minimum, the following types of I/O interface connections must be available on the controller or via port replication:

5.3.3.1 USB: Minimum of four integrated USB Ver 2.0 compliant ports.

5.3.3.2 Ethernet: One 100/1000 BaseT Ethernet RJ45 port.

5.3.3.3 Video: Areo-Capable Graphics card that supports Windows Display Driver Model (WDDM

1.0) with a minimum of 128 MB.

5.3.4 Display Panel: Minimum 17-flat panel LCD Color Display Panel with at least 1280 x 720 resolution for a 16:9 aspect ratio.

5.3.5 Hard Drive: 80 GB (Minimum) Hard Disk Drive.

5.3.6 DVD-ROM/CD-RW Drive:

DVD-R: at least 8X Speed CD-RW: at least 24X Read Speed, 16x Write Speed

5.3.7 Printer:

The printer shall be a laser printer

Height/Width/Depth shall not exceed 12 x 14 x 16 in.

Weight shall not exceed 30 lbs.

Print speed shall at a minimum be 20 ppm.

Printer Resolution shall at a minimum be 600 x 600 dpi.

Printer shall accept Letter size (8.5 x 11 in) Printer interface shall be USB

5.3.8 Operating System and Software: All software must run under or Microsoft Windows7

(Professional operating system or greater, 32 bit, latest version release). All software must run Windows 7 as a non elevated user. Windows XP shall not be provided.

5.3.9 Additional Hardware: Additional hardware required to make this system function as identified by this purchase description shall be required.

5.3.9.1 GPIB Controller Interface Adapters: one minimum.

5.3.9.2 IEEE-488.2 compatible.

5.3.9.3 Speed: Full-speed 12MB/s USB signaling.

5.3.9.4 Switches: Plug and Play configuration.

5.3.9.5 GPIB Transfer Rates: 880kbytes/s or higher

5.4 Source Code:

5.4.1 The source code and all necessary libraries to compile the system, with rights to reuse and modify shall be included.

5.4.2 If the system does NOT compile under the Microsoft .net framework version 3.5 in Microsoft Visual Studio 2010 then a development environment must be provided.

5.4.3 All source code shall have with it an associated set of written “user instructions” for operation. In addition, all “calibration” source code shall have with it an associated set of written step-by-step “calibration instructions” for operation.

6.0 Technical Support: The vendor shall provide technical support for calibration and operation of this Power Sensor EPROM Burning System in the Air Force Primary Standards Lab (AFPSL), through on-line, E-mail and telephone communication, throughout the warranty period.

7.0 On-Site Training: As part of the delivery of this system the contractor shall provide 2 weeks (10 working days) of on-site training (at the AFPSL Heath, Ohio) on the technical operation, physical operation, calibration (including manual methodology verifications, Testing Power Sensors with the supplied software (rho, cal factor, linearity) and Printing labels & Cal Results), Uploading / Programming the EPROMS and features as described by this Purchase Description and required by the contractors Acceptance Test Plan (as approved by WR-ALC/ENHEE). Training will be in the form of hands-on instruction and written visual aids. Additionally, the contractor shall provide any engineering drawings/documents, and upon request up to a maximum five on-site (at the AFPSL Heath, Ohio) visits by the contractors technical staff, required to assist the AFPSL/(WR-ALC/ENHEE) in establishment of self-sufficiency for the Power Sensor EPROM Burning System.

8.0 Special Hardware/ Software/Licenses Requirements: Any special equipment, fixtures, tools, standards, etc., required for calibration and verification and not common to a primary calibration laboratory, must be supplied with the Power Sensor EPROM Burning System. Any unique accessories including software required for calibration and verification of this item shall be furnished by the contractor. Units that fail calibration/verification and are recommended to be aligned or adjusted, either manually or using an automated alignment/adjustment software, the special hardware, manual instructions, software, software instructions and software licenses (where required by the manufacturer, and shall be valid for the life of the instrument) needed to perform the alignment/adjustment process shall be included with the RF Power Sensors /Modules. A list of peripheral support equipment required for accomplishing the calibration, alignment, and adjustments of the unit must be provided with the Power Sensor EPROM Burning System. The Air Force reserves the right to distribute the software to Air Force Base PMELS.

9.0 Full Access to Calibration Software Source Code:

Calibration source code shall be able to be updated and/or modified by AFMETCAL, full edit rights shall be provided.

9.1 Power Sensor Calibration Software: Shall support the calibration of following power sensors.

(See appendix A for a detailed list)

HP/Agilent 84xx Series Sensors, HP/Agilent 11722A/11792A Sensor Modules, HP/Agilent E930x Series Sensors (Average), HP/Agilent E44xx Series Sensors, Agilent N848xA Series Sensors, Giga-tronics Power Sensors (80301A, 80302A, 80303A, 80304A, 80320A, 80321A, 80322A, 80323A, 80324A, 80325A, 80401A, 80410A, 80420A, 80421A, 80422A, 80425A, 80601A, 80621A)

10.0 ANCILLARY/ACCESSORIES: Any-and-All ancillary equipment/items and accessories (not otherwise specified) needed to perform as described shall be provided by the contractor. This would include, but not limited to, special tools, cables, fixtures, programs and Standards.

11.0 Testing: This item will be inspected and accepted at the destination as specified elsewhere in this contract. The item will be checked with test equipment with parameters equal to or better than those required to assure specified performance requirements are met. Tests will be performed to assure the item meets the paragraphs of this purchase description. Such testing at destination does not relieve the contractor of performing all inspections and quality checks at the point of fabrication, as is specified elsewhere in the contract, or, as is necessary, to assure performance as specified herein.

Attachment:

1) Data Requirements

2) Appendix A

Appendix A Power Sensor

N5530 Power Modules Agilent N5532A-504 Agilent N5532A-518 Agilent N5532A-526 Agilent N5532A-550

8902 Power Modules HP\Agilent 11722A HP\Agilent 11792A

848x Series HP\Agilent 8481A HP\Agilent 8481B HP\Agilent 8481D HP\Agilent 8481D-039 HP\Agilent 8481H HP\Agilent 8482A HP\Agilent 8482B HP\Agilent 8482H HP\Agilent 8483A HP\Agilent 8484A HP\Agilent 8485A HP\Agilent 8485A-033 HP\Agilent 8485D HP\Agilent 8487A HP\Agilent 8487D

E-Series (EPROM) HP\Agilent EPC-E18A HP\Agilent EPC-E26A HP\Agilent E4412A HP\Agilent E4413A HP\Agilent E4416A HP\Agilent E4417A HP\Agilent E4418A/B HP\Agilent E9300A HP\Agilent E9300B HP\Agilent E9300H HP\Agilent E9301A HP\Agilent E9301B HP\Agilent E9301H HP\Agilent E9304A HP\Agilent E9304A-H18 HP\Agilent E9321A HP\Agilent E9322A HP\Agilent E9323A HP\Agilent E9325A HP\Agilent E9326A HP\Agilent E9327A

N-Series (EPROM) Agilent N8481A Agilent N8482A Agilent N8485A Agilent N8485A Agilent N8486AR/AQ Agilent N8487A

Peak Average Agilent E9325A Agilent E9326A Agilent E9321A Agilent E9322A Agilent E9323A Agilent E9327A

Thermister Mounts HP\Agilent 478A-H() HP\Agilent 8478A/B

Giga-tronics Giga-tronics 80310A Giga-tronics 80313A Giga-tronics 80314A Giga-tronics 80350A Giga-tronics 80353A Giga-tronics 80354A Giga-tronics 80351A Giga-tronics 80352A Giga-tronics 80355A Giga-tronics 80330A Giga-tronics 80333A Giga-tronics 80334A Giga-tronics 80401A Giga-tronics 80402A Giga-tronics 80410A Giga-tronics 80420A Giga-tronics 80421A Giga-tronics 80422A Giga-tronics 80425A Giga-tronics 80501A Giga-tronics 80502A Giga-tronics 80503A Giga-tronics 80504A Giga-tronics 80601A Giga-tronics 80621A Giga-tronics 80701A

ANRITSU

Anritsu MA2421A Anritsu MA2442B Anritsu MA2472B Anritsu MA2473B Anritsu MA2474B

Rohde and Schwarz Rohde NRVS-B1 Rohde NRVS-Z1 Rohde NRVS-Z4 Rohde NRVS-Z51 Rohde NRVS- Z54 (W) 25DB ATTENUATOR

Attachment to Purchase Description 12E-150A-RF

Sensor EPROM Burning System

DATA REQUIREMENTS

MANUALS: A complete user and service manual and calibration procedure shall be provided with each unit in contractor format. Manuals shall be on CD/DVD-ROM in Indexed Portable Document Format

(iPDF). The manuals shall comply with Data Item Description (DID) DI-TMSS-80527B and the Contract

Data Requirements List (DD Form 1423).

CALIBRATION CERTIFICATE: A calibration certificate, traceable to the National Institute of Standards and Technology (NIST) or other national metrology institutes (NMis) shall be provided with each instrument. This calibration certificate must comply with Data Item Description (DID) DI-QCIC-80798B and the Contract Data Requirements List (DD Form 1423).

ACCEPTANCE TEST PLAN: The acceptance test plan shall be delivered to and approved by the government prior to delivery of the evaluation unit. The acceptance test plan shall detail the criteria, performance objectives, and description of tests to be performed. The plan shall be prepared in accordance with Data Item Description DI-QCIC-80553 and Contract Data Requirements List (DD Form

1423).

SOFTWARE TEST PLAN: The Software Test Plan (STP) describes plans for qualification testing of

Computer Software Configuration Items (CSCis) and software systems. It describes the software test environment to be used for the testing, identifies the tests to be performed and provides schedules for test activities. The software test plan shall be prepared in accordance with Data Item Description DI

IPSC-81438A and Contract Data Requirements List (DD Form 1423).

SOFTWARE USER MANUAL: The software user manual shall contain instructions to execute one or more related computer software configuration items. The software user manual shall be provided in paper format and prepared in accordance with_Data Item Description DI-IPSC-81443A and Contract Data

Requirements List (DD Form 1423).