Exhibit A - Height Adjustable Reclining Tub Specifications - final.pdf
PDF 153 KB Posted
- Attached to
- Height Adjustable Reclining Bath Tub
- Solicitation number
- 1279
- Issued by
- Bonner County, Idaho
About this file
This is a specifications exhibit for a state procurement by the Idaho Department of Veterans Services for height adjustable reclining bath tubs to be used at the Idaho State Veterans Home in Pocatello, Idaho. The specification document outlines detailed requirements for brand new, manufacturer-approved tubs that must meet dimensional constraints including length between 70 to 82 inches, width between 28 to 32 inches, and height between 52 to 60 inches, with access height and seat height specifications of 22 to 24 inches. The tubs must include features such as polished fiberglass or equivalent finish, 110-volt hard-wired electrical connection, adjustable water temperature controls, auto-fill capability, air spa/jet functionality, head and back padding, footrest, and tilt, raise, and lower functions controllable via both panel and hand-held pendant controls. Tubs must accommodate users weighing up to 450 pounds with a safe working load capacity of 660 pounds (patient plus water), and water consumption must range between 24 to 26 gallons at pre-filled/autofill level and 58 to 62 gallons at overflow level. The plumbing system must handle temperatures between 50 and 122 degrees Fahrenheit with accuracy of plus or minus 1.8 degrees Fahrenheit.
The specification document indicates that delivery and placement of the new tubs are required as part of the bid. Bidders must ensure all component parts meet or exceed industry standards for maximum performance and safety, and the tubs must come with standard manufacturer warranty and access to local service and maintenance capability. Minor dimensional deviations may be accepted at the sole discretion of the Maintenance Foreman if the tubs properly fit the available space. The document does not contain pricing information, contract term details, set-aside designations, incumbent vendor information, or specific response and award deadlines within this exhibit.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Exhibit A Event 1279
Specifications for Height Adjustable Reclining Tub(s)
The following specifications are a description of a Height Adjustable Reclining Tub, hereinafter referred to as a "Tub(s)". The Tub(s) will be used to provide quality care for the residents of the Idaho State Veterans Home in Pocatello, Idaho.
It is the responsibility of manufacturers to use materials, components, and design practices, which are the best available, and current to industry standards for the type of operating conditions which this type of Tub(s) may be subject to. All component parts of the Tub(s) shall not merely meet the minimum requirements of this specification, but be selected to give maximum performance, service life and safety of Idaho's Veterans.
A Tub(s) offered in compliance with these specifications must be brand new, not used or refurbished, and be the manufacturer's newest model. Accessories must be engineered, designed, or certified by the manufacturer to be compatible with all other components.
The Tub(s) must come with a standard manufacture warranty and have warranty service and future maintenance onsite or with a local vendor that has the ability and qualifications to repair or service the Tub(s).
Minor deviations of the dimensions and connections may be accepted after a review by the Maintenance Foreman for acceptance, as long as the deviations allow the Tub(s) to properly fit in the space available.
If deviations are outside the space available as deemed solely by the Maintenance Foreman, bid may not be considered. IDVS reserves the right to determine minor deviations.
1. Dimensions
1.1 Length – Acceptable overall length may be no shorter than seventy (70") inches and no more than eight-two (82") inches.
1.2 Width – Acceptable overall width may be no less than twenty-eight (28") inches and no more than thirty-two (32") inches.
1.3 Height – Acceptable overall height may be no shorter than fifty-two (52") inches and no taller than sixty (60") inches.
1.4 Access height must be twenty-two (22") to twenty-four (24") inches to accommodate transfer from standard powered patient lifts.
1.5 Seat height, when lowered must be a minimum of twenty-two (22") Inches and a maximum of twenty-four (24") inches.
2. Electrical Requirement
2.1 One hundred and ten (110) volt hard-wired connection
3. Construction
3.1 Finish must be polished fiberglass or equivalent to meet disinfection protocols in the facility.
3.2 Drain must be located in center of foot bed tub floor.
3.3 Plumbing temperature handling: fifty (50°) Fahrenheit – one hundred and twenty-two
(122°) Fahrenheit, accuracy +/- 1.8° Fahrenheit
3.4 Access door on left side of the user in the tub.
3.5 Exterior must be white or grey. Small accent colors are acceptable.
4. Controls
4.1 Must include panel (on unit) and hand-held (pendant) controls.
4.2 Raise/Lower, Tilt, and Air jet control functions must be adjustable on panel and held-held controls.
5. Capacity
5.1 Must accommodate users weighing ≥ four hundred and fifty (450) lbs.
5.2 Safe working load (patient + water) ≥ six hundred and sixty (660) lbs.
5.3 Water consumption at pre-filled/autofill level must be no less than twenty-four (24) gallons and no more than twenty-six (26) gallons.
5.4 Water consumption at overflow level must be no less fifty-eight (58) gallons and no more than sixty-two (62) gallons.
6. Other Features (must include)
6.1 In-tub water temperature adjustable temperature controls
6.2 Auto Fill
6.3 Handicap Accessible
6.4 Air spa/Air jet functionality
6.5 Head and back padding/cushions
6.6 Footrest
6.7 Tilt, Raise, and Lower functions.
7. Miscellaneous
7.1 No dealership, reseller plates, insignias or emulsions adhered to Tub(s).
7.2 Delivery, placement of new Tub(s) are required.