ESBD_504019_1775484433594_Pressure_transducer_specifications.pdf

PDF 80 KB Posted

Attached to
Pressure Transducer
Solicitation number
580-26-IFB-5138
Issued by
Travis County, Austin City, Texas

About this file

This is a technical specifications document for a pressure transducer device related to a Texas state or local contract opportunity. The pressure transducer is designed for water-level measurement with a measurement time of less than 1.5 seconds and offers multiple output options including SDI-12 (version 1.3) at 1200 bps and RS-232 at 9600 bps. The device operates within a temperature range of 0 to 60 degrees Celsius with a temperature accuracy of 0.2 degrees Celsius and features a water-level resolution of 0.0035% full-scale range. The sensor is equipped with an overpressure rating of 2 times the pressure range and requires power input between 6 to 18 volts DC, with quiescent power consumption below 50 microamps and active measurement/communication consumption up to 8 milliamps for a one-second measurement cycle, with a maximum of 40 milliamps.

The high-accuracy option for this pressure transducer provides a total error band of 0.05% full-scale range, which accounts for combined errors from nonlinearity, hysteresis, nonrepeatability, and thermal effects across the compensated temperature range per ISA S51.1 standards. The device supports a maximum cable length of approximately 457 meters (1500 feet) when a single SDI-12 sensor is connected to a port. A critical operational warning notes that the sensor could be damaged if encased in frozen ice, which is an important consideration for deployment in certain climates.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Pressure Transducer Specifications

Measurement Time < 1.5 s

Output Options SDI-12 (version 1.3) 1200 bps; RS-232 9600 bps

Water-Level Resolution 0.0035% FS

Worst-Case Temperature Resolution 0.006°C

Operating Temperature Range 0° to 60°C WARNING: Sensor could be damaged if encased in frozen ice.

Temperature Accuracy ±0.2°C

Overpressure 2 x pressure range

Power Requirements 6 to 18 Vdc

Power Consumption

Quiescent < 50 µA

Measurement/Communication 8 mA (1 s measurement)

Maximum 40 mA

Accuracy

High Accuracy Option

±0.05% full-scale-range TEB

Total Error Band (TEB) includes the combined errors due to nonlinearity, hysteresis, nonrepeatability, and thermal effects over the compensated temperature range, per ISA S51.1.

Physical

Maximum cable length

• ~457 m (1500 ft)

1 SDI-12 sensor connected to a single port

Diameter • <21. 4 mm (0.85 in.)

Ingress Protection • Exceeds IP68

Barometric compensation • Vented cable

File details come from the government source that posted it. Updated .