ESBD_412936_1745334555382_582-25-08676 Attachment A.pdf
PDF 65 KB Posted
- Attached to
- Air Samplers
- Solicitation number
- 582-25-08676
- Issued by
- Travis County, Austin City, Texas
About this file
This document is a solicitation specification for a Sequential PM2.5 Sampler issued by the Texas Commission on Environmental Quality (TCEQ), detailing technical requirements for acquiring particulate matter monitoring instrumentation. The solicitation seeks a specialized environmental sampling device capable of automated collection of multiple PM2.5 samples over a minimum of two weeks, programmable for various sampling frequencies, and meeting specific EPA reference method standards. The equipment must include a comprehensive set of components such as a stand, pneumatically-activated filter exchanger, active volumetric flow control system, microprocessor with menu-driven software, ten filter cassettes, and multiple manuals and accessories. Delivery is expected within 120 calendar days after order receipt, with the equipment to be shipped to TCEQ's facility in Austin, Texas.
The solicitation specifies that all items must be new and of current production, with a two-year warranty covering onsite system, parts, and labor. Additional requirements include providing two sets of comprehensive manuals, in-house training for TCEQ staff at no additional cost, and technical support available during standard business hours via telephone and email. The sampler must have capabilities for reporting detailed sample information in CSV format, including filter identification, start and stop times, sample duration, average flow rates, and environmental conditions. The procurement involves standard government purchasing protocols, with equipment to be shipped Freight on Board (FOB) Destination, meaning TCEQ will take delivery once goods arrive at their warehouse.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Solicitation Name: Sequential PM2.5 Sampler Solicitation Number: 582-25-08676
Attachment A
Specifications for Sequential Particulate Matter of 2.5 Micrometers in Diameter Sampler
Description: Instrumentation requirements for sequential monitoring of suspended particulate matter of 2.5 micrometers in diameter (PM2.5) in ambient air.
MINIMUM GENERAL SPECIFICATIONS
1. Must meet U.S. Environmental Protection Agency (EPA) federal reference or equivalent methods RFPS-1006-145, RFPS-0498-118, or EQPM-0202-145.
2. Must allow for automated collection of multiple PM2.5 samples over a minimum of two weeks without operator interaction.
3. Must be programmable to allow for every day, 1-in-3 days, and 1-in-6 days sample collection.
4. Must have an operating range from 0 to 5000 micrograms per cubic meter.
5. Must be suitable for ambient sample collection without additional environmental temperature controls.
6. Sample must be collected on filter suitable for gravimetric, ion chromatography, energy dispersive x-ray fluorescence, and thermal optical analysis.
7. The Sequential Particulate Matter Sampler must include the following.
A. Stand
B. Pneumatically-activated filter exchanger in ventilated filter compartment
C. Active volumetric flow control system
D. Adapter to reduce flow to zero to check for leaks in the system
E. Microprocessor with menu-driven software
F. Display and keypad
G. Ten filter cassettes for 47-millimeter filters
H. Three filter cassette magazines with end caps
I. One magazine transport container
J. Sensors for ambient temperature, pressure, relative humidity, and filter temperature
K. Two operating manuals
L. One quick-start guide
M. One service manual
Solicitation Name: Sequential PM2.5 Sampler Solicitation Number: 582-25-08676
N. 9-to-9 pin computer cable
O. AKComm RS232 software (or current Thermo proprietary communications software)
P. Brushless pump
8. Must be able to report filter identification (Id) start date and time stop date and time total sample time, average flow, ambient temp, filter temp, ambient pressure, and any error that occurred during sample. Report must be available in comma-separated values file (CSV) format that can be downloaded to a computer or flash drive.
SPECIAL TERMS AND CONDITIONS
1. All items must be bid new, of current production, and include the manufacturer’s standard parts and accessories.
2. Delivery must be made to the Texas Commission on Environmental Quality, 12100 Park 35 Circle, Building B, Austin, Texas, 78753. Routine delivery response to TCEQ will be within 120 calendar days after receipt of order.
3. Equipment shall be shipped Freight on Board/Free on Board (FOB) Destination. The seller shall be responsible for the cost of transporting merchandise to the TCEQ delivery address listed above and the TCEQ will take delivery of shipped goods once the goods arrive at the TCEQ warehouse.
4. The warranty period should consist of complete coverage of system onsite, parts, and labor for a minimum of two years from the date of acceptance.
5. Two sets of manuals (styled: Instruction Manual, Operator’s Manual, Operation and Service Manual, etc.) must accompany each instrument. The manuals will include complete operating instructions, circuit descriptions, theory of operation, location of controls and switches, software commands, maintenance, and service instructions. The manual will also include schematic diagrams, pneumatic flow charts, lists of commonly replaced parts, warnings, and safety precautions.
6. In-house training for TCEQ staff should be provided at no additional costs to TCEQ for maintenance and diagnostic repair.
7. Technical support must be available for TCEQ staff between the hours of 8 a.m.
and 5 p.m. Central Standard Time (CST) during regular business days, Mondays through Fridays, via both telephone access and electronic mail (e-mail).
| Attachment A |
| Specifications for Sequential Particulate Matter of 2.5 Micrometers in Diameter Sampler |
| SPECIAL TERMS AND CONDITIONS |