Specifications.doc
DOC document 61 KB Posted
- Attached to
- Temperature Bath
- Solicitation number
- EA-133W-14-RQ-0450
About this file
Specifications for Temperature Bath
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications for Temperature Bath
1. Performance Specifications:
The new temperature bath must meet the minimum performance specifications listed below:
Temperature Bath Specifications Temperature range:
-100 °C to +100 °C (Heat transfer fluid type dependent)
Effective Temperature range:
-80 °C to 55 °C (when using Xiameter® PMX 200 silicon fluid)
Heat transfer fluid (current):
Must be able to operate using Dow Corning Xiameter® PMX 200 silicone liquid, Temperature stability:
±0.002 °C after stabilization period Accuracy:
±0.5 °C (when using the thermometer provided with bath)
Circulation Pump Built-in fluid circulation pump or stirring mechanism Cooling capacity:
125 watts at ‑80 °C to 600 watts at 0 °C
Heating capacity:
250 watts (low setting) or 1000 watts (high setting)
Reservoir fluid capacity:
10 gallons (38 liters)
Reservoir working volume:
12” wide × 12” long ×12” deep (units are inches) Control System:
Microprocessor control with at least eight (8) programmable temperature set points. Serial data communications port allowing all bath functions remotely controlled via a computer interface In addition to the minimum specification, preference for selection of the new will go to the manufacturer providing desired features as indicated below:
Temperature Bath Features Ethernet Data Port:
Microprocessor control with an Ethernet data communications port that will allow all remote control of all bath functions over a Local Area Network via a computer interface Data Recording Function:
Built-in data log function that continuously records temperature measurements and a function that allows the user to retrieve this data at a later date Programmable Temperature Control:
Built-in control functions that allow the user to program the system to automatically cycle through a series of different user defined set points. The system will run unattended and then shut down at the conclusion of the test run
Self-Diagnostics:
The control processor will continuously monitor the refrigeration system and notify the user if a malfunction occurs Exterior dimensions:
No more than 32” long × 19” wide × 46” tall
Shipping weight:
No more than 350 pounds (159 kilograms)
Electric power:
Run on 230 VAC, 60 Hz, 20 amps, single phase electric power