Statement of Work

Fluke Automated Pressure Calibration System Upgrades

1.0 Introduction

The Naval Air Systems Command (NAVAIR) Metrology and Calibration (METCAL) Program has a requirement to procure Fluke Electronics Corporation (Fluke) Automated Pressure Calibration System (APC) model number (M/N) 4332A upgrades to support the current NAVAIR METCAL pressure workload. The APCs in the NAVAIR inventory are used to calibrate pressure gauges, transducers, transmitters, and other high-pressure devices. The Fluke model number (M/N) 4332A is being procured as a configuration upgrade that can support the current NAVAIR METCAL pressure calibration workload. The Fluke M/N 4322A configuration is being sought to upgrade legacy Fluke M/N 4322 systems that have lost functionality due to broken and aging internal components.

2.0 Background

The APCs M/N 4322 in the NAVAIR METCAL inventory, enable users to calibrate almost any type of pressure gauge, transducer, transmitter or switch including the following: vacuum, compound, absolute and high-pressure devices. It provides high-end, laboratory-grade performance across a very wide range, on the bench or in situ. For maximum versatility and portability, the Calibrator consists of a Pressure Controller (the Controller), a Pressure Intensifier (the Intensifier), a Nitrogen Cylinder, and an Accessory Kit. At a minimum 90 % of the current Fluke M/N 4322 APC systems in the NAVAIR METCAL inventory are out of service. This configuration upgrade to both firmware and hardware elements will improve operation and maintenance of the system that will result in greater reliability, maintainability and repeatability.

3.0 Scope

The scope of this effort is to procure up to 254 Fluke M/N 4322A APC upgrades.

4.0 Applicable Documents

N/A

5.0 Requirements

5.1 Fluke 4322A Hardware Upgrades

5.1.1 Remove the BaroCheck feature.

5.1.2 Adjust the A150K module from a specified +/- 0.0036 to =/- 0.0054 pounds per square-inch (psi).

5.1.3 Upgrade the Quick Connect (QC) couplings on the model 4322A components of the controller, intensifier, contamination-prevention system (CPS) modules and two test hoses to provided better lock/unlock pressure consistency than the legacy 4322 system and reduced friction so that hand tools are no longer needed to make or break pressure connections.

5.1.4 Upgrade the pressure/vacuum pump assembly by replacing the solenoid and valve components to provide corrosion resistance.

5.1.5 Modify the 4322A pressure controller’s plastic deck to reduce the likelihood of the CPS Driver Circuit and the CPS valve being damaged by a short circuit of the CPS Valve Driver contacts due to an accumulation of liquid on the plastic deck.

5.1.6 Upgrade the CPS circuit by making the fuse holder for the fuse between the CPS contacts and the driver board accessible to an operator without having to remove the pressure controller from its enclosure.

5.1.7 Upgrade the CPS filter assembly so that no tools are required to remove or replace the filters.

5.1.8 Upgrade the Pressure Controller Quick-Connect (QC) Coupling on the supply pressure to the new generation QC coupling found on the 4322A commercial system.

5.1.9 The lid area of the pressure controller shall receive a foam insert stating, “NO STORAGE ON LID”.

5.1.10 Upgrade the pressure controller internals to be protected against liquid spills including oil. The CPS docking on the top surface of the pressure controller shall have improved protection from liquid spills emanating from that area, thus preventing spills from reaching internal components.

5.1.11 Update the Pressure Controller identification plate to reflect the new configuration designation, component and system serialization, and 2-D barcoding.

5.1.12 Upgrade the Intensifier to be case-hardened for a reduction of wear and flaking of the cam.

5.1.13 Isolate the motor encoder sensor to reduce particle contamination.

5.1.14 Upgrade the input (from N2 cylinder) and output (to pressure controller) pressure QC couplings to the new-generation 4322A QC couplings.

5.1.15 Update the Intensifier Identification Plate to reflect the new configuration designation, component and system serialization, and 2-D barcoding.

5.1.16 One (1) each blue and orange Contamination Prevention System (CPS) module shall be returned to Fluke with each system for upgrade and replacement modules shall be issued by Fluke.

5.1.17 Replace each CPS QC coupling with a new-generation 4322A QC coupling found on the commercial system.

5.1.18 Upgrade the purge manifold and filter retainers to stainless-steel versions for corrosion prevention.

5.1.19 One (1) each blue and orange test hose is to be returned to Fluke with each system for upgrade. The QC couplings shall be upgraded to the new 4322A commercial system QC couplings.

5.1.20 Upgrade the spares kits to the new 4322A commercial system spare kits, which is to include a minimum of two (2) additional fuses for each fuse holder and other support items for the upgraded components.

5.1.21 Update the accessory kit identification plate to reflect the new configuration designation, component and system serialization, and 2-D barcoding.

5.1.22 Update the N2 Cylinder Component Identification plate to reflect the new configuration designation, component and system serialization, and 2-D barcoding.

5.1.23 Calibrate all pressure transducer modules from the legacy components of the 4322 shipped by the Navy and the Fluke provided module(s) for the 4322A to the Fluke specification for the M/N 4322A.

5.1.24 Upgraded Removable Parts List. The table below illustrates the remaining parts to be upgraded and key parts that stay the same.

Kit Name
Description
Required For Upgrade
CPS-4322-BLUE
Contamination Prevention System Module for 4322, Blue
Yes
CPS-4322-0RANGE
Contamination Prevention System Module for 4322, Orange
Yes
HOSE-4322-BLUE
High-Pressure Test Hose, 10 ft., blue
Yes
HOSE-4322-0RANGE
High-Pressure Test Hose, 10 ft., orange
Yes
KIT-4322-REM
Remote Pendant
No
4322-5937
Pressure Regulator Assembly
No
WRENCH
10 in Adjustable Wrench
No
8-636
GPIB Cable
No
HOSE, INTERCONNECT
Interconnect Hose, 6 ft.
No

5.2 Fluke 4322A Firmware Upgrades

5.2.1 The required firmware upgrades are listed in the table below.

Classification
Firmware Issue Key
Description
New Feature
NAVNP-31
Need to detect if the THM door switch is unplugged or has faulty wiring
Update
NAVNP-42
The Cal Cert printed shows the temperature in °F even when the unit is set to°C.
Update
NAVNP-43
Changing the time format to 24hr reverts back to 12hr after a power cycle
Performance Improvement
NAVNP-38
A150K Exercise with pressure points to match the calibrated range.
Update
NAVNP-34
False fail indication for calibration results
Update
NAVNP-37
Cannot delete Test Sequence
Update
NAVNP-36
False indications of Outside of Adjustment Tolerance
Update/New Feature
NAVNP-33
140 psi pause during calibration exercise (Update/bug fix) & Option for low speed pressure control at pressures below 150 psi (New Feature)
New Feature
NAVNP-1
Add "1-1/2-1%" and "2-1-2%" tolerance types
New Feature
NAVNP-47
Remember previous user entry edits
New Feature
NAVNP-41
Add ability to select a pressure value for the tolerance type.
Performance Improvement
NAVNP-35
Improved Switch Test sequencing
Update
NAVNP-29
Remote command ABORT causes an error
Update
NAVNP-30
PMAX error could occur when starting a compound test.
New Feature
NAVNP-44
Remove BaroCheck
New Feature
NAVNP-45
Modify A150K alignment points

5.3 Logistics Requirements

5.3.1 Operator/Service Manual(s) (CDRL T001). Technical manuals describing how to operate the M/N 4322A shall have information on operator safety, operating procedures for the system and preventative maintenance. Service manuals shall contain information on safety for servicers of the M/N 4322A and instructions on repairing and maintaining components the original equipment manufacturer (OEM) designates as user serviceable. Technical manual(s) for the operation and/or service of the M/N 4322A APC shall be provided in accordance with CDRL T001.

5.3.1.1 The table below illustrates the key maintenance items that shall be outlined in CDRL T001.

Procedure
Level
4322 Configuration Schedule
4322A Configuration Schedule
Replace seal, and lubricate pump
Depot
Lesser of 2 years, 1,000 hours or as required
Lesser of 2 years, 1,000 hours or as required (same)
Replace pump low- pressure valves
Depot
N/A
Lesser of 2 years, 1,000 hours or as required (new)
Replace CPS fine (fiber) filter
User
1 year or as required
1 year or as required (same)
Replace CPS course (metal) filter
User
1 year or as required
1 year or as required (same)
Flush CPS
User
N/A
As required
Pressure controller fuse replacement
User
As required
As required (same)
CPS circuit fuse replacement
User
N/A
As required (new)
BaroCheck
User
30 days
N/A (eliminated)
Calibration
User
All ranges 1 year
All ranges 1 year
System leak check
User
As required
As required (same)
Intensifier fuse replacement
User
As required
As required (same)
Intensifier pressure generate/leak test
User
Lesser of 1 year, 750 hours or as required
Lesser of 1 year, 750 hours or as required (same)
Intensifier pressure sensor alignment
Depot
Lesser of 2 years, 1,500 hours or as required
Lesser of 2 years, 1,500 hours or as required (same)
Intensifier seals replacement
Depot
Lesser of 2 years, 1,000 hours or as required
Lesser of 2 years, 1,000 hours or as required (same)

5.3.2 Configuration Management/Design Analysis (CDRL E001). The hardware, software and firmware used to operate each and every calibration standard procured on any individual delivery order under this contract shall be of the same release and version. No deviation or substitution of hardware, software and/or firmware shall be done without the approval of the US Government. The contractor shall provide the US Government the release and/or version number of each item of hardware, software and firmware included in the calibration standard in accordance with CDRL E001.

5.3.3 Software/Firmware Change Request (CDRL A001). When an item of firmware or software needs to be changed, a Software/Firmware Change Request shall be submitted to the US Government for approval in accordance with CDRL A001.

5.3.4 Calibration Certification with Report and/or Certificate of Conformance Report (CDRL A002). A calibration certification and/or Certificate of conformance with calibration data shall be provided in accordance with CDRL A002.

5.3.5 Instrument Calibration Procedure (ICP) (CDRL A003). The OEM ICP shall be provided for calibrating the M/N 4322A, for both manual and automated procedures, in accordance with CDRL A003.

5.3.6 Repairable Parts List (CDRL A004). A Repairable Parts list shall be provided to the Government for each standard delivered, in accordance with CDRL A004.

5.3.7 Warranty. The manufacturer shall provide the minimum commercial warranty for the calibration standard of at least one (1) calendar year from date of shipment.

5.3.8 Warranty Sticker. Each calibration standard shall have a sticker affixed to the outside of the unit enclosure.

5.3.9 All systems shall be serialized. A new serial number shall be labeled on each system at the issuance of each delivery order. Each system serial number shall be provided to the Navy within 15 business days after the issuance of each delivery order.

5.3.10 Legacy 4322 components sent to Fluke for upgrade are to be sent in sets, complete with the hardware as agreed upon by the Navy and Fluke. Upgraded 4322A systems will be labeled and returned to the Navy in complete sets.

5.3.11 Navy will be responsible for packaging and shipment of legacy 4322 units to the OEM.

5.3.12 Return shipment to CONUS Navy Warehouse. Address to be provided with each delivery order The address of possible CONUS delivery sites can be found in the table below.

Name of Location
Address
Marine Corps. Air Station Beaufort
NAVAIR Calibration Standards

Attn: Lou Cerrillo (UIC N48535)

BLDG 612, Bay 9 MCAS

Beaufort, SC 29904

Phone: (843)524-1405

Navy Calibration Lab (NCL) Patuxent River
Naval Air Warfare Center Navy Calibration Lab (PRL)

ATTN: Lab Director Greg Miller (301) 342-1654 22113 Fortin Circle

BLDG 1403

NAWC Patuxent River, MD 20670-1118

6.0 Government Furnished Property

6.1 Government Furnished Equipment

All Government furnished equipment will be listed on each delivery order by CAGE Code Model Number and Serial Number.

7.0 Deliverables

7.1 The table below illustrates data deliverable list.

CDRL No.
Nomenclature
Data Item Description (DID)
DID #
T001
Technical Manual(s)
Commercial Off-The-Shelf (COTS) Manuals and Associated Supplemental Data
DI-TMSS-80527C
E001
Configuration Data for Hardware, Software and Firmware
Design Analysis
DI-SESS-80567B
A001
Software Firmware Change Request
Software Firmware Change Request
DI-MISC-81807
A002
Calibration Certificate with Report
Calibration Certificate
DI-QCIC-80798B
A003
Instrument Calibration Procedure
Inspection and Test Plan

DI-QCIC-81110

A004
Repairable Parts List
Proposed Spare Parts List
DI-ILSS-80134A

Note: 1. Contractor Format is acceptable for all CDRLs.