DID-QCIC-80798C Calibration Certificate Report.pdf
PDF 28 KB Posted
- Attached to
- Fluke APCS 4322AUP Upgrade Federal contract opportunity
- Solicitation number
- N6426724Q0260
About this file
The document provided is a Data Item Description (DID) for a Calibration Certificate/Report. It specifies the format and required content for a calibration certificate or report that accompanies the calibration of Department of Defense (DoD) test, measurement, and diagnostic equipment prior to delivery. The DID is consistent with ISO/IEC 17025 requirements for calibration reports or certificates and includes details such as the laboratory information, identification of the calibrated item, calibration results, measurement uncertainty, and traceability evidence.
The related federal contract opportunity is a sole source solicitation for the Fluke APCS 4322AUP Upgrade. The Naval Surface Warfare Center (NSWC) Corona Division intends to award a firm fixed price contract to Fluke Electronics for this all-or-none requirement. Interested parties may submit capability statements, but this is not a request for competitive quotes. Offerors must be registered in SAM, and quotes are due by the response date provided.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Attachment B - GFP Attachment.pdf | ||
| Attachment A - Automated Pressure Calibration System 4322 to 4322AUP Upgrade Specifications.pdf | ||
| A001_Redacted.pdf | ||
| Combined Synopsis-Solicitation N6426724Q0260.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
DATA ITEM DESCRIPTION
Title: CALIBRATION CERTIFICATE/REPORT
Number: DI-QCIC-80798C Approved Date: 20170123 AMSC Number: F9766 Limitation: N/A DTIC Applicable: No GIDEP Applicable: No Preparing Activity: 36 Project Number: QCIC-2017-003 Applicable Forms: N/A Use/Relationship: The Calibration Certificate/Report documents calibration certificates or reports accompanying the calibrations performed on Department of Defense (DoD) test, measurement, and diagnostic equipment.
a. This Data Item Description (DID) contains the format and content preparation instructions for the data product generated by the specific and discrete task requirement as delineated in the contract.
b. This DID is applicable to contracts requiring calibration of test equipment prior to delivery.
This DID is consistent with ISO/IEC 17025, General requirements for the competence of testing and calibration laboratories, requirements for calibration reports or certificates.
(Copies of this document are available online at http://www.iso.org.)
c. This DID supersedes DI-QCIC-80798B.
Requirements:
1. Reference documents. The applicable issue of the documents cited herein, including their approval dates and dates of any applicable amendments, notices, and revisions, shall be as specified in the contract.
2. Format. Contractor format is acceptable.
3. Content. The calibration certificate or report shall include:
a. Title (e.g., calibration certificate or calibration report);
b. The name and address of the laboratory and the location where the calibrations were carried out, if different from the address of the laboratory;
c. Unique identification of the calibration certificate or test report (such as the serial number), and on each page an identification in order to ensure that the page is recognized as a part of the test report or calibration certificate, and a clear identification of the end of the calibration certificate or test report;
d. The name and address of the client or customer;
e. Identification of the method used;
f. A description of, the condition of, and the unambiguous identification of the item(s) tested or calibrated;
g. The date of receipt of the test or calibration item(s) where this is critical to the validity and application of the results, and the date(s) of performance of the test or calibration;
DISTRIBUTION STATEMENT A. Approved for public release. Distribution is unlimited.
Source: https://assist.dla.mil -- Downloaded: 2019-07-16T12:34Z Check the source to verify that this is the current version before use.
DI-QCIC-80798C
h. Reference to the sampling plan and procedures used by the laboratory or other bodies where these are relevant to the validity or application of the results;
i. The test or calibration results with, where appropriate, the units of measurement;
j. The name(s), function(s) and signature(s) or equivalent identification of person(s) authorizing the calibration certificate or test report;
k. Where relevant, a statement to the effect that the results relate only to the items tested or calibrated;
l. The conditions (e.g., environmental) under which the calibrations were made that have an influence on the measurement results;
m. The uncertainty of measurement or a statement of compliance with an identified metrological specification or clauses thereof;
n. Evidence that the measurements are traceable.
End of DI-QCIC-80798C.