Description of Requirements.pdf
PDF 161 KB Posted
- Solicitation number
- W912HZ-21-N-9888
About this file
This document outlines requirements for laboratory equipment and installation services. The US Army Corps of Engineers Engineer Research and Development Center requires one each of a Thermogravimetric Analyzer, Dynamic Mechanical Analyzer, Modulated Differential Scanning Calorimeter, Discovery Hybrid Rheometer, and associated accessories from TA Instruments-Waters L.L.C. to support material analysis and quality control activities. Delivery and installation shall occur at the USACE ERDC campus in Vicksburg, Mississippi. A sole source award is intended to TA Instruments-Waters L.L.C., as they are the only vendor capable of meeting requirements. Interested parties may submit capability statements by November 16, 2020 for consideration, though no solicitation exists and the decision to compete is at the Government's discretion.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Description of Requirements
1.0 BACKGROUND
The Environmental Laboratory (EL) is recognized worldwide as the problem solver in environmental science and engineering research for US Army, the Department of Defense, and the Nation. R&D activities address a multitude of environmental issues to advance civil works, water resources, military installations, and the warfighter. Research activities include ecosystem restoration, ecological resources, and contaminant management, risk and decision science, environmental sensing, ecological modeling and forecasting, environmental security, and environmentally sustainable material.
The instruments purchased must meet the specifications outlined below so that data gathered from previous experiments with identical equipment can be validated for use with data gathered on future experiments within the same projects for use in future journal articles and research publications.
2.0 Equipment needed
2.1 One TA Instruments thermogravimetric analyzer (TGA 5500), one dynamic mechanical analyzer (DMA 850), one modulated differential scanning calorimeter (DSC 25), one discovery hybrid rheometer (HR 20), associated accessories, and installation and basic training from TA Instruments-Waters L.L.C.
1) Thermogravimetric analyzer (Discovery TGA 5500) part number 956000.901 qty of 1 with the following specifications needed: A vertical thermo-balance with an auto-Switching dual range microbalance with a 0-200 mg and 0-1000 mg weight range, an infrared furnace and horizontal gas purge system with a temperature range from ambient to 1200 degrees C, linear heating rates to 500 degrees C with ballistic heating rates to >1,600 degrees C/minimum, a 25 position autosampler with a contamination free sealed pan punching system, an integrated electromagnet for automated Curie-point temperature calibration, a DTA signal for temperature calibration, a gas delivery manifold for switching and control between two different gasses. Requires DSL Router with four port switches (P/N: 251471.002).
TGA 5500 Accessories needed
a) TGA aluminum sealed pan kit part number 957352.901 Qty of 1: Contains 100 aluminum pans, 100 aluminum lids, 30 stainless steel bails, punch alignment tool, crimping die, and crimping tool for use with the tzero sample press (P/N: 901600.901).
2) Modulated differential scanning calorimeter (Discovery DSC 25) part number 973700.903 qty of 1 with the following specifications needed: A maximum operation temperature of 725 °C with a lower temperature limit down to -180 °C with liquid Nitrogen cooling, must be compatible with TRIOS software.
Installation of the DSC requires a DSL Router with four port switch (P/N: 251471.002) which is listed below.
DSC 25 Accessories needed
a) Discovery Refrigerated Cooling System - RCS40 (120VAC/60Hz) part number 972507.901 qty of 1 with the following specifications: Operating temperature range from -40 °C to 400 °C and be compatible with the DSC 25.
b) RCS Gas Dryer part number 200266.001 qty of 1: For removing moisture in purge gases;
should provide dry gas to -90°C dew point.
c) Tzero® Press and Die Sets Kit part number 901600.901 qty of 1 containing the following: Tzero Press, four (4) die sets, and one box each of Tzero Pans and Tzero Lids (100 / box).
3) Dynamic Mechanical Analyzer (Discovery DMA 850) part number 986000.901 qty of 1 with the following specifications needed: Operation range from -160 to 600 degrees C, capable of viscoelastic measurements on materials from 1000 Pa to 1000 GPa, accurate measurements up to 18 N with a sensitivity to 1 mN, accuracy of deformation amplitudes from 0.0005 to 10,000 μm with resolution of 0.1 nm, frequency range of operation from 0.001 to 200 Hz, mechanical or liquid nitrogen cooling, must be compatible with furnace par number 986012.901 and operate with TRIOS software. Requires a DSL Router with a four-port switch (P/N: 251471.002)
• Comes with an accessory kit containing a 35mm single/dual cantilever bending fixture, digital caliper, steel standards, precision torque driver, 1000 g certified calibration weight, and demonstration materials.
DMA 850 Accessories needed
a) Discovery DMA 850 Standard Furnace part number 986012.901 qty of 1: Must be able to control temperature up to 600 °C and be compatible with all DMA 850 clamps.
b) Film / Fiber Tension Kit for DMA 850 part number 985016.901 qty of 1: Kit includes one set of smooth clamp fixtures for testing films and fibers in tensile mode. Sample length range of 5 to 30 mm, width to 6.5 mm, and thickness to 2 mm.
c) Nitrogen Purge Cooler part # 986310.901 qty of 1; Extends temperature range of the DMA 850 to -160°C. The NPC consists of a 2.5 L Dewar flask that contains a copper coil tube.
d) DMA 850 Filter/Regulator part number 986320.901 qty of 1
4) Discovery hybrid rheometer (HR 20) part number 534002.901 qty of 1 with the following specifications needed: Able to perform rheological measurements in rotation and oscillation, a torque range of 1 nN.m to 200 mN.m, resolution of 2 nanoradians, elevated angular velocity to 300 rad/s, and normal force up to 50 N and must be compatible with TRIOS software.
HR 20 Accessories needed
a) HR x0 Environmental Test Chamber part number 534040.901 qty of 1 with the following specifications: Provide heating rates up to 60 °C/min and temperature range from ambient to 600°C, able to use with ETC Geometries, including parallel plates/cones, disposable plates, and torsion clamps for solids, with a temperature range extendable to -160°C using liquid nitrogen control.
b) HR x0 ETC Parallel Plate Accessory Kit part number 534066.901 qty of 1: Includes lower Smart Swap 2 assembly, 25 mm diameter stainless steel parallel plate Smart Swap 2 Geometry, wire brush, brass scraper, and melt rings.
c) HR x0, DHR & AR-Series ETC Flow Meter Kit part number 543311.901 qty of 1; required for the ETC if no cooling accessory is specified.
d) HR x0, DHR & AR Rheometer Filter/Regulator part number 250000.001 qty of 1
5) DSL Router with 4-Port Switch part number 251471.002 qty of 1: for communication between an external controller and Thermal and Rheology instrument; mentioned as needed above.
6) Thermal Installation & Training fee part number 999800.528 qty of 1: Installation provided according to
ISO 9000 procedure. The ISO 9000 procedure will insure consistency and high quality throughout the installation process. The installation (except DMA) will include calibration with a traceable standard and a certificate of calibration. Basic Training is also provided on the equipment after it is installed in the laboratory.
7) HR x0, DHR & AR-Series installation & training fee part number 999800.529 qty of 1: Installation provided according to ISO 9000 procedures. Insures consistency and high quality throughout the installation process. The installation will include calibration with a traceable standard and a certificate of calibration. Basic Training is also provided on the equipment after it is installed in the laboratory.
2.2 Delivery
Ship to:
Alan Kennedy
CEERD-EPR
3909 Halls Ferry Rd.
Vicksburg, MS 39180
Instrument setup and training will take place at the ACOE ERDC Vicksburg campus.
2.2 Warranty and Maintenance
Standard warranty as stated in the quote applies. TA states that there will be guaranteed response times to phone calls, shipment of parts, and equipment repair. Response time is not stated and should be negotiated.
24 hours is reasonable.