LIMITED SOURCE JUSTIFICATION
1. Defense Logistics Agency (DLA) Troop Support, Medical Supply Chain, Capital Equipment Division, is the contracting activity.
2. This justification is for the acquisition of brand name products or features essential to United States Air Force (USAF) requirements under Zoll Medical's Electronic Catalog (ECAT) contract number SPM2D1-l3-D-8220 which expires September 3, 2018.
3. The USAF have a requirement for the following items:
Item Description: Propaq MD Vital Signs Monitor/Defibrillator and accessories
(USAF Advanced Configuration/trade required)
Document Number: FM444482040458 Zoll Medical P/N: 306-2451312-101 Quantity: 200 each Unit Price: $21,495.00 Estimated Dollar Value: $4,299,000.00
Total Estimated Dollar Value: $4,299,000.00
4. The authority waiving the competition requirements is FAR 8.405-6(a) (1) (i) (B): only one source is capable of providing the supplies or services required at the level of quality required because the supplies or services ordered are unique or highly specialized. The Zoll Propaq MD Vital Signs Monitor/Defibrillator identified above is the only product available on ECAT that can meet the USAF’s needs.
The USAF will be purchasing these items as a replacement of older/aged critical care defibrillators and Patient Vital Signs Monitors in the USAF medical Patient Movement Item Centers (PMI) Aeromedical evacuation (Critical Care Air Transport Teams-CCATT; Inflight Kits; and Tactical Critical Care Evacuation Team-TCCET) and ground medical assemblages.
The required Vital Signs Monitor/Defibrillator must satisfy the following minimum requirements that were established by the AF for a monitor/defibrillator in a field environment:
a) Shall provide a minimum of two ports each for temperature monitoring and invasive blood pressure (IBP) monitoring.
b) Shall provide for a Pleth Variability Index (PVI).
c) Shall be capable of operating on line voltages of 110-240V AC, at frequencies of 50/60 Hz, and for aircraft at a line voltage of 110V AC at a frequency of 400Hz.
d) Shall have a Joint Airworthiness Certification (JAC) consisting of a “Safe to Fly” letter from the USAF for fixed and rotary wing aircraft and a Fleet Airworthiness Release (AWR) and an Aeromedical Certification Letter (AWC) from the US Army.
e) Shall have a maximum weight of 15 lbs.
f) The device shall be capable of mounting to the existing Vital Signs Monitor bracket on the
Special Medical Emergency Evacuation Device (SMEED).
g) Shall offer a Night Vision Goggle (NVG) display mode for military and air medical night time operation.
h) Shall have an operating time for at least 6.0 hours on a single battery charge.
i) Shall be a monitor and defibrillator capability that has FDA 510K Pre-Market Approval (PMA) and that is Automated External Defibrillation (AED) capable.
j) Shall have the capability to print patient data to an internal printer or an accessory printer which connects to the Monitor/Defibrillator. The printer shall be suitable for use in aeromedical operations and in a field environment.
The Zoll Propaq MD Vital Signs Monitor/ Defibrillator av ai l ab l e o n E C A T meets these minimum requirements.
5. DLA Troop Support has already determined the prices of supplies under the ECAT multiple award program fair and reasonable. Price reasonableness is based on receiving a 25% discount from Zoll’s commercial catalog price. Zoll's most favored Group Purchasing Organizations receive a 22.5% discount.
A determination has been made by the contracting officer that this order represents the best value to the Government and results in the lowest overall cost alternative (considering price, special features, and administrative cost) consistent with FAR 8.404(d).
6. Market research was performed by the USAF in August 2018. Following is a summary of the results:
• The Physio Control Lifepak 15 defibrillator/monitor was identified as the product that most closely meets the US Army’s minimum requirements; however, the Lifepak 15 does not meet the USAF and the Army’s minimum requirements as stated in paragraph 4 for the following:
(a) Does not provide a minimum of two ports each for temperature monitoring and IBP monitoring.
(b) Does not provide for a PVI. PVI is required for critical care patients to enable non-invasive and continuous assessment of fluid status of critical care patients.
(c) Does not operate on a frequency of 400HZ, which is aircraft power. Aircraft power requires operation at 110V and at a frequency of 400Hz.
(d) Does not have a JAC consisting of a “Safe to Fly” letter from the USAF for fixed and rotary wing aircraft and a Fleet AWR and an AWC from the US Army.
(e) Does not meet the maximum weight of 15 lbs. It weighs 20.1 lbs.
(f) The device does not fit on the existing Vital Signs Monitor bracket on the SMEED.
• The Phillips Healthcare HeartStart MRx model monitor was previously identified however, this model has reached end of life and is no longer available. Philips no longer offers it commercially or to the Government.
Additional market research was performed by the DLA Engineer, USA and USAF in August 2018. An additional potential supplier, Remote Diagnostics Technology (RDT), was identified on ECAT. RDT manufactures two units- the Tempus Pro Monitor and the Tempus ALS monitor/defibrillator. The Tempus Pro Monitor has no Automated External Defibrillation (AED) capability (element 4(i) above) and does not have an internal or accessory printer suitable for use in aeromedical operations and in a field environment (element 4(j) above). The Tempus ALS monitor/defibrillator, on the other hand, has not yet received FDA approval nor is it available to the US market or appear on any Government contract until late 2018 and possibly into early 2019; consequently, the RDT Tempus Pro and Temple ALS do not meet the AF’s minimum requirements.
A review of the contracts on ECAT revealed that the Zoll Propaq MD Vital Signs Monitor/ Defibrillator is the only product currently on ECAT that can meet the AF’s minimum requirements.
7. Any ECAT vendor not currently offering the required product will have the opportunity to add the product, enhancing competition. Furthermore, ECAT vendors have the opportunity to add products that fall within the scope of the ECAT contract at any time.
8. DLA Troop Support will attempt to locate other acceptable sources for future routine requirements as they arise; however, any new item would be required to meet the AF’s rigorous requirements since there are life and death consequences if the item cannot function in adverse conditions.
9. As the Contracting Officer, I hereby certify that the information contained in this Limited Source Justification is accurate and complete to the best of my knowledge and belief.
I hereby certify that the requirements that form the basis of this justification are accurate, that these supplies meet the Government's minimum needs and that this justification is accurate and complete to the best of my knowledge and belief.