This is a technical specification document for the Orthoscan Mini C-Arm X-Ray device, covering three models (1000-0015, 1000-0016, and 1000-0017) contained in the Operators Manual.
The document provides comprehensive technical specifications across multiple categories. For power requirements, the devices operate on 100/120/230 V AC at 50 or 60 Hz with currents under 5A depending on voltage, utilizing specific Bel Fuse ratings per voltage level. The X-ray assembly features a Petrick P030.40.20GW tube with a 42.5-micron focal spot, 80kVp rated peak potential, operational range of 40-80kVp, and tube currents of 40-160µA (Canada: 40-110µA). The system includes a 0.1mm copper optimized dose filter for reducing entrance dose during adult and pediatric imaging. The high-voltage power supply operates in continuous mode for Model 1000-0015 and pulsed/continuous modes for Models 1000-0016/1000-0017, with fluoroscopy pulse rates of 30/15/7.5/2.0 fps for the latter models. Environmental operating conditions range from 15-32°C ambient temperature with 0-90% relative humidity (non-condensing), while storage conditions span 0-55°C with humidity up to 100% and atmospheric pressure of 500-1060hPa.
Imaging specifications vary by model, with touchscreen monitors at 24 inches for Models 1000-0015/1000-0016 and 27 inches for Model 1000-0017. The TAU 1512 detector (Model 1000-0015) provides a 14.5x11.5 cm sensitive area with 1944x1536 pixel matrix, while TAU 1515 (Model 1000-0016) offers 15x15 cm at 1548x1548 pixels, and TAU 2020 (Model 1000-0017) supplies 20.4x21.8 cm at 2236x2064 pixels. The C-Arm system includes 160° orbital range and 430° pivotal range with a 45cm source-to-detector distance. All models support 30 fps standard image and DVR/Cine recording with HDMI video output, 26,000 image capacity, and 14.4 minutes of video storage. The system weighs approximately 215 kg with a 7.62-meter power cord.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Amend-1 B Clause Admin Data 8Pages.docx | DOCX document | |
| 1 RFQ-26-PHX-038 Synopsis-Solicitation.docx | DOCX document | |
| A Price Schedule RFQ-038 1page.xlsx | XLSX spreadsheet | |
| B Clause Admin Data RFQ-038 8Pages.docx | DOCX document | |
| C 2025 AZ Tax Exempt Letter PHX Area 3pages.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Technical Information 10-2
ORTHOSCAN MINI C-ARM OPERATORS MANUAL MODELS: 1000-0015 / 1000-0016 / 1000-0017
050-0041 REV L
Orthoscan TAU Product Specification
NOTE: Product design and specifications are subject to change without Notice
INDUSTRY DEVICE CLASSIFICATIONS
Equipment Classification Class 1 Device
Water ingress protection classification
Ordinary Equipment
Mode of operation Continuous Operation
Electric shock protection classification
Type B applied part
COUNTRY DEVICE CLASSIFICATIONS
Mini C-arm X-Ray Device Class II (FDA) / Class IIb (EU) / Class III (Canada)
AC POWER REQUIREMENTS
Nominal rated line voltage Model: 1000-0015/ 1000-0016 / 1000-0017:
• 100 / 120 / 230 V~ (factory preset)
Frequency (Hertz) 50 Hz or 60 Hz
Line voltage regulation + 10% daily line reg.
Current <5A at 120 V~, ≤5A at 100 V~, ≤3A at 230 V~
Mains Fuse Ratings
Model: 1000-0015/ 1000-0016/ 1000-0017:
• 100 VAC~: Bel Fuse 5ET (P) 5-R
• 120 VAC~: Bel Fuse 5ET (P) 4-R
• 230 VAC~: Bel Fuse 5ET (P) 2-R
LASER SPECIFICATION
Laser Beam Class IIa: 3.5mW, 635nm Wavelength
Laser Light Class 2: IEC 60825-1
FDA Compliance Compliant to FDA CFR Part 1040.10
Equipment Marking Warning (IEC 60825-1)
X-RAY ASSEMBLY SPECIFICATIONS
Duty cycle at 75 kVp / 0.1mA 1.00 (continuous)
X-Ray tube Petrick P030.40.20GW (OS)
Beam filtration 2.5mm Aluminum equivalent
Focal spot 42.5 microns
Rated peak tube potential 80kVp
Tube kVp range 40kVp to 80kVp
Tube current range 40µA to 160µA (Canada: 40µA to 110µA)
Attachment D: Salient Characteristics
Technical Information 10-3
050-0041 REV L
Deviation from indicated values kV +7%, µA +10%
Heat Capacity 3900 J
Target Angle 20°
Max. Continuous Power 50 W
Anode Material Copper w/ Tungsten Target
Cathode Material Nickel
Tube Cooling/ High Voltage Insulation
High Dielectric Transformer Oil
ADDED OPTIMIZED DOSE FILTER SPECIFICATIONS
Material Cu (copper)
Thickness 0.1mm
Method of use User engages for adult and pediatric imaging to reduce entrance Dose
Accessible Not directly accessible to operator
GENERATOR/ HIGH VOLTAGE POWER SUPPLY (HVPS) CHARACTERISTICS
Mode(s) Of Operation Model 1000-0015: Continuous Model 1000-0016/ 1000-0017: Pulsed and Continuous
HVPS Operating Voltage Range 40kV - 78kV
HVPS Operating Current Range 40µA to 160µA
Fluoroscopy Pulse Rate per Second Model 1000-0016/ 1000-0017: 30 / 15 / 7.5 / 2.0 Model 1000-0015: Continuous
ENVIRONMENTAL SPECIFICATIONS
Storage: Temperature Range 0˚C (32˚F) to 55˚C (131˚F)
Storage: Relative Humidity >0% to <100% (no condensation)
Storage: Atmospheric Pressure 500hPa to 1060hPa
Operating: Temperature (Ambient) 15º to 32ºC (59º to 90ºF)
Operating: Humidity (Ambient) 0% – 90% relative humidity, non-condensing
Technical Information 10-4
IMAGING SPECIFICATIONS
Image processing Digital
Monitor Model: 1000-0015: 24” – Touchscreen Model: 1000-0016: 24” – Touchscreen Model: 1000-0017: 27” – Touchscreen
Resolution Model: 1000-0015/1000-0016: 1920x1200 (WUXGA) Model: 1000-0017: 1920x1080 (FHD)
Standard Image 30 Frames per second (fps)
DVR/Cine 30 Frames per second (fps)
Video Output Signal (Ext’l Display) HDMI
Useful Entrance Field Size:
Model: 1000-0015 (12x15) – 14.5 x 11.5 (cm) Model: 1000-0016 (15x15) - 15 x 15 (cm) Model: 1000-0017 (20x20) – 20.4 x 20.4 (cm)
Image Capacity 26,000
Video Capacity 14.4 min
Technical Information 10-5
IMAGE DETECTORS (SSXI) SPECIFICATIONS
TAU 1512 - CMOS SSXI
Sensor Type CMOS Active Pixel Sensor
Scintillator GdOx
Pixel pitch/Size 75μm / 74.8μm
Sensitive area 14.5x 11.5 cm
Pixel Matrix 1944 x 1536 pixels
ADC resolution 14 bits kV Range 12 to 130 kV
Mechanical
Weight 2.2kg (4.8lbs)
Dimension L x W x H 22.3 x 15.0 x 4.2 cm
Operating Temperature Range +100C to +400C
Storage Temperature -100C to +500C
Communication
Cameral link GigE
TAU 1515 -Detector - CMOS SSXI
Sensor Type CMOS Active Pixel Sensor
Scintillator GdOx
Pixel width x height 99 x 99 μm
Sensitive area 15.0 x 15.0 cm
Total Pixels 1548 x 1548 pixels
ADC resolution 14 bits kV Range 40 to 125 kV
Mechanical
Weight 2.0kg (4.4lbs)
Technical Information 10-6
SYSTEM MECHANICAL CHARACTERISTICS
Cabinet Footprint: C-Arm Extended 208cm (82”)
Cabinet Footprint: C-Arm Retracted 101cm (39.8”)
Base Cabinet Footprint – Width 73.7cm (29.0”)
Base Cabinet Footprint – Length 83.8cm (33.0”)
C-Arm Arc Depth 50.8cm (20.0”)
C-Arm Orbital Range 160°
C-Arm Pivotal Range 430°
Dimension L x W x H 22.5 x 19.0 x 3.8 cm
Operating Temperature Range +100C to +400C
Storage Temperature -100C to +550C
Communication
Camera link GigE
TAU 2020 -Detector - CMOS SSXI
Sensor Type CMOS Active Pixel Sensor
Scintillator 600µm CsI
Pixel width x height 99 x 99 μm
Sensitive area 20.4 x 21.8 cm
Total Pixels 2236 x 2064 pixels
ADC resolution 16 bits kV Range 40 to 125 kV
Mechanical
Weight 4.5kg (10lbs)
Dimension H x W x D (cm) 23.5 x 29.2 x 5.0
Operating Temperature Range +100C to +550C
Storage Temperature -100C to +550C
Communication
Camera link GigE
Technical Information 10-7
Lateral Rotation: Wig Wag 320°
Flex Arm Vertical Range 67.3cm (26.5”)
Monitor Arm Rotational Range Shoulder: 216°, Elbow: 370°, Wrist: 200°
Monitor Arm: Horizontal reach 66cm (26“)
Monitor: Vertical height 48.3cm (19“)
Monitor: Vertical travel 35.6cm (14”)
Monitor: Height at lowest position 113cm (48“)
Detector: Highest point above ground
111.8cm (44.0”)
Ground Clearance 10.1cm (4.0“)
Rear Casters Lockable / Bi-directional
Curb Weight (~215 kg ~475 lbs)
Cord Length 7.62m (25 feet)
Printer Spacing 8.9cm x 16cm (3.5“ x 6.25“)
SOURCE TO DETECTOR DISTANCES
Source to Detector Distance (SDD) 45cm (17.72”)
Free Space (Functional SID) 26cm (10.24”) (w/Spacer Cone) 35cm (13.8”) (Optional Spacing w/o Spacer Cone)
Source to Skin Distance (SSD) 10cm (3.94”) (Spacing w/o Spacer Cone) 19cm (7.48”) (Spacing w/Spacer Cone)
File details come from the government source that posted it. Updated .