This document is a Compliance Spreadsheet Template that appears to be related to a federal contract opportunity. The spreadsheet outlines essential characteristics (ECs) for an aeromedically certified suction apparatus, including requirements for operating modes, power sources, controls, certifications, environmental specifications, size/weight, alarms, and maintenance/support.

The related federal contract opportunity is a Request for Information (RFI) from the Defense Logistics Agency Troop Support Medical for aeromedically certified suction apparatuses. The RFI is a market research effort to gather technical information, quotes, and support documentation on available products and components. Responses to the RFI are due by 5:00 PM EST on 05 April 2024.

View the file

Other files

File Type Posted
Suction RFI-2024_v2.pdf PDF
Suction RFI-2024.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sheet1

Essential CharacteristicThresholdObjectiveVendor Compliant (Yes/No)Comments
Key Performance Parameters
EC#1Shall be able to operate in both continuous and intermittent suction modes.YEST=O
EC#2The apparatus shall be capable of operating from multiple power sources including a component battery. Shall be capable of operating from any of the specified external AC or DC power sources while the internal battery is being recharged via the component battery charging circuit. Simultaneous operation and recharge shall not effect recharge time.Shall operate from 110~240 VAC, 50/60 Hz, and an internal rechargeable battery. Automatic switching between these power sources is strongly preferred.Shall operate from 110˜240 VAC, 50/60/400 Hz and 12-28 VDC, and an internal rechargeable battery. Automatic switching between these power sources is strongly preferred.
EC#3Vendor shall provide the option of power plug adapters.YEST=O
EC#4The unit must have an integral battery charger and operate and charge at the same time.YEST=O
EC#5Shall have FDA Clearance and comply with applicable sections of title 21, Code of Federal Regulations (21 CFR).YEST=O
EC#6Controls for the device shall include: Mode control switch - off/continuous/intermittent, On Time Selector Control - Continuously variable, Off Time Selector Control, Vacuum Adjust Control , AC/DC Voltage Protector , and a Safety LockoutControls for the device shall include: Mode control switch - off/continuous/intermittent, On Time Selector Control - Continuously variable, at least 5 thru 40 seconds, Off Time Selector Control - Continuously variable, at least 5 thru 40 seconds at all combinations, Vacuum Adjust Control - Continuously variable over the low and high ranges, AC/DC Voltage Protector - Fuse or circuit breaker, and a Safety Lockout - No intermittent suction in high rangeT=O
Key System Attributes
EC#7Shall be in compliance with the Trades Agreement Act (TAA)YEST=O
EC#8Shall be able to perform suctioning for the following clinical applications:
oropharyngeal, tracheal, wound draining and abdominal/thoracic decompression.YEST=O
EC#9The suction apparatus shall provide two adjustable ranges of vacuum.The low range shall be adjustable from 10 to a maximum of 200 mmHg for intermittent suction procedures. The high range shall be adjustable from 0 to at least 550 mmHg for continuous high vacuum procedures.T=O
EC#10Device must run continuously on battery power on a single charge on max suction2 hrs6 hrs

(System being used during ground transportation, extended lines of evacuation and speed will require systems to last longer. Ground platforms do not have the electrical capacity to manage charging needs en route.)

EC#11Device must fully charge (> = 90 %) a depleted battery6 hrs4 hrs
EC#12Must have spare power adapterNOYES
EC#13Device must have a battery capacity gaugeYEST=O
EC#14Device must have battery shelf Life2 years connected/4 years not connectedT=O
EC#15Certificate of Airworthiness prior to final selection. (Guidance's involved concerning airspace equipment are the following: AFMAN 10-2909, DAFI 48-107, AFMAN 11-2AE (Checklist), AFMAN 11-2AE Vol 1 to 3.)Device shall have a Joint Airworthiness Certification (JAC) consisting of a safe-to-fly letter from the U.S. Air Force (for Fixed and Rotary Wing), Fleet Airworthiness Release (AWR) and an Aeromedical Certification Letter (AWC) from the U.S. Army. (Above testing results are not required prior to being considered for submission by vendor)T=O
EC#16The Device shall have Operational and Storage temperature ranges within allowable limits for cold and hot storage and operational environments per Joint Enroute Care Equipment Test Standard (JECETS) which specifies temperature limits and specifies using the test methods in 810G for Tactical-Standby to Operational environmental Tests.YesT=O
EC#17Securing of device while operationalDevice must be operational while secured vertically and horizontally with accessibility of controlsT=O
EC#18Pressure adjustment - Pediatric to AdultContinuous pressure must be adjustable. Must be = to or > than 550 mmHg , adjustable "0" to maximum settingsGreater than 600mmHg
EC#19Vacuum MeasurementMust be able to measure vacuum in mmHg within 10% of indicated rateT=O
EC#20Device must have a suction gauge.Amount of negative suction provided by device will be easily discernable to the operator. Numeric Negative suction displayed, preferred digitally.T=0 Amount of negative suction provided by device will be easily discernable to the operator.
EC#21Flow Rate VacuumFlow rate at max vacuum, liters per minute of air = 30Greater than 30
EC#22Intermittent pressure adjustabilityThe pressure range should be from 10 to 200 mmHgT=O
EC#23Device must be able to accommodate adult patientsYEST=O
EC#24Device must be able to accommodate pediatric patientsYEST=O
EC#25The Device shall have visual and audible alarms with the ability to silence active alarms and all-alarm suspension; silence duration in seconds configurable by user. Active alarm visual indicators must remain visible while audible alarm is silenced.Alarm conformance IAW IEC 60601-1-8, Medical Electrical Equipment - General requirements for safety - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems. The high and low alarm limits shall be independently adjustable and read on the display.T=O
EC#26Low-light and NVG compatibleThe Device display should have low-light and NVG compatible modalities.T=O
EC#27Equipment WeightLess than or equal to 10lbs5lbs
EC#28Device must be one handed carry with accessories / suppliesYEST=O
EC#29Shall have a portable configuration, with the ability to be mounted onto a IV pole or rail. Shall have the ability to be secured to a standard NATO litter and SMEED.YEST=O
EC#30Device must have operating temperature range0 degrees Fahrenheit to 120 degrees FahrenheitMinus 25 to 150 Fahrenheit
EC#31Shall be ruggedized to withstand transportation over secondary roads and cross country. Will be validated via Environmental Testing events: Transit Drop and Road Shock & Vibration testing in accordance to MIL STD 810G. Airworthiness assessment will be conducted in accordance to the JECETS standard.YEST=O
EC#32Device must be impact resistant /durable. Drop tests are performed to assure that the device can withstand the drop shocks that may be encountered in handling, transportation, and service environments.safe operation after a 3 foot dropsafe operation after a 6 foot drop
EC#33Must have pump contamination preventionBacterial filter between the canister and the pump.Hydrophobic filter between the canister and the pump.
EC#34Must be water resistantYEST=O
EC#35Must operate at low noise levelShall not exceed 60dbA for high and 50dbA for low vacuum (@ one meter from all sides)Less than 40 dbA for high vacuum and 30 dbA for low vacuum (@ one meter from all sides)
EC#36Device and canister must have back flow preventionYEST=O
EC#37Collection canistersCanister for apparatus shall be able to contain up to 2000mL, and apparatus shall be capable of holding a reusable or disposable canister. Collection canisters shall be configurable to operate singly or in tandem with one another. Each canister shall include an overflow shutoff valve cartridge that can be removed and reinstalled without the use of tools to facilitate and make possible tandem use. Collection canisters shall be constructed from clear, non-breakable, autoclavable material if reusable. Canister lids shall be self-sealing to prevent leakage.Canister lids shall be self-sealing to prevent leakage and be autoclavable. Tube fittings and the overflow shutoff valve cartridge and any other components comprising the complete collection canister/lid assembly shall be autoclavable if reusable. Disposable canisters will be collapsible to save space during storage. Device can accept multiple size canisters from 200 ml to 2000ml.
EC#38As an optional capability, a vehicle power cord assembly shall be provided, at least 4 feet in length, having a vehicle cigar lighter plug on one end and a mating suction apparatus plug on the other endNOYES
EC#39Vendor Provided TrainingVendor must provide initial training via classroom, cd, onlineT=O
EC#40Vendor will provide training materials for users on new functionality and enhancementsYEST=O
EC#41Vendor must provide alternative training methods utilizing the latest distance learning technologyYEST=O
EC#42Training material must either be CD or manuals.YEST=O
EC#43Shall contain two (2) sets of operating/maintenance manuals. (Service manual will include theory of operation, trouble shooting, all maintenance and calibration requirements, illustrated parts breakdown (IPB) with part numbers for each replaceable component, all applicable schematics [electronic, pneumatic, hydraulic, etc]; manuals will be the same as provided to OEM service engineers).Electronic and hard copyT=O
EC#44Shall provide a full listing of repair parts and distinguish the items to be considered high-failure repair parts (identified by OEM and Government SMEs). The country of origin, availability, proprietary parts, and purchaseability of each item shall be provided on the list. Requirements for these items would be solicited under separate delivery orders issued in accordance with FAR Part 16.5.YEST=O
EC#45Consumables must be non-proprietary / universalYEST=O
EC#46Shall provide a full listing of consumables, accessories and optional items. The country of origin, availability, proprietary parts, and purchaseability of each item shall be provided on the list. Requirements for these items would be solicited under separate delivery orders issued in accordance with FAR Part 16.5.YEST=O
EC#47Shall agree to support all repair parts, accessories, consumables and optional items under their existing IDIQ contract. Shall indicate which items are on their IDIQ contract and shall agree to add those items that are not currently on their IDIQ contract prior to delivery order selection under this RFQ. This includes agreement to add the items on the DLA Electronic Catalog (ECAT) for direct online ordering by DLA customers, if implemented by DLA. For those items listed as not on contract and which are TAA compliant, the vendor must provide an acceptable contract addition request along with their proposal. Any such items that cannot be added to the IDIQ contract (i.e. pricing cannot be determined as fair and reasonable, etc.) may result in rejection of the quote.YEST=O
EC#48Shall allow biomed personnel who maintain the equipment and the unit supply/logistics support personnel to contact manufacturer to request information, verify, and order supplies as required (i.e. Fed-ex, or DHL, or UPS tracking Number)YEST=O
EC#49Device must be able to be cleaned and disinfected with a solution in accordance with Medical Infection Prevention and Control Program guidance Per AFI 44-108 Infection Control ProgramYEST=O
EC#50Must have manufacture warranty2 years.
Shall provide a manufacturer warranty for all Parts and In-House Labor and Technical Support during the warranty period. Operator / maintainer telephone support shall be provided 24 hours per day, 7 days per week for the entire life of the system.5 years
EC#51Maintainability at military location with Vendor training to military personnelThe Device shall be maintainable by military or civilian BMETs using standard tools (as provided in standard BMET kit; no special tools required), using non-proprietary diagnostic equipment (TMDE) without voiding warranty and shall be capable of being cleaned by commonly accepted methods. Repair parts and/or telephone consultation before and after warranty period will be provided to DoD BMETs without need to attend manufacture specific training.T=O
EC#52Device system software shall be upgradeable in the field by military Biomedical Repair Technicians, with or without internet connection, using a computer interface, the internet or a CD supplied by the OEM without requiring changing hardware components of the device; OEM shall provide notification of product software modifications and/or updates to customer and outdated software shall not render the device unusableDevice system software that may require modification during system life shall be upgradable in the field by military BMETs using a computer interface , the Internet, or a CD supplied by the manufacturer without changing hardware components of the device. Manufacturer notification of product software modification and/or updates sold to customers is required.T=O
EC#53The Device shall have Shipping and Storage temperature ranges within allowable limits for cold and hot storage and operational environments per Joint Enroute Care Equipment Test Standard (JECETS) which specifies temperature limits and specifies using the test methods in 810G for Tactical-Standby to Operational environmental Tests.Conforms to Joint Enroute Care Equipment Test Standard (JECETS) via MIL-STD-810G requirements.T=O
EC#54OEM Service and Operator manuals shall contain the current DoD Performance Specification (MIL-PRF- 32216A) EVALUATION OF COMMERCIAL OFF- THE-SHELF (COTS) MANUALS AND PREPARATION OF SUPPLEMENTAL DATA.YEST=O
EC#55Vendor must notify the Defense Logistics Agency and Services Points of Contact of all field change orders, alerts, recalls and equipment notices affecting the system for the life cycle of the system.YEST=O
EC#56The system shall have Unique Item Identification markings in compliance with DFARS 252.211-7003YesT=O