Sheet1
| Flow Hoods | Pharmacy Locations | ||||||||||||
| Primary Engineering Control - LAFW | Manufacturer Name | Nameplate Model | Mfr. Serial No. | Room No. | Pharmacy Location | Task/Requirement | Specific Tasks | Secondary Engineering Control | Building | Room Number | PECs | Task/Requirement | Specific Tasks |
| LAMINAR FLOW HOOD | NUAIRE INC | NU-301-630 | 129689032409 | 342-18A | Main Inpatient Pharmacy Buffer Area | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Airflow Hoods 2x/year, Recertify and Repair PEC | Support Pharmacy Anteroom | 3600 | L67-8A | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Anteroom Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | |
| LAMINAR FLOW HOOD | NUAIRE INC | NU-301-630 | 145243071311 | TC-216B | CoTo OR Pharmacy Buffer Area | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Airflow Hoods 2x/year, Recertify and Repair PEC | Support Pharmacy Buffer Area | 3600 | L67-8 | 177096111516 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Buffer Area Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC |
| LAMINAR FLOW HOOD | NUAIRE INC | NU-301-630 | 145244071311 | TC-216B | CoTo OR Pharmacy Buffer Area | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Airflow Hoods 2x/year, Recertify and Repair PEC | Nuclear Medicine SCA | 3600 | 136-3 | 2016-106630, 2016-106629 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | SCA Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC |
| LAMINAR FLOW HOOD | NUAIRE INC | NU-301-630 | 129690032409 | 342-18A | Main Inpatient Pharmacy Buffer Area | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Airflow Hoods 2x/year, Recertify and Repair PEC | OR/PACU SCA | 3600 | 239-7A | 177097111516 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | SCA Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC |
| LAMINAR FLOW HOOD | NUAIRE INC | NU-201-630 | 177096111516 | L67-8 | Support Pharmacy Buffer Area | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Airflow Hoods 2x/year, Recertify and Repair PEC | Main Inpatient Pharmacy Anteroom | 3600 | 342-18 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Anteroom Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | |
| LAMINAR FLOW HOOD | NUAIRE INC | NU-201-630 | 177097111516 | 239-7A | OR/PACU SCA | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Airflow Hoods 2x/year, Recertify and Repair PEC | Main Inpatient Pharmacy Buffer Area | 3600 | 342-18A | 129690032409, 129689032409 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Buffer Area Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC |
| LAMINAR FLOW HOOD | NUAIRE INC | 301-630 | 120024122007 | 537-5 | Maternal Child Pharmacy Buffer Area | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Airflow Hoods 2x/year, Recertify and Repair PEC | Maternal / Child Pharmacy Anteroom | 3600 | 537-3 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Anteroom Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | |
| LAMINAR FLOW HOOD | NUAIRE INC | 301-630 | 120025122007 | 537-5 | Maternal Child Pharmacy Buffer Area | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Airflow Hoods 2x/year, Recertify and Repair PEC | Maternal / Child Pharmacy Buffer Area 1 (Left) | 3600 | 537-4 | 134312120209 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Buffer Area Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC |
| Maternal / Child Pharmacy Buffer Area 2 (Right) | 3600 | 537-5 | 120024122007, 120025122007 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Buffer Area Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | ||||||||
| Biological Safety Cabinets | CoTo OR Pharmacy Anteroom | CoTo | TC-216A | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Anteroom Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | ||||||||
| Primary Engineering Control - BSC | Manufacturer Name | Nameplate Model | Mfr. Serial No. | Room No. | Pharmacy Location | Task/Requirement | Specific Task | CoTo OR Pharmacy Buffer Area | CoTo | TC-216B | 145244071311, 145243071311 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Buffer Area Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC |
| CABINET, BIOLOGICAL SAFETY, CLASS II B2 | NUAIRE INC | NU-435-400 | 134312120209 | 537-4 | Maternal Child Pharmacy Buffer Area | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Biological Safety Cabinets 2x/year, Recertify and Repair PEC | Hematology Oncology Pharmacy Anteroom | CoTo | TF-356B | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Anteroom Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | |
| CABINET, BIOLOGICAL SAFETY, CLASS II B2 | NUAIRE INC | NU435-600 | 145678080311 | TF-356A | Hematology Oncology Pharmacy Buffer Area | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Biological Safety Cabinets 2x/year, Recertify and Repair PEC | Hematology Oncology Pharmacy Buffer Area | CoTo | TF-356A | 180153063, 145718080411, 145678080311 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Buffer Area Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC |
| CABINET, BIOLOGICAL SAFETY, CLASS II B2 | NUAIRE INC | NU435-600 | 145718080411 | TF-356A | Hematology Oncology Pharmacy Buffer Area | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Biological Safety Cabinets 2x/year, Recertify and Repair PEC | ||||||
| CABINET, BIOLOGICAL SAFETY, CLASS II A2 | ESCO technologies | LA2-4A1-E | 2016-106629 | 136-3 | Nuclear Medicine SCA | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Biological Safety Cabinets 2x/year, Recertify and Repair PEC | ||||||
| CABINET, BIOLOGICAL SAFETY, CLASS II A2 | ESCO technologies | LA2-4A1-E | 2016-106630 | 136-3 | Nuclear Medicine SCA | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Biological Safety Cabinets 2x/year, Recertify and Repair PEC | ||||||
| CABINET, BIOLOGICAL SAFETY, CLASS II C1 | AXIOM | 304411100 | 180153063 | TF-356A | Hematology Oncology Pharmacy Buffer Area | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Test and Certify Biological Safety Cabinets 2x/year, Recertify and Repair PEC | ||||||
| Pharmacy Locations | |||||||||||||
| Secondary Engineering Control | Building | Room Number | PECs | Task/Requirement | Specific Tasks | ||||||||
| Support Pharmacy Anteroom | 3600 | L67-8A | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Anteroom Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | |||||||||
| Support Pharmacy Buffer Area | 3600 | L67-8 | 177096111516 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Buffer Area Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | ||||||||
| Nuclear Medicine SCA | 3600 | 136-3 | 2016-106630, 2016-106629 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | SCA Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | ||||||||
| OR/PACU SCA | 3600 | 239-7A | 177097111516 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | SCA Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | ||||||||
| Main Inpatient Pharmacy Anteroom | 3600 | 342-18 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Anteroom Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | |||||||||
| Main Inpatient Pharmacy Buffer Area | 3600 | 342-18A | 129690032409, 129689032409 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Buffer Area Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | ||||||||
| Maternal / Child Pharmacy Anteroom | 3600 | 537-3 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Anteroom Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | |||||||||
| Maternal / Child Pharmacy Buffer Area 1 (Left) | 3600 | 537-4 | 134312120209 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Buffer Area Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | ||||||||
| Maternal / Child Pharmacy Buffer Area 2 (Right) | 3600 | 537-5 | 120024122007, 120025122007 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Buffer Area Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | ||||||||
| CoTo OR Pharmacy Anteroom | CoTo | TC-216A | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Anteroom Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | |||||||||
| CoTo OR Pharmacy Buffer Area | CoTo | TC-216B | 145244071311, 145243071311 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Buffer Area Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | ||||||||
| Hematology Oncology Pharmacy Anteroom | CoTo | TF-356B | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Anteroom Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC | |||||||||
| Hematology Oncology Pharmacy Buffer Area | CoTo | TF-356A | 180153063, 145718080411, 145678080311 | PRS #1, #2, #3, #4, #5, #6, #7, #8 | Buffer Area Environmental Viable/Non-Viable Testing 2x/year, Recertify and Repair SEC |